510(k) Report

510(k) · Cleared

Cleared

Press-Mate BX-10 Vital Sign Monitor

K032857510(k) TraditionalCleared Feb 4, 2004145-day reviewClass II21 CFR 870.1130

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Colin Medical Instruments Corp. · DXN System, Measurement, Blood-Pressure, Non-Invasive

Cleared on Feb 4, 2004 as substantially equivalent to K890876 (Colin Medical Instruments Corp., 1989) and 4 other predicates after a 145-day review. Its family tree reaches back 3 generations to 1987. 19 later devices descend from it.

Decided
Feb 4, 2004
Review
145 days
Class
Class II
Product code
DXN
Recalls
None
Predicates
5

Its longest line of descent runs back 3 generations to K872291 (1987). 6 ancestors in all. 19 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1994: 1 other ancestor1993: 1 other ancestor1989: 1 other ancestor2011: 1 descendant cleared2008: 1 descendant cleared2012: 2 descendants cleared2014: 1 descendant cleared2022: 3 descendants cleared2017: 1 descendant cleared2016: 2 descendants cleared2018: 1 descendant cleared2020: 1 descendant cleared2025: 2 descendants cleared2021: 1 descendant cleared2024: 2 descendants cleared2026: 1 descendant clearedK872291 · Escort Portable Patient Monitor · Medical Data Electronics · 1987K973637 · Press-Mate Advantage · Colin Medical Instruments Corp. · 1998K022537 · Press-Mate PM-2100 Vital Sign Monitor · Colin Medical Instruments Corp. · 20021990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 3 descendants3 generations forward: 3 descendants4 generations forward: 3 descendants5 generations forward: 3 descendants6 generations forward: 2 descendants7 generations forward: 2 descendants K022537 · Press-Mate PM-2100 Vital Sign Monitor · Colin Medical Instruments Corp. · 2002K973637 · Press-Mate Advantage · Colin Medical Instruments Corp. · 1998K872291 · Escort Portable Patient Monitor · Medical Data Electronics · 19870+5
  1. Root · 3 generations backK872291 Escort Portable Patient MonitorMedical Data Electronics · 1987
  2. 2 generations backK973637 Press-Mate AdvantageColin Medical Instruments Corp. · 1998
  3. 1 generation backK022537 Press-Mate PM-2100 Vital Sign MonitorColin Medical Instruments Corp. · 2002
  4. This deviceK032857 Press-Mate BX-10 Vital Sign MonitorColin Medical Instruments Corp. · 2004
  5. Descendant · cites it directly · 15 of its ownK113623 Patient MonitorEdan Instruments, Inc. · 2012 · Recalled
  6. Descendant · 2 generations on · 14 of its ownK131971 IM60/IM70 Patient MonitorEdan Instruments, Inc. · 2014 · Recalled
  7. Descendant · 3 generations on · 8 of its ownK160981 Patient MonitorEdan Instruments, Inc. · 2016 · Recalled
  8. Descendant · 4 generations on · 6 of its ownK192514 Patient MonitorEdan Instruments, Inc. · 2020 · Recalled
  9. Descendant · 5 generations on · 3 of its ownK202336 Patient MonitorEdan Instruments, Inc. · 2021 · Recalled

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

2008: 12011: 12012: 22014: 12016: 22017: 12018: 12020: 12021: 12022: 32024: 22025: 22026: 1
2008high 3 (2022) · low 0 (2009)2026

Most influential descendants

Showing 5 of 19 · See all
  • K113623Patient MonitorEdan Instruments, Inc. · 2012 · cites it directly15descendants
  • K131971IM60/IM70 Patient MonitorEdan Instruments, Inc. · 2014 · 2 generations on14descendants
  • K160981Patient MonitorEdan Instruments, Inc. · 2016 · 3 generations on8descendants
  • K192514Patient MonitorEdan Instruments, Inc. · 2020 · 4 generations on6descendants
  • K202336Patient MonitorEdan Instruments, Inc. · 2021 · 5 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K872291Escort Portable Patient MonitorMedical Data Electronics1987—
−2K973637Press-Mate AdvantageColin Medical Instruments Corp.1998—
−1K022537Press-Mate PM-2100 Vital Sign MonitorColin Medical Instruments Corp.2002—
0 (this device)K032857Press-Mate BX-10 Vital Sign MonitorColin Medical Instruments Corp.2004—
+1K113623Patient MonitorEdan Instruments, Inc.20122
+2K131971IM60/IM70 Patient MonitorEdan Instruments, Inc.20142
+3K160981Patient MonitorEdan Instruments, Inc.20161
+4K192514Patient MonitorEdan Instruments, Inc.20201
+5K202336Patient MonitorEdan Instruments, Inc.20211

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199310
199410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201110
201220
201410
201620
201710
201810
202010
202110
202230
202420
202520
202610
FIG. 1 — Lineage of K032857, 1987–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXN device

  1. #1 K222125 Arm Blood Pressure Monitor, models AOJ-30A…score 18.5 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K890876BP-8800 Blood Pressure Monitor
Summary page 2
…from an external AC power source or an internal rechargeable battery. The device uses the same technology and materials as the predicate devices, the Press - Mate 8800 (K890876 cleared 6/22/89),Press - Mate Advantage (K 973637 cleared 9/25/98) and the Press-Mate Prodigy (K022537 cleared 9/13/02) The fo…
Colin Medical Instruments Corp.1989—
*K921048BP8800T/8000IR/8300G Non-Invasive Blood Pres. Syst
Summary page 3
…cated devices with respect to indications for use, device design, materials, and method of manufacture. Colin Corporation, Press-Mate 8800 Monitor (K890876; K921048C; K942871) Colin Corporation, Press-Mate Advantage Monitor (K973637) Colin Medical Instruments, Press-Mate Prodigy(PM-2100…
Colin Medical Instruments Corp.1993—
*K942871Press-Mate BP8800MS Automatic Non-Invasive Blood Pressure System
Summary page 3
…ces with respect to indications for use, device design, materials, and method of manufacture. Colin Corporation, Press-Mate 8800 Monitor (K890876; K921048C; K942871) Colin Corporation, Press-Mate Advantage Monitor (K973637) Colin Medical Instruments, Press-Mate Prodigy(PM-2100) (K022537…
Colin Medical Instruments Corp.1994—
*K973637Press-Mate Advantage
Summary page 2
…rechargeable battery. The device uses the same technology and materials as the predicate devices, the Press - Mate 8800 (K890876 cleared 6/22/89),Press - Mate Advantage (K 973637 cleared 9/25/98) and the Press-Mate Prodigy (K022537 cleared 9/13/02) The following accessories are available for use with the…
Colin Medical Instruments Corp.1998—
*K022537Press-Mate PM-2100 Vital Sign Monitor
Summary page 2
…ogy and materials as the predicate devices, the Press - Mate 8800 (K890876 cleared 6/22/89),Press - Mate Advantage (K 973637 cleared 9/25/98) and the Press-Mate Prodigy (K022537 cleared 9/13/02) The following accessories are available for use with the device: 1. Power cord 2. Printer Paper 3. Operation…
Colin Medical Instruments Corp.2002—

Also cleared against K890876: 1 other device →

What descends from it? (19)

See all 19

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K113623Patient MonitorEdan Instruments, Inc.201215
K110922Patient MonitorEdan Instruments, Inc.20111
K082045Modification to Mcare 300 Vital Signs Monitor, Model 91220Spacelabs Medical, Inc.20080

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 2 companies; a new generation was cleared every 4.0 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K113623 (Edan Instruments, Inc., 2012) has gone furthest, with 15 descendants of its own.
  • 1 other device also cites its predicate K890876.

How does its category compare?

All DXN devices

DXN System, Measurement, Blood-Pressure, Non-Invasive: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DXN

28in 2025

2007: 272008: 232009: 462010: 432011: 342012: 362013: 342014: 292015: 302016: 302017: 332018: 222019: 332020: 242021: 232022: 252023: 352024: 172025: 282026 (partial year): 29
2007high 46 (2009) · low 17 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXN

142in 2025

2007: 118 days median2008: 82 days median2009: 73 days median2010: 94 days median2011: 72 days median2012: 116 days median2013: 150 days median2014: 133 days median2015: 97 days median2016: 144 days median2017: 184 days median2018: 235 days median2019: 156 days median2020: 198 days median2021: 190 days median2022: 137 days median2023: 182 days median2024: 107 days median2025: 142 days median2026 (partial year): 147 days median
2007high 235 (2018) · low 72 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXN

3in 2024

2008: 12009: 32011: 12013: 12016: 12019: 12024: 3
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXN

239in 2025

2007: 212008: 82009: 172010: 142011: 102012: 192013: 722014: 472015: 622016: 542017: 902018: 6172019: 962020: 1302021: 1622022: 1552023: 1622024: 2702025: 239
2007high 617 (2018) · low 8 (2008)2025

MAUDE reports naming product code DXN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200727118021
2008238218
20094673317
20104394014
20113472110
201236116019
201334150172
201429133047
20153097062
201630144154
201733184090
2018222350617
201933156196
2020241980130
2021231900162
2022251370155
2023351820162
2024171073270
2025281420239
2026291470—

What is it indicated for?

The Press-Mate BX-10 and BX-10An are indicated for use on adult, pediatric, and neonatal patients by a trained and qualified healthcare professional in the hospital, outpatient surgery practitioner facilities, or in an environment where patient care is provided by qualified healthcare personnel who will determine when use of this device is indicated, based upon their professional assessment of the patient's medical condition. The Press-Mate BX-10 patient…

Read the full statement (108 words)

The Press-Mate BX-10 and BX-10An are indicated for use on adult, pediatric, and neonatal patients by a trained and qualified healthcare professional in the hospital, outpatient surgery practitioner facilities, or in an environment where patient care is provided by qualified healthcare personnel who will determine when use of this device is indicated, based upon their professional assessment of the patient's medical condition.

The Press-Mate BX-10 patient monitor is intended to monitor a single patient's non-invasive blood pressure and pulse rate.

The Press-Mate BX-1OAn patient monitor is intended to monitor a single patient's non-invasive blood pressure as well as blood oxygen saturation and pulse rate via Nellcor's oximetry technology.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K032857: Press-Mate BX-10 Vital Sign Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K032857

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026