510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Hydrophilic-Antibacterial Intermittent Catheter, Intermittent Catheter Closed System Kit

K033477510(k) TraditionalCleared Jun 10, 2004220-day reviewClass II21 CFR 876.5130Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Rochester Medical Corp. · MJC Catheter, Urological (Antimicrobial) And Accessories

Cleared on Jun 10, 2004 as substantially equivalent to K001143 (Rochester Medical Corp., 2000) and 4 other predicates after a 220-day review. Its family tree reaches back 2 generations to 1998. It has been recalled 3 times.

Decided
Jun 10, 2004
Review
220 days
Class
Class II
Product code
MJC
Recalls
3
Predicates
5

Its longest line of descent runs back 2 generations to K973120 (1998). 7 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2001: 3 other ancestors2000: 2 other ancestorsK973120 · Readycath Catheter · Medical Technologies of Georgia · 1998K013483 · Incare Advance; Intermittent Catheter, Plus Intermittent Catheter, Plus Kit · Hollister, Inc. · 2002199619982000200220042006
Other ancestors, by generation 1 generation back: 4 other ancestors2 generations back: 1 other ancestor K013483 · Incare Advance; Intermittent Catheter, Plus Intermittent Catheter, Plus Kit · Hollister, Inc. · 2002K973120 · Readycath Catheter · Medical Technologies of Georgia · 19980
  1. Root · 2 generations backK973120 Readycath CatheterMedical Technologies of Georgia · 1998
  2. 1 generation backK013483 Incare Advance; Intermittent Catheter, Plus Intermittent Catheter…Hollister, Inc. · 2002
  3. This deviceK033477 Hydrophilic-Antibacterial Intermittent Catheter, Intermittent…Rochester Medical Corp. · 2004
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K973120Readycath CatheterMedical Technologies of Georgia1998—
−1K013483Incare Advance; Intermittent Catheter, Plus Intermittent Catheter, Plus KitHollister, Inc.2002—
0 (this device)K033477Hydrophilic-Antibacterial Intermittent Catheter, Intermittent Catheter Closed…Rochester Medical Corp.2004—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200130
FIG. 1 — Lineage of K033477, 1998–2004Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MJC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MJC device

  1. #1 K233094 Wellead® Latex Foley Catheterscore 11.1 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001143Release NF Antibacterial Foley Catheter, Antibacterial Personal CatheterRochester Medical Corp.2000—
*K000723Hydrophilic Silicone Foley Catheter and Personal Catheter (Hydrophilic and…
Summary page 1
…Closed System Kit Predicate Devices: Rochester Medical Antibacterial Personal®Catheter KOO1 143 Rochester Medical Hydrophilic Personal® Catheter K000723 Mentor Self-Cath Closed System K003873 Hollister InCare Advance Plus Kit K013483 Rusch MMG/O'Neil Cathete…
Rochester Medical Corp.200026
*K003873Mentor Self-Cath Closed System
Summary page 1
…Rochester Medical Antibacterial Personal®Catheter KOO1 143 Rochester Medical Hydrophilic Personal® Catheter K000723 Mentor Self-Cath Closed System K003873 Hollister InCare Advance Plus Kit K013483 Rusch MMG/O'Neil Catheter K010420 Intended Use of the Device…
Mentor Corp.2001—
*K013483Incare Advance; Intermittent Catheter, Plus Intermittent Catheter, Plus Kit
Summary page 1
…theter KOO1 143 Rochester Medical Hydrophilic Personal® Catheter K000723 Mentor Self-Cath Closed System K003873 Hollister InCare Advance Plus Kit K013483 Rusch MMG/O'Neil Catheter K010420 Intended Use of the Device Hydrophilic-Antibacterial Intermittent C…
Hollister, Inc.2002—
*K010420Rusch MMG/O'neil Catheter
Summary page 1
…Hydrophilic Personal® Catheter K000723 Mentor Self-Cath Closed System K003873 Hollister InCare Advance Plus Kit K013483 Rusch MMG/O'Neil Catheter K010420 Intended Use of the Device Hydrophilic-Antibacterial Intermittent Catheter…
Rusch, Inc.20013

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 28, 2017Class IIZ-3226-2017Employee errorSpecific lots labeled as an Antibacterial Hydrophilic Intermittent Catheter may lack the hydrophilic coating.
Oct 5, 2016Class IIZ-0600-2017Mixed-up of materials/componentsMisbranding; the product labeled as an Antibacterial Hydrophilic Intermittent Catheter, is instead an Antibacterial Intermittent Catheter (no Hydrophilic Coating).
Oct 2, 2015Class IIIZ-0198-2016Error in labelingIncorrect unit labeling. The case label and shelf box are correct, Product Code 63512 and Catheter Size 12 French. The unit label (catheter package) may state Product Code 63516 and Catheter Size 16 French.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 3.9 years on median.
  • 2 of the 7 ancestors have a recall of any class.

How does its category compare?

All MJC devices

MJC Catheter, Urological (Antimicrobial) And Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MJC

1in 2024

2024: 1
2007high 1 (2024) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MJC

Too few MJC 510(k)s in any year since 2007 to give a median.

Recalls per year · MJC

2in 2018

2015: 12016: 12018: 22026 (partial year): 1
2007high 2 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MJC

3,747in 2025

2009: 22010: 12012: 52013: 82014: 132015: 252016: 1342017: 7142018: 6132019: 1,2702020: 1,4332021: 1,4642022: 1,7962023: 1,9872024: 1,9602025: 3,747
2007high 3,747 (2025) · low 0 (2007)2025

MAUDE reports naming product code MJC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—02
20100—01
20110—00
20120—05
20130—08
20140—013
20150—125
20160—1134
20170—0714
20180—2613
20190—01,270
20200—01,433
20210—01,464
20220—01,796
20230—01,987
20241—01,960
20250—03,747
20260—1—

What is it indicated for?

Urethral catheter for urological use only. The Rochester Medical Corporation Hydrophilic-Antibacterial Intermittent Catheter is intended for urinary bladder drainage in adult males and females requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures. Efficacy of the Hydrophilic-Antibacterial Intermittent Catheter in preventing urinary tract infection during intermittent use has not been established. The device is not intended to be used as a treatment for active urinary tract infection.

For urological use only. The Intermittent Catheter Closed System Kit is intended for use by patients for the purpose of bladder drainage.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033477: Hydrophilic-Antibacterial Intermittent Catheter, Intermittent Catheter Closed System Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033477

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026