510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Toxcup Drug Screen Cup Amp/Coc/Met/Opi/PCP/THC

K032057510(k) SpecialCleared Aug 21, 200350-day reviewClass II21 CFR 862.3250Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Branan Medical Corp. · DIO Enzyme Immunoassay, Cocaine And Cocaine Metabolites

Cleared on Aug 21, 2003 after a 50-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 2 times.

Decided
Aug 21, 2003
Review
50 days
Class
Class II
Product code
DIO
Recalls
2
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K032057Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DIO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DIO device

  1. #1 K213211 Carolina Liquid Chemistries Cocaine and…score 3.0 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 18, 2020Class IIZ-3045-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Mar 26, 2014Class IIZ-1508-2014Labeling Change ControlMay be labeled with incorrect result interpretation graphics on the ToxCup lid label.

No ancestor or descendant has a Class I recall.

How does its category compare?

All DIO devices

DIO Enzyme Immunoassay, Cocaine And Cocaine Metabolites: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DIO

1in 2022

2009: 12011: 52012: 52013: 22014: 22015: 22016: 32017: 32018: 12020: 12021: 12022: 1
2007high 5 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DIO

267in 2017

2011: 65 days median2012: 139 days median2016: 88 days median2017: 267 days median
2007high 267 (2017) · low 65 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DIO

1in 2013

2013: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DIO

1in 2023

2011: 42014: 12015: 72016: 92017: 162018: 122019: 222022: 12023: 1
2007high 22 (2019) · low 0 (2007)2025

MAUDE reports naming product code DIO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—00
20100—00
201156504
2012513900
20132—10
20142—01
20152—07
201638809
20173267016
20181—012
20190—022
20201—00
20211—00
20221—01
20230—01
20240—00
20250—00
20260—0—

What is it indicated for?

The ToxCupTMDrug Screen Cup is an in vitro screen test that contains chromatographic immunoassays for the rapid detection of amphetamines, cocaine (benzoylecgonine), methamphetamines, opiates, phencyclidine and THC in human urine at the following cutoff concentrations: AMP — d-Amphetamine — 1000 ng/ml COC — Benzoylecgonine — 300 ng/ml MET500 — d-Methamphetamine — 500 ng/ml 0PI2000 — Morphine — 2000 ng/ml OPI300* — Morphine — 300ng/mI * PCP —…

Read the full statement (135 words)

The ToxCupTMDrug Screen Cup is an in vitro screen test that contains chromatographic immunoassays for the rapid detection of amphetamines, cocaine (benzoylecgonine), methamphetamines, opiates, phencyclidine and THC in human urine at the following cutoff concentrations:

AMP — d-Amphetamine — 1000 ng/ml

COC — Benzoylecgonine — 300 ng/ml

MET500 — d-Methamphetamine — 500 ng/ml

0PI2000 — Morphine — 2000 ng/ml

OPI300* — Morphine — 300ng/mI *

PCP — Phencyclidine — 25 ng/ml

THC — 11-nor-A9-Tetrahydrocannabinol-9-carboxylicacid — 50 ng/ml

*The Opiates test is offered at the SAMHSA mandated cut-off concentration of 2000 n g h . and the optional cut-off concentration indicated.

The ToxCupTMDrug Screen Cup provides visual, qualitative results for multiple drugs-of-abused in human urine. The device is intended for professional in vitro diagnostic use only. It is not intended for over-the-counter sale to lay persons.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K032057: Toxcup Drug Screen Cup Amp/Coc/Met/Opi/PCP/THC.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K032057

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)