510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Modification to Dash 3000/4000 Patient Monitor

K040458510(k) SpecialCleared Jun 3, 2004101-day reviewClass II21 CFR 870.1025Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems Information Technologies · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Jun 3, 2004 as substantially equivalent to K033304 (Ge Medical Systems Information Technologies, 2003) after a 101-day review. Its family tree reaches back 4 generations to 2000. It has been recalled once.

Decided
Jun 3, 2004
Review
101 days
Class
Class II
Product code
MHX
Recalls
1
Predicates
1

Its longest line of descent runs back 4 generations to K001359 (2000). No later device cites it as a predicate yet.

  1. Root · 4 generations backK001359 Dash 3000/4000 Patient MonitorGeneral Electric Medical Systems Information Techn · 2000 · Recalled
  2. 3 generations backK020290 Dash 3000/4000 Patient MonitorGeneral Electric Medical Systems Information Techn · 2002 · Recalled
  3. 2 generations backK030431 Modification to Dash 3000/4000 Patient MonitorGe Medical Systems Information Technologies · 2003 · Recalled
  4. 1 generation backK033304 Modification to Dash 3000/4000 Patient MonitorGe Medical Systems Information Technologies · 2003 · Recalled
  5. This deviceK040458 Modification to Dash 3000/4000 Patient MonitorGe Medical Systems Information Technologies · 2004
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K001359Dash 3000/4000 Patient MonitorGeneral Electric Medical Systems Information Techn20001
−3K020290Dash 3000/4000 Patient MonitorGeneral Electric Medical Systems Information Techn20021
−2K030431Modification to Dash 3000/4000 Patient MonitorGe Medical Systems Information Technologies20031
−1K033304Modification to Dash 3000/4000 Patient MonitorGe Medical Systems Information Technologies20031
0 (this device)K040458Modification to Dash 3000/4000 Patient MonitorGe Medical Systems Information Technologies2004—
FIG. 1 — Lineage of K040458, 2000–2004Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033304Modification to Dash 3000/4000 Patient Monitor
Summary page 1
…ythmia Detection Or MH-X Alarms) 21 CFR 870.2770 Plethysmograph, Impedance DSB Predicate Device: K033304 DASH 3000/4000 Patient Monitor Device Description: The DASH 3000/4000 Patient Monitor is a device that is…
Ge Medical Systems Information Technologies20031

Also cleared against K033304: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 31, 2018Class IIZ-0520-2019Software designThe patient monitors may simultaneously restart as designed if all are connected to the same network and a network overload occurs for a prolonged time.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 2 companies; a new generation was cleared every 9 months on median.
  • 4 of the 4 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K033304.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The DASH 3000/4000 Patient Monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash is designed as a bedside, portable, and transport monitor that can operate in all professional medical facilities and medical transport modes including but not limited to: emergency department, operating room…

Read the full statement (221 words)

The DASH 3000/4000 Patient Monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash is designed as a bedside, portable, and transport monitor that can operate in all professional medical facilities and medical transport modes including but not limited to: emergency department, operating room, post anesthesia recovery, critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care areas located in hospitals, outpatient clinics, freestanding surgical centers, and other alternate care facilities, intra-hospital patient transport, inter-hospital patient transport via ground vehicles (i.e., ambulance, etc.) and fixed and rotary winged aircraft, and pre-hospital emergency response.

Physiologic data includes but is not restricted to: electrocardiogram, invasive blood pressure, noninvasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, oxygen, and anesthetic agents as summarized in the operator's manual.

The DASH 3000/4000 Patient Monitor is also intended to provide physiologic data over the Unity network to clinical information systems and allow the user to access hospital data at the point-of-care.

This information can be displayed, trended, stored, and printed.

The DASH 3000/4000 Patient Monitor was developed to interface with non-proprietary third party peripheral devices that support serial data outputs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040458: Modification to Dash 3000/4000 Patient Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K040458

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026