510(k) · Cleared
Recalled · 2009Bowie-Dick Mini Pak, Model BD 115
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Steritec Products, Inc. · JOJ Indicator, Physical/Chemical Sterilization Process
Cleared on Aug 11, 2003 after a 230-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 1 later device descends from it. It has been recalled once.
- Decided
- Aug 11, 2003
- Review
- 230 days
- Class
- Class II
- Product code
- JOJ
- Recalls
- 1
- Predicates
- 0
1 later device descends from it, 1 citing it directly.
- This deviceK024293 Bowie-Dick Mini Pak, Model BD 115Steritec Products, Inc. · 2003
- Descendant · cites it directly · 0 of its ownK250172 Green Card Bowie-Dick Test (BD115)Steritec Products, Inc. (A Getinge Company) · 2025
Most influential descendants
Showing 1 of 1 · See all- K250172Green Card Bowie-Dick Test (BD115)Steritec Products, Inc. (A Getinge Company) · 2025 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K024293 | Bowie-Dick Mini Pak, Model BD 115 | Steritec Products, Inc. | 2003 | — |
| +1 | K250172 | Green Card Bowie-Dick Test (BD115) | Steritec Products, Inc. (A Getinge Company) | 2025 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JOJ device
- #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K250172 | Green Card Bowie-Dick Test (BD115) | 2025 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 15, 2009 | Not classified | Z-1531-2010 | Other | This action notifies affected SteriTec customers as well as private label distributors of the necessity to modify the shelf life of their sterilization indicator products. There is insufficient data on file to support the shelf life… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K250172 (Steritec Products, Inc. (A Getinge Company), 2025) has gone furthest, with 0 descendants of its own.
How does its category compare?
All JOJ devicesJOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JOJ
6in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JOJ
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JOJ
1in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JOJ
12in 2025
MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 8 | 140 | 0 | 1 |
| 2008 | 9 | 57 | 2 | 1 |
| 2009 | 7 | 188 | 2 | 9 |
| 2010 | 6 | 115 | 0 | 3 |
| 2011 | 13 | 151 | 2 | 1 |
| 2012 | 7 | 224 | 0 | 8 |
| 2013 | 6 | 212 | 0 | 6 |
| 2014 | 6 | 198 | 0 | 11 |
| 2015 | 8 | 188 | 0 | 7 |
| 2016 | 4 | 150 | 0 | 11 |
| 2017 | 7 | 172 | 0 | 35 |
| 2018 | 5 | 150 | 0 | 12 |
| 2019 | 12 | 38 | 0 | 8 |
| 2020 | 7 | 108 | 0 | 6 |
| 2021 | 4 | 87 | 1 | 29 |
| 2022 | 8 | 90 | 0 | 23 |
| 2023 | 4 | 53 | 1 | 39 |
| 2024 | 12 | 71 | 0 | 25 |
| 2025 | 6 | 90 | 0 | 12 |
| 2026 | 3 | 30 | 0 | — |
What is it indicated for?
SteriTec Bowie-Dick Mini Pak is designed to detect the presence of residual air in pre-vacuum steam sterilizers operating at 132° C for 3.5 minutes. The indicator will demonstrate a uniform color change from purple to green when the proper sterilization conditions are met and no air is present. If enough air is present to create a 2° C or greater temperature difference in a standard Bowie-Dick test towel pack, the indicator will demonstrate a non-uniform color change.
Indications for use as stated in the FDA 510(k) statement.
https://510kreport.com/device/K024293 · printed Nov 3, 2026
Cite this page
510(k) Report. “K024293: Bowie-Dick Mini Pak, Model BD 115.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K024293
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026