510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Toxcup Drug Screen Cup

K122633510(k) TraditionalCleared Feb 7, 2013162-day reviewClass II21 CFR 862.3870Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Branan Medical Corp. · LDJ Enzyme Immunoassay, Cannabinoids

Cleared on Feb 7, 2013 as substantially equivalent to K082898 (Amedica Biotech, Inc., 2008) after a 162-day review. Its family tree reaches back 1 generation to 2008. It has been recalled 8 times.

Decided
Feb 7, 2013
Review
162 days
Class
Class II
Product code
LDJ
Recalls
8
Predicates
1

Its longest line of descent runs back 1 generation to K082898 (2008). No later device cites it as a predicate yet.

K082898 · Amedica Drug Screen Test Cup · Amedica Biotech, Inc. · 2008 · recalled20062008201020122014
  1. Root · 1 generation backK082898 Amedica Drug Screen Test CupAmedica Biotech, Inc. · 2008 · Recalled
  2. This deviceK122633 Toxcup Drug Screen CupBranan Medical Corp. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K082898Amedica Drug Screen Test CupAmedica Biotech, Inc.20082
0 (this device)K122633Toxcup Drug Screen CupBranan Medical Corp.2013—
FIG. 1 — Lineage of K122633, 2008–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LDJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LDJ device

  1. #1 K232732 Rapid Marijuana (THC) Test Strip 20, Rapid…score 9.4 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082898Amedica Drug Screen Test Cup
Summary page 2
…JXN 11 862.3700 Propoxyphene Test System Device Classification Class 11 Substantially Equivalent Devices K082898 - Amedica Home Drug Test U~p 140 Technology Dr, Suite 4oo, Irvine, CA 92618 e-mail: in…
Amedica Biotech, Inc.20082

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 18, 2020Class IIZ-3032-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Aug 18, 2020Class IIZ-3047-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Aug 18, 2020Class IIZ-3033-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Aug 18, 2020Class IIZ-3039-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Aug 18, 2020Class IIZ-3031-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Aug 18, 2020Class IIZ-3035-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Aug 18, 2020Class IIZ-3040-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Aug 18, 2020Class IIZ-3045-2020Process controlVarious immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.

How does its category compare?

All LDJ devices

LDJ Enzyme Immunoassay, Cannabinoids: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LDJ

2in 2024

2008: 22009: 12011: 42012: 42013: 32014: 22015: 32016: 12019: 12023: 12024: 2
2007high 4 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LDJ

48in 2015

2011: 219 days median2012: 193 days median2013: 162 days median2015: 48 days median
2007high 219 (2011) · low 48 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LDJ

1in 2023

2008: 42023: 1
2007high 4 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LDJ

1in 2015

2015: 1
2007high 1 (2015) · low 0 (2007)2025

MAUDE reports naming product code LDJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—40
20091—00
20100—00
2011421900
2012419300
2013316200
20142—00
201534801
20161—00
20170—00
20180—00
20191—00
20200—00
20210—00
20220—00
20231—10
20242—00
20250—00
20260—0—

What is it indicated for?

The ToxCup® Drug Screen Cup is an in vitro screening test for the rapid detection of multiple drugs and drug metabolites in human urine at or above the following cutoff concentration: AMP Amphetamine 500 ng/ml BAR Secobarbital 300 nglml BUPG Buprenorphine Glucuronide 10 ng/ml BZO Oxazepam 300 ng/ml COC Benzoylecgonine 150 nglml MDMA 3,4-methylenedioxymethamphetamine 500 ng/ml MET Methamphetamine 500 ng/ml MTD Methadone 300 ng/ml OPI Morphine 300 ng/ml…

Read the full statement (218 words)

The ToxCup® Drug Screen Cup is an in vitro screening test for the rapid detection of multiple drugs and drug metabolites in human urine at or above the following cutoff concentration:

AMP Amphetamine 500 ng/ml

BAR Secobarbital 300 nglml

BUPG Buprenorphine Glucuronide 10 ng/ml

BZO Oxazepam 300 ng/ml

COC Benzoylecgonine 150 nglml

MDMA 3,4-methylenedioxymethamphetamine 500 ng/ml

MET Methamphetamine 500 ng/ml

MTD Methadone 300 ng/ml

OPI Morphine 300 ng/ml

OXY Oxycodone 100 ng/ml

PCP Phencyclidine 25 ng/rnl

PPX Propoxyphene 300 ng/ml

TCA Nortriptyline 1000 ng/ml

THC 11-nor-A9-Tetrahydrocannabinol-9-carboxylic acid 50 ng/ml

These tests provide visual qualitative results and are intended for in vitro diagnostic use only. The ToxCup® Drug Screen Cup is available in double drug analyte cassette dip format. It is intended for prescription point-of-care use and over-the-counter consumer use.

These tests provide only a preliminary test result and are the first step in a two-step process for detecting drugs of abuse in urine. The second step is confirming the results in a certified laboratory. For a quantitative result or to confirm preliminary positive results obtained by the ToxCup® Drug Screen Cup, a more specific alternative method such as Gas Chromatography/Mass Spectrometry (GC/MS) must be used. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when a preliminary positive result is indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K122633: Toxcup Drug Screen Cup.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K122633

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026