Zoll Aed Plus

K033474510(k) TraditionalCleared May 21, 2004200-day reviewClass III21 CFR 870.5310Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zoll Medical Corp · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on May 21, 2004 as substantially equivalent to K022732 (Medtronic Physio-Control Corp., 2002) after a 200-day review. Its family tree reaches back 2 generations to 2001. 1 of the 75 devices in its family, counting itself, has a Class I recall. 72 later devices descend from it. It has been recalled once.

Decided
May 21, 2004
Review
200 days
Class
Class III
Product code
MKJ
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K003819 (2001). 72 later devices descend from it, 5 citing it directly. 1 of the 75 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2011: 8 descendants cleared2010: 4 descendants cleared2005: 2 descendants cleared2006: 1 descendant cleared2009: 5 descendants cleared2008: 4 descendants cleared2012: 6 descendants cleared2013: 1 descendant cleared2015: 5 descendants cleared2016: 1 descendant cleared2014: 6 descendants cleared2017: 3 descendants cleared2020: 6 descendants cleared2018: 2 descendants cleared2021: 2 descendants cleared2023: 4 descendants cleared2019: 3 descendants cleared2022: 1 descendant cleared2024: 5 descendants cleared2026: 2 descendants cleared2025: 1 descendant clearedK003819 · Heartstream Attenuated Defibrillation Pads, Model M3870A · Agilent Technologies, Inc. · 2001 · recalledK022732 · Infant/Child Reduced Energy Electrodes · Medtronic Physio-Control Corp. · 2002 · Class I recall200020052010201520202025
Descendants, by generation 1 generation forward: 5 descendants2 generations forward: 12 descendants3 generations forward: 16 descendants4 generations forward: 14 descendants5 generations forward: 8 descendants6 generations forward: 14 descendants7 generations forward: 3 descendants K022732 · Infant/Child Reduced Energy Electrodes · Medtronic Physio-Control Corp. · 2002 · Class I recallK003819 · Heartstream Attenuated Defibrillation Pads, Model M3870A · Agilent Technologies, Inc. · 2001 · recalled0+5
  1. Root · 2 generations backK003819 Heartstream Attenuated Defibrillation Pads, Model M3870AAgilent Technologies, Inc. · 2001 · Recalled
  2. 1 generation backK022732 Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp. · 2002 · Recalled (Class I)
  3. This deviceK033474 Zoll Aed PlusZoll Medical Corp · 2004
  4. Descendant · cites it directly · 64 of its ownK042007 Zoll E Series DefibrillatorZoll Medical Corp · 2005 · Recalled
  5. Descendant · 2 generations on · 27 of its ownK090989 Zoll R Series with Nibp and ETCO2 Options, Zoll R SeriesZoll Medical Corp · 2009
  6. Descendant · 3 generations on · 24 of its ownK151644 Masimo Root Vital Signs Monitoring System and AccessoriesMasimo Corporation · 2016
  7. Descendant · cites it directly · 23 of its ownK041892 Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005
  8. Descendant · 2 generations on · 21 of its ownK060559 Zoll R Series DefibrillatorZOLL Medical Corporation · 2006 · Recalled

Showing 5 of 72 descendants. See all descendants

Descendants cleared per year

2005: 22006: 12008: 42009: 52010: 42011: 82012: 62013: 12014: 62015: 52016: 12017: 32018: 22019: 32020: 62021: 22022: 12023: 42024: 52025: 12026: 2
2005high 8 (2011) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 72 · See all
  • K042007Zoll E Series DefibrillatorZoll Medical Corp · 2005 · cites it directly64descendants
  • K090989Zoll R Series with Nibp and ETCO2 Options, Zoll R SeriesZoll Medical Corp · 2009 · 2 generations on27descendants
  • K151644Masimo Root Vital Signs Monitoring System and AccessoriesMasimo Corporation · 2016 · 3 generations on24descendants
  • K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005 · cites it directly23descendants
  • K060559Zoll R Series DefibrillatorZOLL Medical Corporation · 2006 · 2 generations on21descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K003819Heartstream Attenuated Defibrillation Pads, Model M3870AAgilent Technologies, Inc.20014
−1K022732Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp.2002Class I
0 (this device)K033474Zoll Aed PlusZoll Medical Corp2004—
+1K042007Zoll E Series DefibrillatorZoll Medical Corp20053
+2K090989Zoll R Series with Nibp and ETCO2 Options, Zoll R SeriesZoll Medical Corp2009—
+3K151644Masimo Root Vital Signs Monitoring System and AccessoriesMasimo Corporation2016—
+1K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation2005—
+2K060559Zoll R Series DefibrillatorZOLL Medical Corporation20061

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200520
200610
200840
200950
201040
201180
201260
201310
201460
201550
201610
201730
201820
201930
202060
202120
202210
202340
202450
202510
202620
FIG. 1 — Lineage of K033474, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022732Infant/Child Reduced Energy Electrodes
Summary page 2
…Medtronic Physio-Control Infant / Child Reduced Energy Electrodes used with the LIFEPAK CR Plus or biphasic LIFEPAK 500 AEDs cleared under premarket 510(k) notification k022732. Performance Data The ZOLL AED Plus external defibrillator was subject to extensive safety and performance testing prior to rel…
Medtronic Physio-Control Corp.2002Class I

Also cleared against K022732: 2 other devices →

What descends from it? (72)

See all 72

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042007Zoll E Series DefibrillatorZoll Medical Corp200564
K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation200523
K110978Onestep Pediatric Mfe'sZOLL Medical Corporation20117
K110154Zoll Aed Plus with 2010 Aha Guidlines Software UpdateZoll Medical Corporation, World Wide Headquarters20112
K091561Zoll Aed Plus Version 5.32 Software ReleaseZOLL Medical Corporation20100

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 12, 2009Not classifiedZ-1206-2009Process designDevice fails to discharge the defibrillation energy.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K022732Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp. · 2002 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 3 years and passes through 3 companies; a new generation was cleared every 1.6 years on median.
  • 2 of the 2 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K042007 (Zoll Medical Corp, 2005) has gone furthest, with 64 descendants of its own.
  • 2 other devices also cite its predicate K022732.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

Use the AED when a suspected cardiac arrest victim has an apparent LACK OF CIRCULATION as indicated by:

Unconsciousness and

Absence of normal breathing and

Absence of a pulse or signs of circulation.

When a victim is less than 8 years of age, or weighs less than 55 lbs (25kg), the ZOLL AED Plus should be used with the ZOLL AED Plus Pediatric Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.

The ZOLL AED Plus external defibrillator is intended to be used by personnel who are qualified by training in basic life support, advanced life support, or other physician-authorized emergency medical response to defibrillate victims of cardiac arrest.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033474: Zoll Aed Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K033474

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026