510(k) · Cleared

Recalled · 2011

BD Phoenix Automated Microbiology System - Confirmatory Esbl Test

K033458510(k) TraditionalCleared May 28, 2004211-day reviewClass II21 CFR 866.1645Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Becton, Dickinson & CO · LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Cleared on May 28, 2004 as substantially equivalent to K031545 (Trek Diagnostic Systems, Inc., 2003) and 3 other predicates after a 211-day review. Its family tree reaches back 1 generation to 2002. 1 later device descends from it. It has been recalled once.

Decided
May 28, 2004
Review
211 days
Class
Class II
Product code
LON
Recalls
1
Predicates
4

Its longest line of descent runs back 1 generation to K020322 (2002). 4 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2002: 2 other ancestors2018: 1 descendant clearedK020322 · BD Phoenix Automated Microbiology System · Becton, Dickinson & CO · 2002200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 1 descendant K020322 · BD Phoenix Automated Microbiology System · Becton, Dickinson & CO · 20020
  1. Root · 1 generation backK020322 BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO · 2002
  2. This deviceK033458 BD Phoenix Automated Microbiology System - Confirmatory Esbl TestBecton, Dickinson & CO · 2004
  3. Descendant · cites it directly · 0 of its ownK181665 BD Phoenix Automated Microbiology System - BD Phoenix CPO detect - GNBecton, Dickinson and Company · 2018 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K181665BD Phoenix Automated Microbiology System - BD Phoenix CPO detect - GNBecton, Dickinson and Company · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K020322BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO2002—
0 (this device)K033458BD Phoenix Automated Microbiology System - Confirmatory Esbl TestBecton, Dickinson & CO2004—
+1K181665BD Phoenix Automated Microbiology System - BD Phoenix CPO detect - GNBecton, Dickinson and Company20182

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
FIG. 1 — Lineage of K033458, 2002–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LON? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LON device

  1. #1 K260619 VITEK 2 AST-Streptococcus Cefepime (<=0.06 –…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031545Sensititre 18-24 Hour Susceptibility Plate
Summary page 1
…Antimicrobial Susceptibility Device, 21 CFR 866.1645 PREDICATE DEVICES: Sensititre® ESBL confirmatory test plate (K03 1545) and BD PhoeniXTm Automated Microbiology System with…
Trek Diagnostic Systems, Inc.2003—
*K020321BD Phoenix Automated Microbiology System
Summary page 1
…st plate (K03 1545) and BD PhoeniXTm Automated Microbiology System with Gatifloxacin (K02032 1, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322,…
Becton, Dickinson & CO20024
*K020323BD Phoenix Automated Microbiology System
Summary page 1
…Automated Microbiology System with Gatifloxacin (K02032 1, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). INTENDED USE:…
Becton, Dickinson & CO2002—
*K020322BD Phoenix Automated Microbiology System
Summary page 1
…Gatifloxacin (K02032 1, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). INTENDED USE: The BD Phoenix Tm Automate…
Becton, Dickinson & CO2002—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181665BD Phoenix Automated Microbiology System - BD Phoenix CPO detect - GNBecton, Dickinson and Company20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 28, 2011Not classifiedZ-1847-2011Mixed-up of materials/componentsIn vitro diagnostic test kit is defective and may cause false positive or false negative results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K181665 (Becton, Dickinson and Company, 2018) has gone furthest, with 0 descendants of its own.

How does its category compare?

All LON devices

LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LON

5in 2025

2007: 192008: 72009: 82010: 12011: 82012: 102013: 72014: 42016: 32017: 102018: 42019: 82020: 52021: 52022: 82023: 112024: 22025: 52026 (partial year): 4
2007high 19 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LON

90in 2025

2007: 50 days median2008: 66 days median2009: 76 days median2011: 85 days median2012: 200 days median2013: 115 days median2014: 90 days median2016: 242 days median2017: 89 days median2018: 88 days median2019: 84 days median2020: 81 days median2021: 268 days median2022: 189 days median2023: 210 days median2025: 90 days median2026 (partial year): 146 days median
2007high 268 (2021) · low 50 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LON

1in 2025

2008: 12010: 1162011: 32012: 32014: 2, 2 Class I2016: 22017: 12018: 62019: 12020: 82022: 152024: 12025: 12026 (partial year): 6
2007high 116 (2010) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LON

690in 2025

2007: 32008: 12010: 32011: 42012: 92013: 32014: 102015: 412016: 942017: 1972018: 5122019: 3152020: 2182021: 2322022: 1802023: 4532024: 4182025: 690
2007high 690 (2025) · low 0 (2009)2025

MAUDE reports naming product code LON, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007195003
200876611
200987600
20101—1163
201188534
20121020039
2013711503
20144902 (2)10
20150—041
20163242294
201710891197
20184886512
20198841315
20205818218
202152680232
2022818915180
2023112100453
20242—1418
20255901690
202641466—

What is it indicated for?

The BD PhoenixTM Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus. This premarket notification…

Read the full statement (102 words)

The BD PhoenixTM Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

This premarket notification is for the addition of the confirmatory ESBL Test to Gram-negative ID/AST or AST only Phoenix panels. The Phoenix confirmatory ESBL Test is a confirmatory test for the detection of the organisms that produce extended spectrum P3-lacta ..es (ESBL) in Escherichia coli, Klebsiella pneumoniae, and Klebsiella oxytoca.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033458: BD Phoenix Automated Microbiology System - Confirmatory Esbl Test.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033458

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026