510(k) · Cleared
Recalled · 2011Elekta Synergy System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Elekta Oncology Systems, Ltd. · IYE Accelerator, Linear, Medical
Cleared on Oct 23, 2003 as substantially equivalent to K963624 (Philips Medical Systems, Inc., 1996) after a 28-day review. Its family tree reaches back 1 generation to 1996. 85 later devices descend from it. It has been recalled 3 times.
- Decided
- Oct 23, 2003
- Review
- 28 days
- Class
- Class II
- Product code
- IYE
- Recalls
- 3
- Predicates
- 1
Its longest line of descent runs back 1 generation to K963624 (1996). 85 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK963624 Philips Sli Series Linear AcceleratorPhilips Medical Systems, Inc. · 1996 · Recalled
- This deviceK032996 Elekta Synergy SystemElekta Oncology Systems, Ltd. · 2003
- Descendant · cites it directly · 79 of its ownK040192 ON-Board Imager DeviceVarian Medical Systems, Inc. · 2004 · Recalled
- Descendant · 2 generations on · 76 of its ownK042720 ON-Board Imager DeviceVarian Medical Systems · 2004 · Recalled
- Descendant · 3 generations on · 61 of its ownK092871 Trilogy MXVarian Medical Systems, Inc. · 2009 · Recalled
- Descendant · 4 generations on · 43 of its ownK111106 TruebeamVarian Medical Systems, Inc. · 2011 · Recalled
- Descendant · 5 generations on · 37 of its ownK123291 TruebeamVarian Medical Systems, Inc. · 2012
Showing 5 of 85 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 85 · See all- K040192ON-Board Imager DeviceVarian Medical Systems, Inc. · 2004 · cites it directly79descendants
- K042720ON-Board Imager DeviceVarian Medical Systems · 2004 · 2 generations on76descendants
- K092871Trilogy MXVarian Medical Systems, Inc. · 2009 · 3 generations on61descendants
- K111106TruebeamVarian Medical Systems, Inc. · 2011 · 4 generations on43descendants
- K123291TruebeamVarian Medical Systems, Inc. · 2012 · 5 generations on37descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K963624 | Philips Sli Series Linear Accelerator | Philips Medical Systems, Inc. | 1996 | 2 |
| 0 (this device) | K032996 | Elekta Synergy System | Elekta Oncology Systems, Ltd. | 2003 | — |
| +1 | K040192 | ON-Board Imager Device | Varian Medical Systems, Inc. | 2004 | 2 |
| +2 | K042720 | ON-Board Imager Device | Varian Medical Systems | 2004 | 5 |
| +3 | K092871 | Trilogy MX | Varian Medical Systems, Inc. | 2009 | 5 |
| +4 | K111106 | Truebeam | Varian Medical Systems, Inc. | 2011 | 5 |
| +5 | K123291 | Truebeam | Varian Medical Systems, Inc. | 2012 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 3 | 0 |
| 2005 | 1 | 0 |
| 2008 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 3 | 0 |
| 2011 | 3 | 0 |
| 2012 | 3 | 0 |
| 2013 | 3 | 0 |
| 2014 | 3 | 0 |
| 2015 | 3 | 0 |
| 2016 | 2 | 0 |
| 2017 | 7 | 0 |
| 2018 | 5 | 0 |
| 2019 | 5 | 0 |
| 2020 | 5 | 0 |
| 2021 | 8 | 0 |
| 2022 | 6 | 0 |
| 2023 | 11 | 0 |
| 2024 | 3 | 0 |
| 2025 | 2 | 0 |
| 2026 | 7 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYE device
- #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K963624 | Philips Sli Series Linear AcceleratorSummary page 1…The Elekta Synergy"'MSystem is an enhancement to the previously reported Precise Treatment Systems Digital Accelerator (formally designated the SLISLi Series, D.C. K963624) and its commercially available options. These devices have an established and proven track record for safety. The primary… | 1996 |
What descends from it? (85)
See all 853 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 6, 2011 | Not classified | Z-3079-2011 | Other | A small number of the Modular Reverse Diodes (D1A), a component in the equipment room of the Elekta Linear Accelerator, have ruptured. |
| Mar 18, 2008 | Not classified | Z-1441-2010 | Finished device change control | Installation of a touchguard to prevent injury to patients or staff by stopping machine movements in the case of accidental collision. |
| Oct 1, 2005 | Not classified | Z-1159-2010 | Device Design | Customer reported larger than expected variations in wedge factory with gantry rotation. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All IYE devicesIYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · IYE
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYE
170in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYE
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · IYE
101in 2025
MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 23 | 44 | 78 | 155 |
| 2008 | 14 | 69 | 49 | 114 |
| 2009 | 20 | 63 | 22 | 174 |
| 2010 | 23 | 74 | 27 | 246 |
| 2011 | 11 | 94 | 73 | 280 |
| 2012 | 28 | 106 | 20 | 128 |
| 2013 | 18 | 88 | 11 | 96 |
| 2014 | 20 | 100 | 14 | 124 |
| 2015 | 20 | 108 | 8 | 100 |
| 2016 | 21 | 97 | 7 | 97 |
| 2017 | 23 | 94 | 7 | 125 |
| 2018 | 21 | 58 | 7 | 85 |
| 2019 | 19 | 83 | 2 | 91 |
| 2020 | 20 | 67 | 4 | 74 |
| 2021 | 19 | 86 | 5 | 68 |
| 2022 | 12 | 172 | 3 | 66 |
| 2023 | 18 | 129 | 7 | 128 |
| 2024 | 9 | 239 | 1 | 81 |
| 2025 | 12 | 170 | 1 | 101 |
| 2026 | 14 | 120 | 4 | — |
What is it indicated for?
The Elekta Synergym System is intended to be used for radiation therapy treatment of malignant neophstic diseases, as determinedby a licensed medical practitioner.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K032996 · printed Nov 5, 2026
Cite this page
510(k) Report. “K032996: Elekta Synergy System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K032996
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026