510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Maxim Accel Knee System

K023546510(k) TraditionalCleared Jan 17, 200387-day reviewClass II21 CFR 888.3560Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biomet, Inc. · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Cleared on Jan 17, 2003 as substantially equivalent to K915132 (Biomet, Inc., 1992) and 1 other predicate after an 87-day review. Its family tree reaches back 1 generation to 1992. 305 later devices descend from it. It has been recalled 13 times.

Decided
Jan 17, 2003
Review
87 days
Class
Class II
Product code
JWH
Recalls
13
Predicates
2

Its longest line of descent runs back 1 generation to K915132 (1992). 2 ancestors in all. 305 later devices descend from it, 15 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor2011: 6 descendants cleared2006: 4 descendants cleared2008: 3 descendants cleared2005: 2 descendants cleared2019: 20 descendants cleared2013: 16 descendants cleared2004: 4 descendants cleared2012: 10 descendants cleared2015: 16 descendants cleared2021: 26 descendants cleared2014: 19 descendants cleared2017: 11 descendants cleared2018: 12 descendants cleared2010: 5 descendants cleared2020: 19 descendants cleared2007: 1 descendant cleared2023: 21 descendants cleared2022: 23 descendants cleared2016: 15 descendants cleared2026: 22 descendants cleared2024: 22 descendants cleared2025: 28 descendants clearedK915132 · MCK Maximum Congruent Knee) Knee System · Biomet, Inc. · 1992 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 15 descendants2 generations forward: 32 descendants3 generations forward: 55 descendants4 generations forward: 51 descendants5 generations forward: 64 descendants6 generations forward: 48 descendants7 generations forward: 29 descendants8 generations forward: 9 descendants9 generations forward: 1 descendant10 generations forward: 1 descendant K915132 · MCK Maximum Congruent Knee) Knee System · Biomet, Inc. · 1992 · recalled0+5+10
  1. Root · 1 generation backK915132 MCK Maximum Congruent Knee) Knee SystemBiomet, Inc. · 1992 · Recalled
  2. This deviceK023546 Maxim Accel Knee SystemBiomet, Inc. · 2003
  3. Descendant · cites it directly · 255 of its ownK050222 Vanguard Anterior Stabilized Tibial BearingsBiomet, Inc. · 2005 · Recalled
  4. Descendant · cites it directly · 249 of its ownK041046 Maxim Accel (Vanguard) PS+ BearingsBiomet, Inc. · 2004 · Recalled
  5. Descendant · cites it directly · 248 of its ownK080528 E-Poly Tibial BearingsBiomet Orthopedics, Inc. · 2008
  6. Descendant · 2 generations on · 119 of its ownK091956 Highly Cross-Linked Vitamin E Uhmwpe Tibial Inserts, Model…Encore Medical L.P. · 2010 · Recalled
  7. Descendant · 2 generations on · 114 of its ownK100048 E1 Antioxidant Infused TechnologyBiomet Manufacturing Corp · 2010

Showing 5 of 305 descendants. See all descendants

Descendants cleared per year

2004: 42005: 22006: 42007: 12008: 32010: 52011: 62012: 102013: 162014: 192015: 162016: 152017: 112018: 122019: 202020: 192021: 262022: 232023: 212024: 222025: 282026: 22
2004high 28 (2025) · low 0 (2009)2026

Most influential descendants

Showing 5 of 305 · See all
  • K050222Vanguard Anterior Stabilized Tibial BearingsBiomet, Inc. · 2005 · cites it directly255descendants
  • K041046Maxim Accel (Vanguard) PS+ BearingsBiomet, Inc. · 2004 · cites it directly249descendants
  • K080528E-Poly Tibial BearingsBiomet Orthopedics, Inc. · 2008 · cites it directly248descendants
  • K091956Highly Cross-Linked Vitamin E Uhmwpe Tibial Inserts, Model 391/392-09_19-702_712Encore Medical L.P. · 2010 · 2 generations on119descendants
  • K100048E1 Antioxidant Infused TechnologyBiomet Manufacturing Corp · 2010 · 2 generations on114descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K915132MCK Maximum Congruent Knee) Knee SystemBiomet, Inc.19926
0 (this device)K023546Maxim Accel Knee SystemBiomet, Inc.2003—
+1K050222Vanguard Anterior Stabilized Tibial BearingsBiomet, Inc.20051
+1K041046Maxim Accel (Vanguard) PS+ BearingsBiomet, Inc.20041
+1K080528E-Poly Tibial BearingsBiomet Orthopedics, Inc.2008—
+2K091956Highly Cross-Linked Vitamin E Uhmwpe Tibial Inserts, Model 391/392-09_19-702_712Encore Medical L.P.20104
+2K100048E1 Antioxidant Infused TechnologyBiomet Manufacturing Corp2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200440
200520
200640
200710
200830
201050
201160
2012100
2013160
2014190
2015160
2016150
2017110
2018120
2019200
2020190
2021260
2022230
2023210
2024220
2025280
2026220
FIG. 1 — Lineage of K023546, 1992–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JWH device

  1. #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K915132MCK Maximum Congruent Knee) Knee System
Summary page 1
…ined polymerimetaVpolymer (888.3560) Device Classification: Class I1 Device Product Code: 87 JWH Substantially Equivalent Devices: MCK (K915 132); Performance (K936274) Device Description: The Maxim Accel Knee System is intended to replace the articular portions.…
Biomet, Inc.19926
*K936274Kirschner Performance Posterior Stabilized Total Knee System
Summary page 1
…r (888.3560) Device Classification: Class I1 Device Product Code: 87 JWH Substantially Equivalent Devices: MCK (K915 132); Performance (K936274) Device Description: The Maxim Accel Knee System is intended to replace the articular portions. of the knee…
Kirschner Medical Corp.1995—

Also cleared against K915132: 13 other devices →

What descends from it? (305)

See all 305

15 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050222Vanguard Anterior Stabilized Tibial BearingsBiomet, Inc.2005255
K041046Maxim Accel (Vanguard) PS+ BearingsBiomet, Inc.2004249
K080528E-Poly Tibial BearingsBiomet Orthopedics, Inc.2008248
K033489Agc Total Knee System Knee, Ascent Total Knee System, Maxim Complete Knee System, Maxim…Biomet, Inc.200460
K060303Vanguard PS Open Box Porous Femoral ComponentsBiomet Manufacturing Corp200647
K113550Vanguard Complete Knee SystemBiomet Manufacturing Corp201246
K042757Vanguard SSK Knee SystemBiomet, Inc.200523
K053505Regenerex Ultra Porous Construct-Femoral and Tibial Knee AugmentsBiomet Manufacturing, Inc.200619
K033260AequosAap Implantate AG200419
K060525Vanguard Removable Molded Poly TibiaBiomet Manufacturing Corp200616

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 17, 2020Class IIZ-2150-2020Process controlPotential presence of elevated endotoxin levels that exceed the specification limit
Apr 17, 2020Class IIZ-2153-2020Process controlPotential presence of elevated endotoxin levels that exceed the specification limit
Oct 22, 2018Class IIZ-0495-2019Labeling mix-upsLabel on the outer carton or the patient labels does not match label on the inner sterile packaging
Oct 22, 2018Class IIZ-0496-2019Labeling mix-upsLabel on the outer carton or the patient labels does not match label on the inner sterile packaging
Sep 29, 2017Class IIZ-0673-2018Labeling mix-upsKnee implant components (tibial bearings) are labeled with the incorrect size.
Sep 29, 2017Class IIZ-0674-2018Labeling mix-upsKnee implant components (tibial bearings) are labeled with the incorrect size.
Jan 4, 2017Class IIZ-1115-2017Process change controlpackaging of some femoral components was incorrectly labeled as either smaller or larger and/or as the incorrect side (right/left). Risk delay in surgery.
Aug 9, 2016Class IIZ-0111-2017Labeling mix-upsContents of the package do not match the product labeling. Product is labeled: Vanguard Distal Femoral Augment 75 X 5 RL/LM Carton contains: Vanguard Distal Femoral Augment 75 X 5 LL/RM Risks: Adverse health outcomes that could occur as a…
Jul 6, 2011Not classifiedZ-2945-2011Employee errorThe recalled device was manufactured with two distal pegs but package labeling indicates the device is pegless.
Apr 20, 2010Not classifiedZ-2204-2010Process controlA knee implant labeled as left was opened and the package actually contained a right knee implant.
Feb 12, 2008Not classifiedZ-1375-2008Packaging process controlMislabeled: The packaging of the two products was mixed up, resulting in the products being labeled with an incorrect size. If implanted, the product may not fit properly, which could require repeat surgery. The problem could also result…
Feb 12, 2008Not classifiedZ-1376-2008Packaging process controlMislabeled: The packaging of the two products was mixed up, resulting in the products being labeled with an incorrect size. If implanted, the product may not fit properly, which could require repeat surgery. The problem could also result…
Aug 31, 2007Not classifiedZ-0123-2008Packaging process controlMislabeled as to size; labeled as 67.5 mm, actually contains 62.5 mm size femoral component.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K050222 (Biomet, Inc., 2005) has gone furthest, with 255 descendants of its own.
  • 13 other devices also cite its predicate K915132.

How does its category compare?

All JWH devices

JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JWH

20in 2025

2007: 182008: 172009: 112010: 192011: 252012: 332013: 312014: 332015: 272016: 182017: 252018: 252019: 282020: 222021: 242022: 172023: 202024: 192025: 202026 (partial year): 12
2007high 33 (2012) · low 11 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JWH

55in 2025

2007: 74 days median2008: 81 days median2009: 168 days median2010: 85 days median2011: 87 days median2012: 85 days median2013: 128 days median2014: 98 days median2015: 87 days median2016: 88 days median2017: 118 days median2018: 90 days median2019: 133 days median2020: 59 days median2021: 61 days median2022: 56 days median2023: 57 days median2024: 89 days median2025: 55 days median2026 (partial year): 63 days median
2007high 168 (2009) · low 55 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JWH

2in 2025

2007: 142008: 102009: 1092010: 322011: 272012: 852013: 312014: 132015: 112016: 172017: 612018: 142019: 52020: 22021: 82022: 122023: 672024: 72025: 22026 (partial year): 4
2007high 109 (2009) · low 2 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JWH

9,306in 2025

2007: 1,3052008: 1,4832009: 1,8702010: 3,2862011: 5,1332012: 5,2032013: 6,1352014: 6,7222015: 6,8812016: 7,5862017: 12,2632018: 14,0712019: 14,9872020: 12,5922021: 9,4082022: 8,3642023: 8,9472024: 10,8282025: 9,306
2007high 14,987 (2019) · low 1,305 (2007)2025

MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071874141,305
20081781101,483
2009111681091,870
20101985323,286
20112587275,133
20123385855,203
201331128316,135
20143398136,722
20152787116,881
20161888177,586
2017251186112,263
201825901414,071
201928133514,987
20202259212,592
2021246189,408
20221756128,364
20232057678,947
20241989710,828
2025205529,306
202612634—

What is it indicated for?

1) Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.

2) Correction of varus, valgus, or posttraumatic deformity.

3) Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

The device is a single use implant intended for implantation with bone cement.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023546: Maxim Accel Knee System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023546

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026