Elekta Synergy; Elekta Synergy S and XVIR3.5

K051932510(k) SpecialCleared Aug 12, 200530-day reviewClass II21 CFR 892.5050Recalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Elekta, Inc. · IYE Accelerator, Linear, Medical

Cleared on Aug 12, 2005 as substantially equivalent to K032996 (Elekta Oncology Systems, Ltd., 2003) after a 30-day review. Its family tree reaches back 2 generations to 1996. 17 later devices descend from it. It has been recalled 14 times.

Decided
Aug 12, 2005
Review
30 days
Class
Class II
Product code
IYE
Recalls
14
Predicates
1

Its longest line of descent runs back 2 generations to K963624 (1996). 17 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2018: 2 descendants cleared2008: 1 descendant cleared2010: 2 descendants cleared2013: 1 descendant cleared2019: 2 descendants cleared2011: 1 descendant cleared2016: 2 descendants cleared2015: 1 descendant cleared2017: 1 descendant cleared2021: 1 descendant cleared2026: 1 descendant cleared2022: 1 descendant cleared2025: 1 descendant clearedK963624 · Philips Sli Series Linear Accelerator · Philips Medical Systems, Inc. · 1996 · recalledK032996 · Elekta Synergy System · Elekta Oncology Systems, Ltd. · 2003 · recalled1995200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 3 descendants3 generations forward: 5 descendants4 generations forward: 3 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 1 descendant K032996 · Elekta Synergy System · Elekta Oncology Systems, Ltd. · 2003 · recalledK963624 · Philips Sli Series Linear Accelerator · Philips Medical Systems, Inc. · 1996 · recalled0+5
  1. Root · 2 generations backK963624 Philips Sli Series Linear AcceleratorPhilips Medical Systems, Inc. · 1996 · Recalled
  2. 1 generation backK032996 Elekta Synergy SystemElekta Oncology Systems, Ltd. · 2003 · Recalled
  3. This deviceK051932 Elekta Synergy; Elekta Synergy S and XVIR3.5Elekta, Inc. · 2005
  4. Descendant · cites it directly · 9 of its ownK100115 XVI R4.5Elekta, Ltd. · 2010 · Recalled
  5. Descendant · 2 generations on · 8 of its ownK131965 XVI R5.0Elekta Limited · 2013
  6. Descendant · 3 generations on · 4 of its ownK151561 Leksell Gamma Knife IconElekta Instrument AB · 2015 · Recalled
  7. Descendant · cites it directly · 3 of its ownK182138 Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDEleckta Limited · 2018
  8. Descendant · 4 generations on · 3 of its ownK160440 Leksell Gamma Knife IconElekta Instrument AB · 2016 · Recalled

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2008: 12010: 22011: 12013: 12015: 12016: 22017: 12018: 22019: 22021: 12022: 12025: 12026: 1
2008high 2 (2010) · low 0 (2009)2026

Most influential descendants

Showing 5 of 17 · See all
  • K100115XVI R4.5Elekta, Ltd. · 2010 · cites it directly9descendants
  • K131965XVI R5.0Elekta Limited · 2013 · 2 generations on8descendants
  • K151561Leksell Gamma Knife IconElekta Instrument AB · 2015 · 3 generations on4descendants
  • K182138Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDEleckta Limited · 2018 · cites it directly3descendants
  • K160440Leksell Gamma Knife IconElekta Instrument AB · 2016 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K963624Philips Sli Series Linear AcceleratorPhilips Medical Systems, Inc.19962
−1K032996Elekta Synergy SystemElekta Oncology Systems, Ltd.20033
0 (this device)K051932Elekta Synergy; Elekta Synergy S and XVIR3.5Elekta, Inc.2005—
+1K100115XVI R4.5Elekta, Ltd.20104
+2K131965XVI R5.0Elekta Limited2013—
+3K151561Leksell Gamma Knife IconElekta Instrument AB20151
+1K182138Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDEleckta Limited2018—
+4K160440Leksell Gamma Knife IconElekta Instrument AB20161

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201020
201110
201310
201510
201620
201710
201820
201920
202110
202210
202510
202610
FIG. 1 — Lineage of K051932, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032996Elekta Synergy System
Summary page 1
…ar Accelerator (with Patient Imaging) Classification Name: Medical Linear Accelerator Accessory 90 lYE Predicate Device: Elekta Synergy® System (K032996) Product Description: This Premarket Notification Special 510(k) describes modifications to the Elekta Synergy® Syst…
Elekta Oncology Systems, Ltd.20033

Also cleared against K032996: 2 other devices →

What descends from it? (17)

See all 17

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K100115XVI R4.5Elekta, Ltd.20109
K182138Precise Treatment System; Synergy platform; Synergy; Infinity; VersaHDEleckta Limited20183
K080585Desktop ProElekta, Ltd.20082

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 1, 2017Class IIZ-2840-2018Component design/selectionThere is a potential for an uncontrolled extension of iViewGT / XVI detector arms.
Jul 8, 2016Class IIZ-2721-2016OtherThere have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
Jan 7, 2015Class IIZ-1088-2015OtherElekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
Nov 25, 2013Class IIZ-1061-2014Software designPotential collision risk when using XVI and external beam shaping devices.
May 28, 2013Class IIZ-1065-2014Device DesignCustomers are resetting the default values outside recommended factory settings.
Dec 19, 2012Class IIZ-0931-2013Software designThe DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
Aug 6, 2012Class IIZ-2405-2012Nonconforming Material/ComponentThere have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.
May 6, 2011Not classifiedZ-3079-2011OtherA small number of the Modular Reverse Diodes (D1A), a component in the equipment room of the Elekta Linear Accelerator, have ruptured.
Mar 9, 2011Not classifiedZ-0305-2012Software designIf "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs to MOSAIQ a a zero…
Jan 5, 2011Not classifiedZ-1252-2011OtherFlexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180.
Oct 25, 2010Not classifiedZ-1723-2011OtherIf another patient is selected on the imaging system (XVI) release 4.5) during the transmission of end of treatment data from Desktop Pro R7.01 to R&V system, the XVI information is prioritized and the end of treatment data is never…
Sep 1, 2010Not classifiedZ-0350-2011OtherIt is possible to do couch corrections twice if both Couch Move Assistant in MOSAIQ and Remote Automatic Table in XVI are configured when in SYNERGISTIQ mode.
Dec 12, 2008Not classifiedZ-1405-2010Software designIt is possible that two (or more) CT reference data sets are necessary for one patient, for example, if you have to scan a patient during breath hold and then again during free breathing.
Sep 19, 2007Not classifiedZ-1411-2010Device DesignFirm became aware of a failure in the position detection system which could result in incorrect positioning of the detector in the longitudinal or lateral direction.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 6.9 years on median.
  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K100115 (Elekta, Ltd., 2010) has gone furthest, with 9 descendants of its own.
  • 2 other devices also cite its predicate K032996.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The Elekta Synergy®, Elekta Synergy® S, and XVI R3.5 are intended

to be used for radiation therapy treatment of malignant neoplastic

diseases, as determined by a licensed physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051932: Elekta Synergy; Elekta Synergy S and XVIR3.5.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051932

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026