510(k) Report

510(k) · Cleared

Recalled · 2009

Tidal Wave SP Model 710/715 and 715

K032971510(k) TraditionalCleared Aug 26, 2004339-day reviewClass II21 CFR 870.2700Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics Novametrix, Inc. · DQA Oximeter

Cleared on Aug 26, 2004 as substantially equivalent to K963327 (Novametrix Medical Systems, Inc., 1996) and 2 other predicates after a 339-day review. Its family tree reaches back 1 generation to 1992. 1 of the 61 devices in its family, counting itself, has a Class I recall. 57 later devices descend from it. It has been recalled 7 times.

Decided
Aug 26, 2004
Review
339 days
Class
Class II
Product code
DQA
Recalls
7
Predicates
3

Its longest line of descent runs back 1 generation to K920379 (1992). 3 ancestors in all. 57 later devices descend from it, 4 citing it directly. 1 of the 61 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1993: 1 other ancestor2009: 1 descendant cleared2004: 1 descendant cleared2010: 4 descendants cleared2013: 3 descendants cleared2018: 5 descendants cleared2015: 1 descendant cleared2016: 4 descendants cleared2008: 2 descendants cleared2006: 1 descendant cleared2020: 4 descendants cleared, 1 with a Class I recall12012: 4 descendants cleared2011: 2 descendants cleared2023: 3 descendants cleared2014: 2 descendants cleared2022: 5 descendants cleared2021: 4 descendants cleared2017: 2 descendants cleared2019: 1 descendant cleared2024: 4 descendants cleared2026: 2 descendants cleared2025: 2 descendants clearedK920379 · Novametrix Model 7100 Comb. End Tidal CO2 & Pul OX · Novametrix Medical Systems, Inc. · 1992 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 4 descendants2 generations forward: 13 descendants3 generations forward: 14 descendants, 1 with a Class I recall14 generations forward: 9 descendants5 generations forward: 7 descendants6 generations forward: 8 descendants7 generations forward: 2 descendants K920379 · Novametrix Model 7100 Comb. End Tidal CO2 & Pul OX · Novametrix Medical Systems, Inc. · 1992 · recalled0+5
  1. Root · 1 generation backK920379 Novametrix Model 7100 Comb. End Tidal CO2 & Pul OXNovametrix Medical Systems, Inc. · 1992 · Recalled
  2. This deviceK032971 Tidal Wave SP Model 710/715 and 715Respironics Novametrix, Inc. · 2004
  3. Descendant · cites it directly · 39 of its ownK083821 Vital Signs MonitorEdan Instruments, Inc. · 2009
  4. Descendant · 2 generations on · 16 of its ownK093016 Vital Signs MonitorShenzhen Creative Industry Co., Ltd. · 2010
  5. Descendant · 2 generations on · 15 of its ownK113623 Patient MonitorEdan Instruments, Inc. · 2012 · Recalled
  6. Descendant · cites it directly · 15 of its ownK042601 Capnostat 5Respironics Novametrix, Inc. · 2004 · Recalled
  7. Descendant · 3 generations on · 14 of its ownK131971 IM60/IM70 Patient MonitorEdan Instruments, Inc. · 2014 · Recalled

Showing 5 of 57 descendants. See all descendants

Descendants cleared per year

2004: 12006: 12008: 22009: 12010: 42011: 22012: 42013: 32014: 22015: 12016: 42017: 22018: 52019: 12020: 42021: 42022: 52023: 32024: 42025: 22026: 2
2004high 5 (2018) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 57 · See all
  • K083821Vital Signs MonitorEdan Instruments, Inc. · 2009 · cites it directly39descendants
  • K093016Vital Signs MonitorShenzhen Creative Industry Co., Ltd. · 2010 · 2 generations on16descendants
  • K113623Patient MonitorEdan Instruments, Inc. · 2012 · 2 generations on15descendants
  • K042601Capnostat 5Respironics Novametrix, Inc. · 2004 · cites it directly15descendants
  • K131971IM60/IM70 Patient MonitorEdan Instruments, Inc. · 2014 · 3 generations on14descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K920379Novametrix Model 7100 Comb. End Tidal CO2 & Pul OXNovametrix Medical Systems, Inc.19923
0 (this device)K032971Tidal Wave SP Model 710/715 and 715Respironics Novametrix, Inc.2004—
+1K083821Vital Signs MonitorEdan Instruments, Inc.2009—
+2K093016Vital Signs MonitorShenzhen Creative Industry Co., Ltd.2010—
+2K113623Patient MonitorEdan Instruments, Inc.20122
+1K042601Capnostat 5Respironics Novametrix, Inc.20041
+3K131971IM60/IM70 Patient MonitorEdan Instruments, Inc.20142

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200610
200820
200910
201040
201120
201240
201330
201420
201510
201640
201720
201850
201910
202041
202140
202250
202330
202440
202520
202620
FIG. 1 — Lineage of K032971, 1992–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQA device

  1. #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K963327Tidal Wave Carbon Dioxide Monitor, Model 610
Summary page 1
…ly published accuracy specifications with the specified predicate devices. The predicate devices are as follows: 1. Tidal Wave Model 610 (510(k) No. K963327) dated 11/20/1996 2. C0 2SMO, Model 7100 (510(k) No. K920379/A), dated 07/15/1992 3. Spot Check Model 510 (51…
Novametrix Medical Systems, Inc.19964
*K920379Novametrix Model 7100 Comb. End Tidal CO2 & Pul OX
Summary page 1
…cate devices. The predicate devices are as follows: 1. Tidal Wave Model 610 (510(k) No. K963327) dated 11/20/1996 2. C0 2SMO, Model 7100 (510(k) No. K920379/A), dated 07/15/1992 3. Spot Check Model 510 (510(k) No. K924626), dated 12/03/1993. e. Device Description…
Novametrix Medical Systems, Inc.19923
*K924626Novametrix Model 510
Summary page 1
…e Model 610 (510(k) No. K963327) dated 11/20/1996 2. C0 2SMO, Model 7100 (510(k) No. K920379/A), dated 07/15/1992 3. Spot Check Model 510 (510(k) No. K924626), dated 12/03/1993. e. Device Description The Tidal Wave Models 710/715 handheld combined pulse oximeter/capnograph…
Novametrix Medical Systems, Inc.1993—

Also cleared against K963327: 3 other devices →

What descends from it? (57)

See all 57

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083821Vital Signs MonitorEdan Instruments, Inc.200939
K042601Capnostat 5Respironics Novametrix, Inc.200415
K092611Sonarmed Airway Monitoring System, Model 0901Sonarmed, Inc.20102
K123797NT1D Vital Signs MonitorNew Tech, Inc.20130

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 19, 2009Not classifiedZ-1279-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1278-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1292-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1276-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1290-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1277-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1263-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K193058SonarMed AirWave Airway Monitoring SystemSonarmed, Inc. · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K083821 (Edan Instruments, Inc., 2009) has gone furthest, with 39 descendants of its own.
  • 3 other devices also cite its predicate K963327.

How does its category compare?

All DQA devices

DQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQA

15in 2025

2007: 202008: 222009: 352010: 292011: 152012: 142013: 152014: 112015: 132016: 142017: 92018: 182019: 192020: 152021: 102022: 202023: 132024: 142025: 152026 (partial year): 8
2007high 35 (2009) · low 9 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQA

182in 2025

2007: 122 days median2008: 97 days median2009: 117 days median2010: 127 days median2011: 195 days median2012: 120 days median2013: 182 days median2014: 175 days median2015: 158 days median2016: 246 days median2017: 200 days median2018: 260 days median2019: 158 days median2020: 126 days median2021: 174 days median2022: 265 days median2023: 317 days median2024: 285 days median2025: 182 days median2026 (partial year): 187 days median
2007high 317 (2023) · low 97 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQA

3in 2024

2007: 62008: 52009: 292010: 42011: 12014: 22015: 12018: 322024: 3
2007high 32 (2018) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQA

399in 2025

2007: 3252008: 3122009: 2812010: 1852011: 1642012: 3222013: 4682014: 1,1242015: 2,0822016: 1,7752017: 1,1902018: 6942019: 5482020: 4262021: 5532022: 4782023: 3922024: 5452025: 399
2007high 2,082 (2015) · low 164 (2011)2025

MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007201226325
200822975312
20093511729281
2010291274185
2011151951164
2012141200322
2013151820468
20141117521,124
20151315812,082
20161424601,775
2017920001,190
20181826032694
2019191580548
2020151260426
2021101740553
2022202650478
2023133170392
2024142853545
2025151820399
202681870—

What is it indicated for?

The intended use of the TidalWave Sp (Models 710/715) is to provide short term monitoring of carbon dioxide and oxygen saturation during anesthesia / recovery, in the intensive care unit (ICU). and in Emergency Medicine/Transport or Respiratory care. Separate airway adapters are provided for pediatric/adult and neonatal/pediatric use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K032971: Tidal Wave SP Model 710/715 and 715.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K032971

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026