510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

MODEL 780

K031383510(k) SpecialCleared Jul 7, 200367-day reviewClass II21 CFR 868.5450Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hudson Respiratory Care, Inc. · BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Cleared on Jul 7, 2003 as substantially equivalent to K881625 (Respiratory Care, Inc., 1988) after a 67-day review. Its family tree reaches back 1 generation to 1988. 2 of the 6 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 3 times.

Decided
Jul 7, 2003
Review
67 days
Class
Class II
Product code
BTT
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K881625 (1988). 4 later devices descend from it, 2 citing it directly. 2 of the 6 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2014: 2 descendants cleared, 1 with a Class I recall12006: 1 descendant cleared2017: 1 descendant cleared, 1 with a Class I recall1K881625 · Conchatherm Iii Plus Dual Heat Wire Servo Controll · Respiratory Care, Inc. · 19881990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 1 descendant, 1 with a Class I recall13 generations forward: 1 descendant, 1 with a Class I recall1 K881625 · Conchatherm Iii Plus Dual Heat Wire Servo Controll · Respiratory Care, Inc. · 19880
  1. Root · 1 generation backK881625 Conchatherm Iii Plus Dual Heat Wire Servo ControllRespiratory Care, Inc. · 1988
  2. This deviceK031383 MODEL 780Hudson Respiratory Care, Inc. · 2003
  3. Descendant · cites it directly · 2 of its ownK061736 Comfort Flo Humidification System and Nasal CannulaTeleflex Medical · 2006
  4. Descendant · 2 generations on · 1 of its ownK131912 Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification…Teleflex, Inc. · 2014 · Recalled (Class I)
  5. Descendant · 3 generations on · 0 of its ownK162242 Comfort Flo Humidification SystemTeleflexmedical, Inc. · 2017 · Recalled (Class I)
  6. Descendant · cites it directly · 0 of its ownK140556 Conchasmart Breathing CircuitTeleflex, Inc. · 2014

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K061736Comfort Flo Humidification System and Nasal CannulaTeleflex Medical · 2006 · cites it directly2descendants
  • K131912Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification System, Conchasmart…Teleflex, Inc. · 2014 · 2 generations on1descendant
  • K162242Comfort Flo Humidification SystemTeleflexmedical, Inc. · 2017 · 3 generations on0descendants
  • K140556Conchasmart Breathing CircuitTeleflex, Inc. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K881625Conchatherm Iii Plus Dual Heat Wire Servo ControllRespiratory Care, Inc.1988—
0 (this device)K031383MODEL 780Hudson Respiratory Care, Inc.2003—
+1K061736Comfort Flo Humidification System and Nasal CannulaTeleflex Medical2006—
+2K131912Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification System…Teleflex, Inc.2014Class I
+3K162242Comfort Flo Humidification SystemTeleflexmedical, Inc.2017Class I
+1K140556Conchasmart Breathing CircuitTeleflex, Inc.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
201421
201711
FIG. 1 — Lineage of K031383, 1988–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BTT device

  1. #1 K222197 F&P Optiflow Junior 2/2+ Nasal Cannula…score 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K881625Conchatherm Iii Plus Dual Heat Wire Servo Controll
Summary page 1
…7.1 (Circuit) Breathing Circuit 8.0 Predicate Devices 8.1 (Circuit) Current Hudson RCI Adult Heated Wire Circuit cleared in 510(k) submission K881625 9.0 Device Description Breathing circuit that is intended to administer medical gases to a patient. It provides both an inh…
Respiratory Care, Inc.1988—

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061736Comfort Flo Humidification System and Nasal CannulaTeleflex Medical20062
K140556Conchasmart Breathing CircuitTeleflex, Inc.20140

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 13, 2026Class IIZ-2991-2026Under Investigation by firmNeonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage.…
Jul 9, 2014Class IIZ-2107-2014Nonconforming Material/ComponentThe temperature probe does not properly connect to the temperature port.
Apr 17, 2014Class IIZ-0419-2016Process controlProduct is being recalled due to complaints that the inspiratory and expiratory tubes are transposed.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K131912Conchatherm Neptune Heated Humidifier, Comfort Flo Humidification System, Conchasmart ColumnTeleflex, Inc. · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K162242Comfort Flo Humidification SystemTeleflexmedical, Inc. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K061736 (Teleflex Medical, 2006) has gone furthest, with 2 descendants of its own.

How does its category compare?

All BTT devices

BTT Humidifier, Respiratory Gas, (Direct Patient Interface): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BTT

4in 2025

2007: 62008: 42009: 12010: 42011: 32012: 22013: 62014: 72015: 12016: 32017: 32018: 22019: 12020: 22021: 22022: 32023: 72024: 72025: 42026 (partial year): 4
2007high 7 (2014) · low 1 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BTT

104in 2025

2007: 73 days median2008: 199 days median2010: 113 days median2011: 116 days median2013: 171 days median2014: 231 days median2016: 157 days median2017: 303 days median2022: 290 days median2023: 258 days median2024: 219 days median2025: 104 days median2026 (partial year): 184 days median
2007high 303 (2017) · low 73 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BTT

1in 2025

2007: 12008: 12010: 22011: 12014: 5, 4 Class I2025: 12026 (partial year): 1
2007high 5 (2014) · low 0 (2009)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · BTT

1,743in 2025

2007: 3652008: 4562009: 5582010: 5622011: 6652012: 5832013: 7182014: 7182015: 6702016: 1,0942017: 1,0072018: 9182019: 8942020: 1,1882021: 1,5102022: 7492023: 7162024: 5012025: 1,743
2007high 1,743 (2025) · low 365 (2007)2025

MAUDE reports naming product code BTT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076731365
200841991456
20091—0558
201041132562
201131161665
20122—0583
201361710718
201472315 (4)718
20151—0670
2016315701,094
2017330301,007
20182—0918
20191—0894
20202—01,188
20212—01,510
202232900749
202372580716
202472190501
2025410411,743
202641841—

What is it indicated for?

The Hudson Heated Wire Ventilator Circuit is intended as a conduit for respiratory gas between a patient and a ventilator, and includes heated wires for use with a Concha Column Humidifier; the heated wires are intended to minimize condensation in the ventilator tubing.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031383: MODEL 780.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K031383

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026