510(k) Report

510(k) · Cleared

Recalled · 2009

Accu-Chek Insulin Pump Configuration Software (Standard), Model 04625137001; Accu-Chek Insulin Pump Configuration Softwa

K060876510(k) TraditionalCleared Jun 15, 200676-day reviewClass II21 CFR 880.5725Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Disetronic Medical Systems AG · LZG Pump, Infusion, Insulin

Cleared on Jun 15, 2006 as substantially equivalent to K042887 (Disetronic Medical Systems AG, 2005) after a 76-day review. Its family tree reaches back 2 generations to 2003. 1 of the 42 devices in its family, counting itself, has a Class I recall. 38 later devices descend from it. It has been recalled once.

Decided
Jun 15, 2006
Review
76 days
Class
Class II
Product code
LZG
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K031390 (2003). 3 ancestors in all. 38 later devices descend from it, 1 citing it directly. 1 of the 42 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2012: 1 descendant cleared2015: 2 descendants cleared2013: 2 descendants cleared2014: 1 descendant cleared2016: 2 descendants cleared2023: 4 descendants cleared2022: 2 descendants cleared2018: 2 descendants cleared2020: 5 descendants cleared, 1 with a Class I recall12017: 2 descendants cleared2025: 5 descendants cleared2024: 3 descendants cleared2021: 2 descendants cleared2026: 5 descendants clearedK031390 · Medtronic Minimed Paradigm Insulin Pump, Model MMT-712 · Minimed, Inc. · 2003 · recalledK042887 · Accu-Chek Spirit · Disetronic Medical Systems AG · 2005 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 3 descendants5 generations forward: 7 descendants6 generations forward: 9 descendants, 1 with a Class I recall17 generations forward: 5 descendants8 generations forward: 5 descendants9 generations forward: 2 descendants10 generations forward: 1 descendant K042887 · Accu-Chek Spirit · Disetronic Medical Systems AG · 2005 · recalledK031390 · Medtronic Minimed Paradigm Insulin Pump, Model MMT-712 · Minimed, Inc. · 2003 · recalled0+5+10
  1. Root · 2 generations backK031390 Medtronic Minimed Paradigm Insulin Pump, Model MMT-712Minimed, Inc. · 2003 · Recalled
  2. 1 generation backK042887 Accu-Chek SpiritDisetronic Medical Systems AG · 2005 · Recalled
  3. This deviceK060876 Accu-Chek Insulin Pump Configuration Software (Standard), Model…Disetronic Medical Systems AG · 2006
  4. Descendant · cites it directly · 37 of its ownK111353 Accu-Check Combo SystemRoche Diagnostics · 2012 · Recalled
  5. Descendant · 2 generations on · 33 of its ownK141929 Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation · 2015 · Recalled
  6. Descendant · 3 generations on · 32 of its ownK150910 ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc. · 2015 · Recalled
  7. Descendant · 4 generations on · 21 of its ownK160629 InPenCompanion Medical, Inc. · 2016 · Recalled
  8. Descendant · 5 generations on · 10 of its ownK190487 InPen Dose CalculatorCompanion Medical, Inc. · 2020 · Recalled

Showing 5 of 38 descendants. See all descendants

Descendants cleared per year

2012: 12013: 22014: 12015: 22016: 22017: 22018: 22020: 52021: 22022: 22023: 42024: 32025: 52026: 5
2012high 5 (2020) · low 0 (2019)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 38 · See all
  • K111353Accu-Check Combo SystemRoche Diagnostics · 2012 · cites it directly37descendants
  • K141929Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation · 2015 · 2 generations on33descendants
  • K150910ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc. · 2015 · 3 generations on32descendants
  • K160629InPenCompanion Medical, Inc. · 2016 · 4 generations on21descendants
  • K190487InPen Dose CalculatorCompanion Medical, Inc. · 2020 · 5 generations on10descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K031390Medtronic Minimed Paradigm Insulin Pump, Model MMT-712Minimed, Inc.20032
−1K042887Accu-Chek SpiritDisetronic Medical Systems AG20051
0 (this device)K060876Accu-Chek Insulin Pump Configuration Software (Standard), Model 04625137001…Disetronic Medical Systems AG2006—
+1K111353Accu-Check Combo SystemRoche Diagnostics20123
+2K141929Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation20151
+3K150910ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc.20156
+4K160629InPenCompanion Medical, Inc.20161
+5K190487InPen Dose CalculatorCompanion Medical, Inc.20201

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201320
201410
201520
201620
201720
201820
202051
202120
202220
202340
202430
202550
202650
FIG. 1 — Lineage of K060876, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZG device

  1. #1 K253534 Pivot Insulin Delivery Systemscore 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042887Accu-Chek Spirit
Summary page 2
…mp in this submission is a modification to the predicate device of the same name with 510(k): 510(k) #: K042887 ACCU-CHEK Spirit Insulin Infusion Pump Disetronic Medical Systems AG…
Disetronic Medical Systems AG20051

What descends from it? (38)

See all 38

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111353Accu-Check Combo SystemRoche Diagnostics201237

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 30, 2009Not classifiedZ-1646-2009Device DesignThe "up" and "down" buttons, which are used for changing the program in the menu or to administer additional insulin through a bolus delivery, may experience intermittent failure or completely cease to operate.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K193483Basal-IQ TechnologyTandem Diabetes Care, Inc. · 2020 · 6 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 3 years and passes through 2 companies; a new generation was cleared every 1.7 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K111353 (Roche Diagnostics, 2012) has gone furthest, with 37 descendants of its own.

How does its category compare?

All LZG devices

LZG Pump, Infusion, Insulin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LZG

2in 2024

2007: 32008: 62009: 22010: 22011: 42012: 32013: 22014: 42015: 12016: 42017: 12018: 12019: 22021: 12022: 12023: 12024: 22026 (partial year): 2
2007high 6 (2008) · low 0 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZG

116in 2016

2007: 112 days median2008: 129 days median2011: 189 days median2012: 73 days median2014: 123 days median2016: 116 days median
2007high 189 (2011) · low 73 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZG

1in 2022

2007: 12008: 132009: 22010: 42011: 22012: 42013: 3, 1 Class I2014: 2, 1 Class I2015: 2, 2 Class I2018: 3, 2 Class I2019: 52022: 1, 1 Class I2026 (partial year): 2, 2 Class I
2007high 13 (2008) · low 0 (2016)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · LZG

12,004in 2025

2007: 5,1102008: 5,7112009: 9,4712010: 10,1752011: 12,8582012: 18,3672013: 30,0592014: 49,1882015: 69,1362016: 56,5812017: 44,4722018: 32,6522019: 29,0272020: 32,5882021: 31,5222022: 28,2252023: 21,9722024: 18,2282025: 12,004
2007high 69,136 (2015) · low 5,110 (2007)2025

MAUDE reports naming product code LZG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007311215,110
20086129135,711
20092—29,471
20102—410,175
20114189212,858
2012373418,367
20132—3 (1)30,059
201441232 (1)49,188
20151—2 (2)69,136
20164116056,581
20171—044,472
20181—3 (2)32,652
20192—529,027
20200—032,588
20211—031,522
20221—1 (1)28,225
20231—021,972
20242—018,228
20250—012,004
20262—2 (2)—

What is it indicated for?

The ACCU-CHEK Spirit Insulin Infusion Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician. The ACCU-CHEK Insulin Pump Configuration Software facilitates the monitoring and programming of the pump settings.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060876: Accu-Chek Insulin Pump Configuration Software (Standard), Model 04625137001; Accu-Chek Insulin Pump Configuration Softwa.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K060876

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026