510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Medtronic Minimed Paradigm Insulin Pump, Model MMT-712

K031390510(k) TraditionalCleared Jul 23, 200382-day reviewClass II21 CFR 880.5725Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Minimed, Inc. · LGZ Warmer, Thermal, Infusion Fluid

Cleared on Jul 23, 2003 after an 82-day review. The FDA summary names its predicate without a number. 1 of the 41 devices in its family, counting itself, has a Class I recall. 40 later devices descend from it. It has been recalled 2 times.

Decided
Jul 23, 2003
Review
82 days
Class
Class II
Product code
LGZ
Recalls
2
Predicates
0

40 later devices descend from it, 1 citing it directly. 1 of the 41 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2005: 1 descendant cleared2006: 1 descendant cleared2012: 1 descendant cleared2015: 2 descendants cleared2013: 2 descendants cleared2014: 1 descendant cleared2016: 2 descendants cleared2023: 4 descendants cleared2022: 2 descendants cleared2018: 2 descendants cleared2020: 5 descendants cleared, 1 with a Class I recall12017: 2 descendants cleared2025: 5 descendants cleared2024: 3 descendants cleared2021: 2 descendants cleared2026: 5 descendants cleared200020052010201520202025
  1. This deviceK031390 Medtronic Minimed Paradigm Insulin Pump, Model MMT-712Minimed, Inc. · 2003
  2. Descendant · cites it directly · 39 of its ownK042887 Accu-Chek SpiritDisetronic Medical Systems AG · 2005 · Recalled
  3. Descendant · 2 generations on · 38 of its ownK060876 Accu-Chek Insulin Pump Configuration Software (Standard), Model…Disetronic Medical Systems AG · 2006 · Recalled
  4. Descendant · 3 generations on · 37 of its ownK111353 Accu-Check Combo SystemRoche Diagnostics · 2012 · Recalled
  5. Descendant · 4 generations on · 33 of its ownK141929 Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation · 2015 · Recalled
  6. Descendant · 5 generations on · 32 of its ownK150910 ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc. · 2015 · Recalled

Showing 5 of 40 descendants. See all descendants

Descendants cleared per year

2005: 12006: 12012: 12013: 22014: 12015: 22016: 22017: 22018: 22020: 52021: 22022: 22023: 42024: 32025: 52026: 5
2005high 5 (2020) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 40 · See all
  • K042887Accu-Chek SpiritDisetronic Medical Systems AG · 2005 · cites it directly39descendants
  • K060876Accu-Chek Insulin Pump Configuration Software (Standard), Model 04625137001; Accu-Chek…Disetronic Medical Systems AG · 2006 · 2 generations on38descendants
  • K111353Accu-Check Combo SystemRoche Diagnostics · 2012 · 3 generations on37descendants
  • K141929Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation · 2015 · 4 generations on33descendants
  • K150910ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc. · 2015 · 5 generations on32descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K031390Medtronic Minimed Paradigm Insulin Pump, Model MMT-712Minimed, Inc.2003—
+1K042887Accu-Chek SpiritDisetronic Medical Systems AG20051
+2K060876Accu-Chek Insulin Pump Configuration Software (Standard), Model 04625137001…Disetronic Medical Systems AG20061
+3K111353Accu-Check Combo SystemRoche Diagnostics20123
+4K141929Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation20151
+5K150910ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc.20156

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200610
201210
201320
201410
201520
201620
201720
201820
202051
202120
202220
202340
202430
202550
202650
FIG. 1 — Lineage of K031390, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LGZ device

  1. #1 K260988 °M Warmer System with °M Stationscore 12.7 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Medtronic MiniMed Paradigm Model 511 Insulin Pump
Named without a submission number
—

What descends from it? (40)

See all 40

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042887Accu-Chek SpiritDisetronic Medical Systems AG200539

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 13, 2026Class IIZ-1748-2026Device DesignAll Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site…
Apr 27, 2007Not classifiedZ-1007-2007OtherExposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K193483Basal-IQ TechnologyTandem Diabetes Care, Inc. · 2020 · 8 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K042887 (Disetronic Medical Systems AG, 2005) has gone furthest, with 39 descendants of its own.

How does its category compare?

All LGZ devices

LGZ Warmer, Thermal, Infusion Fluid: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LGZ

2in 2025

2007: 22008: 42009: 22010: 22011: 32012: 72013: 52015: 22016: 22017: 22018: 12019: 32022: 32023: 12024: 12025: 22026 (partial year): 2
2007high 7 (2012) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LGZ

546in 2022

2008: 99 days median2011: 77 days median2012: 162 days median2013: 143 days median2019: 209 days median2022: 546 days median
2007high 546 (2022) · low 77 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LGZ

1in 2016

2009: 12010: 12011: 82015: 12016: 12026 (partial year): 1
2007high 8 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LGZ

376in 2025

2007: 112008: 102009: 122010: 432011: 542012: 522013: 162014: 232015: 342016: 152017: 582018: 942019: 3402020: 1,2212021: 8852022: 1,5482023: 1,0332024: 8622025: 376
2007high 1,548 (2022) · low 10 (2008)2025

MAUDE reports naming product code LGZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—011
2008499010
20092—112
20102—143
2011377854
20127162052
20135143016
20140—023
20152—134
20162—115
20172—058
20181—094
201932090340
20200—01,221
20210—0885
2022354601,548
20231—01,033
20241—0862
20252—0376
20262—1—

What is it indicated for?

The Medtronic MiniMed Paradigm Model 712 Insulin Pump is indicated for the continuous delivery of insulin, at set and variable rates, for thc management of diabetes mellitus in persons requiring insulin.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K031390: Medtronic Minimed Paradigm Insulin Pump, Model MMT-712.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031390

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026