510(k) · Cleared
Recalled · 2009Zoll Aed Plus
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Zoll Medical Corp · MKJ Automated External Defibrillators (Non-Wearable)
Cleared on May 21, 2004 as substantially equivalent to K022732 (Medtronic Physio-Control Corp., 2002) after a 200-day review. Its family tree reaches back 2 generations to 2001. 1 of the 75 devices in its family, counting itself, has a Class I recall. 72 later devices descend from it. It has been recalled once.
- Decided
- May 21, 2004
- Review
- 200 days
- Class
- Class III
- Product code
- MKJ
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 2 generations to K003819 (2001). 72 later devices descend from it, 5 citing it directly. 1 of the 75 devices in this family, counting this one, has a Class I recall.
- Root · 2 generations backK003819 Heartstream Attenuated Defibrillation Pads, Model M3870AAgilent Technologies, Inc. · 2001 · Recalled
- 1 generation backK022732 Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp. · 2002 · Recalled (Class I)
- This deviceK033474 Zoll Aed PlusZoll Medical Corp · 2004
- Descendant · cites it directly · 64 of its ownK042007 Zoll E Series DefibrillatorZoll Medical Corp · 2005 · Recalled
- Descendant · 2 generations on · 27 of its ownK090989 Zoll R Series with Nibp and ETCO2 Options, Zoll R SeriesZoll Medical Corp · 2009
- Descendant · 3 generations on · 24 of its ownK151644 Masimo Root Vital Signs Monitoring System and AccessoriesMasimo Corporation · 2016
- Descendant · cites it directly · 23 of its ownK041892 Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005
- Descendant · 2 generations on · 21 of its ownK060559 Zoll R Series DefibrillatorZOLL Medical Corporation · 2006 · Recalled
Showing 5 of 72 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K022732 Infant/Child Reduced Energy Electrodesdirect predicate · 2002
Most influential descendants
Showing 5 of 72 · See all- K042007Zoll E Series DefibrillatorZoll Medical Corp · 2005 · cites it directly64descendants
- K090989Zoll R Series with Nibp and ETCO2 Options, Zoll R SeriesZoll Medical Corp · 2009 · 2 generations on27descendants
- K151644Masimo Root Vital Signs Monitoring System and AccessoriesMasimo Corporation · 2016 · 3 generations on24descendants
- K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005 · cites it directly23descendants
- K060559Zoll R Series DefibrillatorZOLL Medical Corporation · 2006 · 2 generations on21descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K003819 | Heartstream Attenuated Defibrillation Pads, Model M3870A | Agilent Technologies, Inc. | 2001 | 4 |
| −1 | K022732 | Infant/Child Reduced Energy Electrodes | Medtronic Physio-Control Corp. | 2002 | Class I |
| 0 (this device) | K033474 | Zoll Aed Plus | Zoll Medical Corp | 2004 | — |
| +1 | K042007 | Zoll E Series Defibrillator | Zoll Medical Corp | 2005 | 3 |
| +2 | K090989 | Zoll R Series with Nibp and ETCO2 Options, Zoll R Series | Zoll Medical Corp | 2009 | — |
| +3 | K151644 | Masimo Root Vital Signs Monitoring System and Accessories | Masimo Corporation | 2016 | — |
| +1 | K041892 | Zoll Aed Pro External Defibrillator | ZOLL Medical Corporation | 2005 | — |
| +2 | K060559 | Zoll R Series Defibrillator | ZOLL Medical Corporation | 2006 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 2 | 0 |
| 2006 | 1 | 0 |
| 2008 | 4 | 0 |
| 2009 | 5 | 0 |
| 2010 | 4 | 0 |
| 2011 | 8 | 0 |
| 2012 | 6 | 0 |
| 2013 | 1 | 0 |
| 2014 | 6 | 0 |
| 2015 | 5 | 0 |
| 2016 | 1 | 0 |
| 2017 | 3 | 0 |
| 2018 | 2 | 0 |
| 2019 | 3 | 0 |
| 2020 | 6 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 4 | 0 |
| 2024 | 5 | 0 |
| 2025 | 1 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K022732 | Infant/Child Reduced Energy ElectrodesSummary page 2…Medtronic Physio-Control Infant / Child Reduced Energy Electrodes used with the LIFEPAK CR Plus or biphasic LIFEPAK 500 AEDs cleared under premarket 510(k) notification k022732. Performance Data The ZOLL AED Plus external defibrillator was subject to extensive safety and performance testing prior to rel… | 2002 |
What descends from it? (72)
See all 725 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 12, 2009 | Not classified | Z-1206-2009 | Process design | Device fails to discharge the defibrillation energy. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K022732Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp. · 2002 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
What does this lineage say?
- The line of descent spans 3 years and passes through 3 companies; a new generation was cleared every 1.6 years on median.
- 2 of the 2 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K042007 (Zoll Medical Corp, 2005) has gone furthest, with 64 descendants of its own.
- 2 other devices also cite its predicate K022732.
How does its category compare?
All MKJ devicesMKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MKJ
48in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MKJ
156in 2015
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MKJ
1in 2025
Class I (23 since 2007) Class II/III not classified
Adverse event reports per year · MKJ
13,315in 2025
MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 69 | 12 | 3,958 |
| 2008 | 13 | 141 | 24 | 4,752 |
| 2009 | 8 | 165 | 37 | 6,782 |
| 2010 | 4 | 250 | 35 | 7,399 |
| 2011 | 11 | 36 | 7 | 8,170 |
| 2012 | 10 | 197 | 5 (1) | 9,732 |
| 2013 | 7 | 191 | 4 | 14,955 |
| 2014 | 11 | 196 | 24 (15) | 15,399 |
| 2015 | 6 | 156 | 1 | 16,129 |
| 2016 | 0 | — | 2 (1) | 17,062 |
| 2017 | 4 | — | 6 (4) | 16,684 |
| 2018 | 20 | — | 3 | 16,591 |
| 2019 | 25 | — | 6 (2) | 16,861 |
| 2020 | 52 | — | 2 | 16,197 |
| 2021 | 41 | — | 4 | 16,300 |
| 2022 | 19 | — | 3 | 14,867 |
| 2023 | 24 | — | 11 | 13,766 |
| 2024 | 27 | — | 5 | 15,109 |
| 2025 | 48 | — | 1 | 13,315 |
| 2026 | 31 | — | 3 | — |
What is it indicated for?
Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION as indicated by:
Unconsciousness and
Absence of normal breathing and
Absence of a pulse or signs of circulation.
When a victim is less than 8 years of age, or weighs less than 55 lbs (25kg), the ZOLL AED Plus should be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
The ZOLL AED Plus external defibrillator is intended to be used by personnel who are qualified by training in basic life support, advanced life support, or other physician-authorized emergency medical response to defibrillate victims of cardiac arrest.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K033474 · printed Nov 3, 2026
Cite this page
510(k) Report. “K033474: Zoll Aed Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K033474
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026