510(k) Report

510(k) · Cleared

Recalled · 2010

Sonoline Antares with Clarify VE

K033196510(k) TraditionalCleared Oct 16, 200314-day reviewClass II21 CFR 892.1550Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IYN System, Imaging, Pulsed Doppler, Ultrasonic

Cleared on Oct 16, 2003 after a 14-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 991 devices in its family, counting itself, have a Class I recall. 990 later devices descend from it. It has been recalled once.

Decided
Oct 16, 2003
Review
14 days
Class
Class II
Product code
IYN
Recalls
1
Predicates
0

990 later devices descend from it, 5 citing it directly. 2 of the 991 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 3 descendants cleared2006: 5 descendants cleared2007: 11 descendants cleared2005: 5 descendants cleared2008: 14 descendants cleared2009: 21 descendants cleared2011: 29 descendants cleared2010: 17 descendants cleared2015: 59 descendants cleared2013: 42 descendants cleared2012: 36 descendants cleared2014: 50 descendants cleared2016: 59 descendants cleared2019: 49 descendants cleared2017: 77 descendants cleared2018: 68 descendants cleared2020: 75 descendants cleared2021: 54 descendants cleared2023: 57 descendants cleared2024: 65 descendants cleared2025: 80 descendants cleared, 2 with a Class I recall22022: 59 descendants cleared2026: 55 descendants cleared200020052010201520202025
  1. This deviceK033196 Sonoline Antares with Clarify VESiemens Medical Solutions USA, Inc. · 2003
  2. Descendant · cites it directly · 988 of its ownK041319 Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2004
  3. Descendant · 2 generations on · 979 of its ownK051139 Acuson Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005
  4. Descendant · 3 generations on · 972 of its ownK052410 Acuson Sequoia Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005
  5. Descendant · 2 generations on · 932 of its ownK042770 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  6. Descendant · 3 generations on · 927 of its ownK050240 Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2005 · Recalled

Showing 5 of 990 descendants. See all descendants

Descendants cleared per year

2004: 32005: 52006: 52007: 112008: 142009: 212010: 172011: 292012: 362013: 422014: 502015: 592016: 592017: 772018: 682019: 492020: 752021: 542022: 592023: 572024: 652025: 802026: 55
2004high 80 (2025) · low 3 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 990 · See all
  • K041319Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2004 · cites it directly988descendants
  • K051139Acuson Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005 · 2 generations on979descendants
  • K052410Acuson Sequoia Ultrasound SystemSiemens Medical Solutions USA, Inc. · 2005 · 3 generations on972descendants
  • K042770Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2004 · 2 generations on932descendants
  • K050240Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc. · 2005 · 3 generations on927descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K033196Sonoline Antares with Clarify VESiemens Medical Solutions USA, Inc.2003—
+1K041319Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2004—
+2K051139Acuson Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2005—
+3K052410Acuson Sequoia Ultrasound SystemSiemens Medical Solutions USA, Inc.2005—
+2K042770Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc.20041
+3K050240Modification to Acuson CV70 Cardiovascular SystemSiemens Medical Solutions USA, Inc.20051

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200430
200550
200650
2007110
2008140
2009210
2010170
2011290
2012360
2013420
2014500
2015590
2016590
2017770
2018680
2019490
2020750
2021540
2022590
2023570
2024650
2025802
2026550
FIG. 1 — Lineage of K033196, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYN device

  1. #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (990)

See all 990

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041319Sequoia Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2004988
K043016Sonoline Orchid Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc.2004927
K063138Modification to Acuson Antares Ultrasound SystemSiemens Medical Solutions USA, Inc.2006802
K063803Acuson Antares Diagnostic Ultrasound SystemSiemens Medical Solutions USA, Inc., Ultrasound DI2007801
K060992Syngo US WorkplaceSiemens Medical Solutions200697

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 1, 2010Not classifiedZ-1853-2011Software designImage captured from the device may be erroneous.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K242670Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 9 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K242519Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 9 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K041319 (Siemens Medical Solutions USA, Inc., 2004) has gone furthest, with 988 descendants of its own.

How does its category compare?

All IYN devices

IYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYN

57in 2025

2007: 132008: 252009: 182010: 292011: 392012: 432013: 512014: 562015: 542016: 602017: 642018: 592019: 472020: 572021: 462022: 532023: 442024: 502025: 572026 (partial year): 44
2007high 64 (2017) · low 13 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYN

124in 2025

2007: 26 days median2008: 18 days median2009: 29 days median2010: 67 days median2011: 46 days median2012: 37 days median2013: 60 days median2014: 39 days median2015: 55 days median2016: 56 days median2017: 83 days median2018: 85 days median2019: 79 days median2020: 72 days median2021: 105 days median2022: 105 days median2023: 115 days median2024: 119 days median2025: 124 days median2026 (partial year): 107 days median
2007high 124 (2025) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYN

10in 2025

2007: 162008: 122009: 82010: 102011: 82012: 32013: 82014: 22015: 32016: 22018: 72020: 42021: 82022: 102023: 42024: 12025: 102026 (partial year): 2, 1 Class I
2007high 16 (2007) · low 0 (2017)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · IYN

610in 2025

2007: 32008: 22009: 92010: 172011: 182012: 202013: 182014: 482015: 1162016: 1452017: 1462018: 852019: 592020: 672021: 2132022: 1862023: 2292024: 2662025: 610
2007high 610 (2025) · low 2 (2008)2025

MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071326163
20082518122
2009182989
201029671017
20113946818
20124337320
20135160818
20145639248
201554553116
201660562145
201764830146
20185985785
20194779059
20205772467
2021461058213
20225310510186
2023441154229
2024501191266
20255712410610
2026441072 (1)—
Cite this page

510(k) Report. “K033196: Sonoline Antares with Clarify VE.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033196

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026