510(k) · Cleared
Recalled · Class II · 2014FG-36UX Fiber Ultrasound Gastroscope
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Pentax Precision Instrument Corp. · ITX Transducer, Ultrasonic, Diagnostic
Cleared on Jun 9, 2004 after a 14-day review. The FDA summary names its predicate without a number. It has been recalled once.
- Decided
- Jun 9, 2004
- Review
- 14 days
- Class
- Class II
- Product code
- ITX
- Recalls
- 1
- Predicates
- 0
The FDA summary names its predicate without a number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in ITX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new ITX device
- #1 K260114 Align® Cover and Pad Guidance Toolscore 13.6 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | FG-36UX, Fiber Ultrasound Gastroscope Named without a submission number | |
| * | — | EUB-5500 Ultrasound Diagnostic Scanner Named without a submission number | |
| * | — | EUB-8500 Ultrasound Diagnostic Scanner Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 12, 2014 | Class II | Z-2116-2014 | Device Design | Aspiration needles used in combination with Ultrasound Gastroscopes can derail from the accessory elevator at the distal end of the endoscope. This may result in patient injury if the Instruction for Use is not followed carefully by the… |
No ancestor or descendant has a Class I recall.
How does its category compare?
All ITX devicesITX Transducer, Ultrasonic, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · ITX
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · ITX
126in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · ITX
2in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · ITX
571in 2025
MAUDE reports naming product code ITX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 10 | 26 | 0 | 16 |
| 2008 | 12 | 73 | 4 | 39 |
| 2009 | 17 | 28 | 4 | 23 |
| 2010 | 10 | 42 | 4 | 30 |
| 2011 | 10 | 16 | 1 | 34 |
| 2012 | 7 | 25 | 1 | 29 |
| 2013 | 3 | 105 | 0 | 38 |
| 2014 | 4 | 41 | 1 | 43 |
| 2015 | 1 | — | 5 | 183 |
| 2016 | 4 | 151 | 0 | 269 |
| 2017 | 6 | 110 | 1 | 412 |
| 2018 | 2 | — | 15 | 130 |
| 2019 | 1 | — | 0 | 119 |
| 2020 | 3 | 120 | 3 | 116 |
| 2021 | 3 | 139 | 10 (10) | 146 |
| 2022 | 2 | — | 6 | 87 |
| 2023 | 6 | 213 | 1 | 138 |
| 2024 | 3 | 81 | 1 | 988 |
| 2025 | 4 | 126 | 2 | 571 |
| 2026 | 6 | 158 | 0 | — |
What is it indicated for?
The FG-36UX, Fiber Ultrasound Gastroscope, is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract. The Upper Gastrointestinal Tract includes but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for the procedure…
Read the full statement (138 words)
The FG-36UX, Fiber Ultrasound Gastroscope, is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract. The Upper Gastrointestinal Tract includes but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for the procedure are observed in Adult and Pediatric patient populations.
The FG-36UX, Fiber Ultrasound Gastroscope, is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract. The Upper Gastrointestinal Tract includes but is not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for the procedure are observed in Adult and Pediatric patient populations.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K041396 · printed Nov 5, 2026
Cite this page
510(k) Report. “K041396: FG-36UX Fiber Ultrasound Gastroscope.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K041396
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026