510(k) · Cleared
Recalled · 2011Barbiturate Enzyme Immunoassay, CA. # 0140 (500 Tests Kit); Cat# 0141 (5000 Tests Kit)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Lin-Zhi International, Inc. · DIS Enzyme Immunoassay, Barbiturate
Cleared on Nov 3, 2003 as substantially equivalent to K010934 (Syva Co., Dade Behring, Inc., 2001) and 1 other predicate after a 59-day review. Its family tree reaches back 1 generation to 1994. 1 later device descends from it. It has been recalled once.
- Decided
- Nov 3, 2003
- Review
- 59 days
- Class
- Class II
- Product code
- DIS
- Recalls
- 1
- Predicates
- 2
Its longest line of descent runs back 1 generation to K935101 (1994). 2 ancestors in all. 1 later device descends from it, 1 citing it directly.
- Root · 1 generation backK935101 Urine Calibrators and Controls ModificationDiagnostic Reagents, Inc. · 1994
- This deviceK032764 Barbiturate Enzyme Immunoassay, CA. # 0140 (500 Tests Kit); Cat#…Lin-Zhi International, Inc. · 2003
- Descendant · cites it directly · 0 of its ownK033885 BMBP Enzyme Immunoassay, Cat #0610, 0611 (500, 5000 Test Kit)Lin-Zhi International, Inc. · 2004
Most influential descendants
Showing 1 of 1 · See all- K033885BMBP Enzyme Immunoassay, Cat #0610, 0611 (500, 5000 Test Kit)Lin-Zhi International, Inc. · 2004 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K935101 | Urine Calibrators and Controls Modification | Diagnostic Reagents, Inc. | 1994 | — |
| 0 (this device) | K032764 | Barbiturate Enzyme Immunoassay, CA. # 0140 (500 Tests Kit); Cat# 0141 (5000… | Lin-Zhi International, Inc. | 2003 | — |
| +1 | K033885 | BMBP Enzyme Immunoassay, Cat #0610, 0611 (500, 5000 Test Kit) | Lin-Zhi International, Inc. | 2004 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2001 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DIS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K010934 | Emit II Plus Barbituate Assay, Model 0SR9D0229 | 2001 |
| * | K935101 | Urine Calibrators and Controls ModificationSummary page 6…and controls included in the Multi- Drugs Urine Calibrators and Controls (Diagnostic Reagents, Inc., now Microgenics Corporation), cleared under premarket notifications (K935 101) for Drugs of Abuse Urine Calibrators and Controls.… | 1994 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K033885 | BMBP Enzyme Immunoassay, Cat #0610, 0611 (500, 5000 Test Kit) | 2004 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 8, 2011 | Not classified | Z-0846-2012 | Nonconforming Material/Component | Catalog #0140 and 0141 Barbiturate EIA small and large kits were not correctly made to specification such that the two antibodies used in production did not have curves that correctly aligned. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- None of the 2 ancestors has a recall.
- Of its descendants, K033885 (Lin-Zhi International, Inc., 2004) has gone furthest, with 0 descendants of its own.
How does its category compare?
All DIS devicesDIS Enzyme Immunoassay, Barbiturate: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DIS
1in 2016
2026 to Oct 25, outlined.
Median 510(k) review, days · DIS
29in 2015
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DIS
1in 2016
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DIS
1in 2018
MAUDE reports naming product code DIS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 331 | 0 | 1 |
| 2008 | 0 | — | 0 | 1 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 2 | — | 1 | 0 |
| 2012 | 0 | — | 1 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 2 | — | 0 | 0 |
| 2015 | 3 | 29 | 0 | 0 |
| 2016 | 1 | — | 1 | 0 |
| 2017 | 0 | — | 0 | 0 |
| 2018 | 0 | — | 0 | 1 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 0 | 0 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 0 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Barbiturate Drug of Abuse Calibrators are intended for in vitro diagnostic use for the calibration of the Barbiturate Enzyme Immunoassay to detect barbiturates in human urine. The Barbiturate Drug of Abuse Controls are intended for in vitro diagnostic use for the validation of the Barbiturate Enzyme Immunoassay to detect barbiturates in human urine. The Barbiturate Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 200 ng/mL and/or 300…
Read the full statement (185 words)
The Barbiturate Drug of Abuse Calibrators are intended for in vitro diagnostic use for the calibration of the Barbiturate Enzyme Immunoassay to detect barbiturates in human urine.
The Barbiturate Drug of Abuse Controls are intended for in vitro diagnostic use for the validation of the Barbiturate Enzyme Immunoassay to detect barbiturates in human urine.
The Barbiturate Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 200 ng/mL and/or 300 ng/mL cutoffs. The assay is intended for use in the qualitative and semi-quantitative analyses of barbiturates in human urine. The assay is designed for professional use with a number of automated clinical chemistry analyzers.
Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.
The Barbiturate Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GCMS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K032764 · printed Nov 5, 2026
Cite this page
510(k) Report. “K032764: Barbiturate Enzyme Immunoassay, CA. # 0140 (500 Tests Kit); Cat# 0141 (5000 Tests Kit).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K032764
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026