510(k) · Cleared
Recalled · 2010Modification to Mcare 300 Vital Signs Monitor, Model 91220
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Spacelabs Medical, Inc. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)
Cleared on Aug 1, 2008 as substantially equivalent to K062095 (Spacelabs Medical, Inc., 2006) and 1 other predicate after a 14-day review. Its family tree reaches back 8 generations to 1997. It has been recalled once.
- Decided
- Aug 1, 2008
- Review
- 14 days
- Class
- Class II
- Product code
- MHX
- Recalls
- 1
- Predicates
- 2
Its longest line of descent runs back 8 generations to K962603 (1997). 34 ancestors in all, the oldest K872291 (1987). No later device cites it as a predicate yet.
- Root · 8 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
- 7 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
- 6 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
- 5 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
- 4 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
- 3 generations backK021090 Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc. · 2002
- 2 generations backK031740 VSM - Vital Signs Monitor, Model 53000 SeriesWelch Allyn Protocol, Inc. · 2003
- 1 generation backK062095 Mcare 300 Vital Signs Monitor, Model 91220Spacelabs Medical, Inc. · 2006
- This deviceK082045 Modification to Mcare 300 Vital Signs Monitor, Model 91220Spacelabs Medical, Inc. · 2008
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −8 | K962603 | Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables | Masimo Corp. | 1997 | — |
| −7 | K973887 | Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables | Masimo Corp. | 1998 | — |
| −6 | K982255 | Model 2000 Pulse Oximeter | Ivy Biomedical Systems, Inc. | 1998 | — |
| −5 | K991823 | N-395 Pulse Oximeter, Model N-395 | Nellcor Puritan Bennett, Inc. | 1999 | — |
| −4 | K012891 | Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and… | Nellcor Puritan Bennett, Inc. | 2002 | 1 |
| −3 | K021090 | Oximax N-550 Pulse Oximeter,Model N-550 Pulse Oximeter | Nellcor Puritan Bennett, Inc. | 2002 | — |
| −2 | K031740 | VSM - Vital Signs Monitor, Model 53000 Series | Welch Allyn Protocol, Inc. | 2003 | — |
| −1 | K062095 | Mcare 300 Vital Signs Monitor, Model 91220 | Spacelabs Medical, Inc. | 2006 | — |
| 0 (this device) | K082045 | Modification to Mcare 300 Vital Signs Monitor, Model 91220 | Spacelabs Medical, Inc. | 2008 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1987 | 1 | 0 |
| 1989 | 1 | 0 |
| 1990 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 2 | 0 |
| 1994 | 1 | 0 |
| 1995 | 1 | 0 |
| 1996 | 4 | 0 |
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 2 | 0 |
| 2002 | 3 | 0 |
| 2003 | 2 | 0 |
| 2004 | 4 | 0 |
| 2006 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MHX device
- #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K062095 | Mcare 300 Vital Signs Monitor, Model 91220Summary page 2…substantially equivalent in design concepts, Discussion: technologies and materials to the mCare 300 that was originally cleared under 5 10(k) number K062095 and the Press-Mate BX-10 Vital Signs Monitor (BX- 10) by Colin (now Omron) that was cleared under 510(k)… | 2006 |
| * | K032857 | Press-Mate BX-10 Vital Sign MonitorSummary page 2…er 5 10(k) number K062095 and the Press-Mate BX-10 Vital Signs Monitor (BX- 10) by Colin (now Omron) that was cleared under 510(k) number K032857. The mCare 300 originally incorporated a Welch Allyn NIBP module, the revised mCare 300 and the b… | 2004 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 14, 2010 | Not classified | Z-3185-2011 | Component design/selection | There have been reports that the Spacelabs Medical Vital Signs Monitor, Model 91220, will not power up with AC power and will no longer charge the batteries. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 11 years and passes through 5 companies; a new generation was cleared every 1.3 years on median.
- 12 of the 34 ancestors have a recall of any class.
- 2 other devices also cite its predicate K062095.
How does its category compare?
All MHX devicesMHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · MHX
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MHX
192in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MHX
21so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MHX
3,761in 2025
MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 15 | 96 | 3 | 131 |
| 2008 | 16 | 58 | 2 | 232 |
| 2009 | 13 | 57 | 7 | 382 |
| 2010 | 20 | 90 | 13 | 955 |
| 2011 | 12 | 81 | 16 | 949 |
| 2012 | 16 | 88 | 11 | 676 |
| 2013 | 21 | 120 | 14 | 1,997 |
| 2014 | 13 | 248 | 6 | 1,609 |
| 2015 | 14 | 203 | 5 | 1,284 |
| 2016 | 13 | 149 | 7 | 1,105 |
| 2017 | 14 | 134 | 6 | 1,115 |
| 2018 | 11 | 203 | 12 | 2,157 |
| 2019 | 7 | 243 | 6 | 1,462 |
| 2020 | 11 | 196 | 0 | 1,714 |
| 2021 | 5 | 195 | 0 | 2,423 |
| 2022 | 15 | 189 | 6 | 1,470 |
| 2023 | 8 | 229 | 3 | 2,254 |
| 2024 | 5 | 181 | 3 | 2,814 |
| 2025 | 8 | 192 | 15 | 3,761 |
| 2026 | 14 | 205 | 21 | — |
What is it indicated for?
The Spacelabs Medical, mCare 300, Model 91220, Vital Signs Monitor is indicated for use in adult, pediatric and neonate patient populations wherever there is a need for the monitoring of ECG, respiration, invasive or noninvasive blood pressures, body temperature, functional arterial oxygen saturation, or expired or minimum inspire CO 2.
The Spacelabs Medical, mCare 300, Model 91220, Vital Signs Monitor is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K082045 · printed Nov 6, 2026
Cite this page
510(k) Report. “K082045: Modification to Mcare 300 Vital Signs Monitor, Model 91220.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082045
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026