510(k) Report

510(k) · Cleared

Recalled · 2009

Surgassist Circular Stapler Digital Loading Units (CS21, CS25, CS29, CS33)

K032701510(k) SpecialCleared Sep 30, 200328-day reviewClass II21 CFR 878.4750Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Power Medical Interventions, Inc. · GDW Staple, Implantable

Cleared on Sep 30, 2003 as substantially equivalent to K003277 (Powermed, Inc., 2000) and 1 other predicate after a 28-day review. Its family tree reaches back 1 generation to 1992. 1 later device descends from it. It has been recalled 3 times.

Decided
Sep 30, 2003
Review
28 days
Class
Class II
Product code
GDW
Recalls
3
Predicates
2

Its longest line of descent runs back 1 generation to K920752 (1992). 2 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2004: 1 descendant clearedK920752 · Endopath Ils Endoscopic Circular Stapler · ETHICON, Inc. · 19921990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant K920752 · Endopath Ils Endoscopic Circular Stapler · ETHICON, Inc. · 19920
  1. Root · 1 generation backK920752 Endopath Ils Endoscopic Circular StaplerETHICON, Inc. · 1992
  2. This deviceK032701 Surgassist Circular Stapler Digital Loading Units (CS21, CS25, CS29…Power Medical Interventions, Inc. · 2003
  3. Descendant · cites it directly · 0 of its ownK040024 Surgassist Circular Stapler Digital Loading Units - 29 MM, Model CS29Power Medical Interventions, Inc. · 2004 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K040024Surgassist Circular Stapler Digital Loading Units - 29 MM, Model CS29Power Medical Interventions, Inc. · 2004 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K920752Endopath Ils Endoscopic Circular StaplerETHICON, Inc.1992—
0 (this device)K032701Surgassist Circular Stapler Digital Loading Units (CS21, CS25, CS29, CS33)Power Medical Interventions, Inc.2003—
+1K040024Surgassist Circular Stapler Digital Loading Units - 29 MM, Model CS29Power Medical Interventions, Inc.20041

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
FIG. 1 — Lineage of K032701, 1992–2004Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GDW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GDW device

  1. #1 K253657 Tri-staple 2.0™ Reloads; Endo GIA™ Reloads…score 15.1 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K003277Surgassist System with Circular Stapler Disposable Loading Unit, Model CS21,CS25,CS29,CS33
Summary page 1
…mm, 25 mm, 29 mm, 33 mm), with Titanium Implantable Staple, Power Medical Interventions, Inc., New Hope, PA. REF CS21, CS25, CS29, CS33 (K003277). B. Endopath ILS Endoscopic Circular Staplers, 21 mm, 25 mm, 29 mm, 33 mm, Ethicon Endo-Surgery, Cincinnat…
Powermed, Inc.20003
*K920752Endopath Ils Endoscopic Circular Stapler
Summary page 1
…CS29, CS33 (K003277). B. Endopath ILS Endoscopic Circular Staplers, 21 mm, 25 mm, 29 mm, 33 mm, Ethicon Endo-Surgery, Cincinnati, OH. (K920752). 000014 Power Medical Interventions, Inc. SurgA…
ETHICON, Inc.1992—

Also cleared against K003277: 6 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K040024Surgassist Circular Stapler Digital Loading Units - 29 MM, Model CS29Power Medical Interventions, Inc.20040

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 23, 2009Not classifiedZ-0393-2010Nonconforming Material/ComponentLack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in extended surgical…
Nov 20, 2003Not classifiedZ-0252-04Otherimproper staple formation or anvil jam
Nov 20, 2003Not classifiedZ-0253-04Otherimproper staple formation or anvil jam

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K040024 (Power Medical Interventions, Inc., 2004) has gone furthest, with 0 descendants of its own.
  • 6 other devices also cite its predicate K003277.

How does its category compare?

All GDW devices

GDW Staple, Implantable: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GDW

3in 2025

2007: 82008: 72009: 132010: 102011: 112012: 92013: 132014: 152015: 82016: 112017: 132018: 72019: 122020: 102021: 92022: 72023: 42024: 22025: 32026 (partial year): 3
2007high 15 (2014) · low 2 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GDW

111in 2025

2007: 56 days median2008: 57 days median2009: 27 days median2010: 39 days median2011: 49 days median2012: 45 days median2013: 49 days median2014: 63 days median2015: 111 days median2016: 96 days median2017: 103 days median2018: 90 days median2019: 205 days median2020: 88 days median2021: 233 days median2022: 89 days median2023: 267 days median2025: 111 days median2026 (partial year): 141 days median
2007high 267 (2023) · low 27 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GDW

3in 2025

2007: 92009: 222010: 22011: 92012: 11, 1 Class I2013: 12015: 62016: 22017: 32018: 32019: 4, 4 Class I2020: 112023: 12025: 3
2007high 22 (2009) · low 0 (2008)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · GDW

11,905in 2025

2007: 2,4012008: 2,3802009: 4,0142010: 3,7682011: 3,8262012: 3,7172013: 3,7772014: 4,7842015: 4,8992016: 5,1062017: 11,6432018: 14,9142019: 16,6002020: 12,3322021: 13,0152022: 13,6052023: 14,6232024: 13,2802025: 11,905
2007high 16,600 (2019) · low 2,380 (2008)2025

MAUDE reports naming product code GDW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200785692,401
200875702,380
20091327224,014
2010103923,768
2011114993,826
201294511 (1)3,717
2013134913,777
2014156304,784
2015811164,899
2016119625,106
201713103311,643
2018790314,914
2019122054 (4)16,600
202010881112,332
20219233013,015
2022789013,605
20234267114,623
20242—013,280
20253111311,905
202631410—

What is it indicated for?

The SurgASSIST Circular Stapler Digital Loading Units have applications throughout the alimentary tract for end-to-end, end-to-side, and side-to-side anastomoses.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K032701: Surgassist Circular Stapler Digital Loading Units (CS21, CS25, CS29, CS33).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K032701

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026