510(k) · Cleared
Recalled · Class I · 2020Infant/Child Reduced Energy Electrodes
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medtronic Physio-Control Corp. · MKJ Automated External Defibrillators (Non-Wearable)
Cleared on Dec 13, 2002 as substantially equivalent to K003819 (Agilent Technologies, Inc., 2001) after a 119-day review. Its family tree reaches back 1 generation to 2001. 1 of the 77 devices in its family, counting itself, has a Class I recall. 75 later devices descend from it. It has been recalled 4 times, including 2 Class I recalls.
- Decided
- Dec 13, 2002
- Review
- 119 days
- Class
- Class III
- Product code
- MKJ
- Recalls
- 4 (2 Class I)
- Predicates
- 2
Its longest line of descent runs back 1 generation to K003819 (2001). 75 later devices descend from it, 3 citing it directly. 1 of the 77 devices in this family, counting this one, has a Class I recall.
- Root · 1 generation backK003819 Heartstream Attenuated Defibrillation Pads, Model M3870AAgilent Technologies, Inc. · 2001 · Recalled
- This deviceK022732 Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp. · 2002
- Descendant · cites it directly · 72 of its ownK033474 Zoll Aed PlusZoll Medical Corp · 2004 · Recalled
- Descendant · 2 generations on · 64 of its ownK042007 Zoll E Series DefibrillatorZoll Medical Corp · 2005 · Recalled
- Descendant · 3 generations on · 27 of its ownK090989 Zoll R Series with Nibp and ETCO2 Options, Zoll R SeriesZoll Medical Corp · 2009
- Descendant · 4 generations on · 24 of its ownK151644 Masimo Root Vital Signs Monitoring System and AccessoriesMasimo Corporation · 2016
- Descendant · 2 generations on · 23 of its ownK041892 Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005
Showing 5 of 75 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 75 · See all- K033474Zoll Aed PlusZoll Medical Corp · 2004 · cites it directly72descendants
- K042007Zoll E Series DefibrillatorZoll Medical Corp · 2005 · 2 generations on64descendants
- K090989Zoll R Series with Nibp and ETCO2 Options, Zoll R SeriesZoll Medical Corp · 2009 · 3 generations on27descendants
- K151644Masimo Root Vital Signs Monitoring System and AccessoriesMasimo Corporation · 2016 · 4 generations on24descendants
- K041892Zoll Aed Pro External DefibrillatorZOLL Medical Corporation · 2005 · 2 generations on23descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K003819 | Heartstream Attenuated Defibrillation Pads, Model M3870A | Agilent Technologies, Inc. | 2001 | 4 |
| 0 (this device) | K022732 | Infant/Child Reduced Energy Electrodes | Medtronic Physio-Control Corp. | 2002 | — |
| +1 | K033474 | Zoll Aed Plus | Zoll Medical Corp | 2004 | 1 |
| +2 | K042007 | Zoll E Series Defibrillator | Zoll Medical Corp | 2005 | 3 |
| +3 | K090989 | Zoll R Series with Nibp and ETCO2 Options, Zoll R Series | Zoll Medical Corp | 2009 | — |
| +4 | K151644 | Masimo Root Vital Signs Monitoring System and Accessories | Masimo Corporation | 2016 | — |
| +2 | K041892 | Zoll Aed Pro External Defibrillator | ZOLL Medical Corporation | 2005 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 2 | 0 |
| 2005 | 2 | 0 |
| 2006 | 1 | 0 |
| 2008 | 4 | 0 |
| 2009 | 5 | 0 |
| 2010 | 4 | 0 |
| 2011 | 8 | 0 |
| 2012 | 7 | 0 |
| 2013 | 1 | 0 |
| 2014 | 6 | 0 |
| 2015 | 5 | 0 |
| 2016 | 1 | 0 |
| 2017 | 3 | 0 |
| 2018 | 2 | 0 |
| 2019 | 3 | 0 |
| 2020 | 6 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 4 | 0 |
| 2024 | 5 | 0 |
| 2025 | 1 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.
| Type | Number | Device | Decided |
|---|---|---|---|
| Kept off the family tree | K979301 | Medtronic Physio-Control Pediatric QUIK-COMBO electrodes Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree Summary page 2…e Infant / Child Reduced Energy Electrodes are substantially equivalent to those of the Medtronic Physio- Control Pediatric QUIK-COMBO electrodes, 51O(k) no. K979301, (cleared March 6, 1997) and the Heartstream Attenuated Defibrillation Pads, 510(k) no. KO03819 (cleared May 2, 2001… | |
| * | K003819 | Heartstream Attenuated Defibrillation Pads, Model M3870A | 2001 |
What descends from it? (75)
See all 753 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 24, 2020 | Class II | Z-2360-2020 | Packaging change control | Packaging of infant child reduced energy electrodes was not properly sealed. |
| Apr 24, 2020 | Class II | Z-2359-2020 | Packaging change control | Packaging of infant child reduced energy electrodes was not properly sealed. |
| Oct 27, 2017 | Class I | Z-0186-2018 | Process control | The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols. |
| Oct 27, 2017 | Class I | Z-0187-2018 | Process control | The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All MKJ devicesMKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MKJ
48in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MKJ
156in 2015
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MKJ
1in 2025
Class I (23 since 2007) Class II/III not classified
Adverse event reports per year · MKJ
13,315in 2025
MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 69 | 12 | 3,958 |
| 2008 | 13 | 141 | 24 | 4,752 |
| 2009 | 8 | 165 | 37 | 6,782 |
| 2010 | 4 | 250 | 35 | 7,399 |
| 2011 | 11 | 36 | 7 | 8,170 |
| 2012 | 10 | 197 | 5 (1) | 9,732 |
| 2013 | 7 | 191 | 4 | 14,955 |
| 2014 | 11 | 196 | 24 (15) | 15,399 |
| 2015 | 6 | 156 | 1 | 16,129 |
| 2016 | 0 | — | 2 (1) | 17,062 |
| 2017 | 4 | — | 6 (4) | 16,684 |
| 2018 | 20 | — | 3 | 16,591 |
| 2019 | 25 | — | 6 (2) | 16,861 |
| 2020 | 52 | — | 2 | 16,197 |
| 2021 | 41 | — | 4 | 16,300 |
| 2022 | 19 | — | 3 | 14,867 |
| 2023 | 24 | — | 11 | 13,766 |
| 2024 | 27 | — | 5 | 15,109 |
| 2025 | 48 | — | 1 | 13,315 |
| 2026 | 31 | — | 3 | — |
What is it indicated for?
Use the Infant/Child defibrillation electrodes on an infant or child up to 8 years old or up to 25kg (551bs). Do not delay therapy to determine child’s exact age or weight. These electrodes are to be used only with LIFEPAK CR PLUS defibrillators and biphasic LIFEPAK 500 AEDs configured with a pink cable connector. The LIFEPAK CR PLUS and LIFEPAK 500 biphasic defibrillators are indicated for use on patients in cardiopulmonary arrest. The patient must be…
Read the full statement (143 words)
Use the Infant/Child defibrillation electrodes on an infant or child up to 8 years old or up to 25kg (551bs). Do not delay therapy to determine child’s exact age or weight.
These electrodes are to be used only with LIFEPAK CR PLUS defibrillators and biphasic LIFEPAK 500 AEDs configured with a pink cable connector.
The LIFEPAK CR PLUS and LIFEPAK 500 biphasic defibrillators are indicated for use on patients in cardiopulmonary arrest. The patient must be unconscious, not breathing normally, and showing no signs of circulation (for example, no pulse, and/or no coughing, no movement) before the device is used to analyze the patient’s ECG rhythm.
The LIFEPAK CR PLUS and LIFEPAK 500 biphasic defibrillators are intended to be used by personnel who have been trained on the device operation and in basic life support or other physician authorized emergency medical response system.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K022732 · printed Nov 5, 2026
Cite this page
510(k) Report. “K022732: Infant/Child Reduced Energy Electrodes.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K022732