510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Periarticular Locking Plates and Screws, 2358, 2359 Series

K040593510(k) TraditionalCleared Apr 12, 200435-day reviewClass II21 CFR 888.3030Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · HRS Plate, Fixation, Bone

Cleared on Apr 12, 2004 as substantially equivalent to K012114 (Synthes (Usa), 2001) and 1 other predicate after a 35-day review. Its family tree reaches back 2 generations to 1996. 162 later devices descend from it. It has been recalled 2 times.

Decided
Apr 12, 2004
Review
35 days
Class
Class II
Product code
HRS
Recalls
2
Predicates
2

Its longest line of descent runs back 2 generations to K962616 (1996). 4 ancestors in all. 162 later devices descend from it, 14 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1996: 1 other ancestor2004: 2 descendants cleared2006: 1 descendant cleared2008: 2 descendants cleared2013: 10 descendants cleared2015: 6 descendants cleared2011: 4 descendants cleared2014: 11 descendants cleared2005: 3 descendants cleared2012: 3 descendants cleared2016: 4 descendants cleared2007: 3 descendants cleared2010: 3 descendants cleared2017: 12 descendants cleared2026: 14 descendants cleared2018: 11 descendants cleared2021: 7 descendants cleared2025: 14 descendants cleared2020: 10 descendants cleared2022: 10 descendants cleared2024: 14 descendants cleared2019: 8 descendants cleared2023: 10 descendants clearedK962616 · Synthes Dorsal Distal Radius Plate (DDRP) System · Synthes (Usa) · 1996K982732 · Synthes (Usa) Distal Radius Plate System (DRPS) · Synthes · 19981995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 14 descendants2 generations forward: 24 descendants3 generations forward: 27 descendants4 generations forward: 46 descendants5 generations forward: 30 descendants6 generations forward: 15 descendants7 generations forward: 4 descendants8 generations forward: 2 descendants K982732 · Synthes (Usa) Distal Radius Plate System (DRPS) · Synthes · 1998K962616 · Synthes Dorsal Distal Radius Plate (DDRP) System · Synthes (Usa) · 19960+5
  1. Root · 2 generations backK962616 Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa) · 1996
  2. 1 generation backK982732 Synthes (Usa) Distal Radius Plate System (DRPS)Synthes · 1998
  3. This deviceK040593 Periarticular Locking Plates and Screws, 2358, 2359 SeriesZimmer, Inc. · 2004
  4. Descendant · cites it directly · 80 of its ownK100271 Vari AX Distal Radius Line Extensions of XXL PlatesHowmedica Osteonics Corp. · 2010
  5. Descendant · 2 generations on · 56 of its ownK112455 Aiming Blocks for Variax Distal Radius PlatesHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2011
  6. Descendant · 3 generations on · 40 of its ownK113760 Variax Superior and Anterior, Lateral and Midshaft Clavicle Plates…Howmedica Osteonics Corp., Dba Stryker Orthopaedics · 2012
  7. Descendant · cites it directly · 31 of its ownK042598 Periarticular Locking Plates and Screws, 2357 and 2359 SeriesZimmer, Inc. · 2004 · Recalled
  8. Descendant · 2 generations on · 27 of its ownK050121 Zimmer Periarticular Locking Plate System, Model 2357 & 2359Zimmer, Inc. · 2005

Showing 5 of 162 descendants. See all descendants

Descendants cleared per year

2004: 22005: 32006: 12007: 32008: 22010: 32011: 42012: 32013: 102014: 112015: 62016: 42017: 122018: 112019: 82020: 102021: 72022: 102023: 102024: 142025: 142026: 14
2004high 14 (2024) · low 0 (2009)2026

Most influential descendants

Showing 5 of 162 · See all
  • K100271Vari AX Distal Radius Line Extensions of XXL PlatesHowmedica Osteonics Corp. · 2010 · cites it directly80descendants
  • K112455Aiming Blocks for Variax Distal Radius PlatesHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2011 · 2 generations on56descendants
  • K113760Variax Superior and Anterior, Lateral and Midshaft Clavicle Plates Variax Aiming Blocks…Howmedica Osteonics Corp., Dba Stryker Orthopaedics · 2012 · 3 generations on40descendants
  • K042598Periarticular Locking Plates and Screws, 2357 and 2359 SeriesZimmer, Inc. · 2004 · cites it directly31descendants
  • K050121Zimmer Periarticular Locking Plate System, Model 2357 & 2359Zimmer, Inc. · 2005 · 2 generations on27descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K962616Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa)1996—
−1K982732Synthes (Usa) Distal Radius Plate System (DRPS)Synthes1998—
0 (this device)K040593Periarticular Locking Plates and Screws, 2358, 2359 SeriesZimmer, Inc.2004—
+1K100271Vari AX Distal Radius Line Extensions of XXL PlatesHowmedica Osteonics Corp.2010—
+2K112455Aiming Blocks for Variax Distal Radius PlatesHowmedica Osteonics Corp., Dba Stryker Orthopaedics2011—
+3K113760Variax Superior and Anterior, Lateral and Midshaft Clavicle Plates Variax…Howmedica Osteonics Corp., Dba Stryker Orthopaedics2012—
+1K042598Periarticular Locking Plates and Screws, 2357 and 2359 SeriesZimmer, Inc.20041
+2K050121Zimmer Periarticular Locking Plate System, Model 2357 & 2359Zimmer, Inc.2005—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200420
200530
200610
200730
200820
201030
201140
201230
2013100
2014110
201560
201640
2017120
2018110
201980
2020100
202170
2022100
2023100
2024140
2025140
2026140
FIG. 1 — Lineage of K040593, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HRS device

  1. #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K012114Synthes Locking Distal Radius Plating System
Summary page 1
…Preamendment Device Synthes Locking Distal Radius Plating System, K012114, cleared 9-28-01 Synthes Distal Radius Plate System, K982732,…
Synthes (Usa)20013
*K982732Synthes (Usa) Distal Radius Plate System (DRPS)
Summary page 1
…ting System, K012114, cleared 9-28-01 Synthes Distal Radius Plate System, K982732, cleared 10-8-98 Device Description: The Zimmer Peri…
Synthes1998—
*—Zimmer ECT® Internal Fracture Fixation System
Named without a submission number
—

Also cleared against K012114: 15 other devices →

What descends from it? (162)

See all 162

14 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K100271Vari AX Distal Radius Line Extensions of XXL PlatesHowmedica Osteonics Corp.201080
K042598Periarticular Locking Plates and Screws, 2357 and 2359 SeriesZimmer, Inc.200431
K063672Fibula Plate Prolock with Angular StabilityI.T.S. Implantat-Technologie-Systeme GmbH200725
K080184Distal Humeral Plates with Angular StabilityI.T.S. Implantat-Technologie-Systeme GmbH200815
K051098Periarticular Locking Plates and Screws, Model 2358 SeriesZimmer, Inc.200513
K043560Periarticular Locking Plates and Screws, Series 2358 and 2359Zimmer, Inc.200513
K043227Periarticular Locking Plates, 2358 SeriesZimmer, Inc.200413
K060969Polyax Locked Plating SystemDePuy Orthopaedics, Inc.200610
K121601Synthes 2.7/3.5MM Variable Angle LCP Ankle Trauma System-Anterolateral Distal Tibia PlateSynthes USA20126
K142938G3 Active Plate Large Fragment SystemGenesis Fracture Care, Inc.20143

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 9, 2015Class IIZ-0646-2016Labeling Change ControlProduct on the package label did not match what was inside the package. The label was for part number 00-2359-058-35 (Zimmer Periarticular Locking Plate System 3.5mm Locking Screw 58mm Length) lot number 62918019 while the product inside…
Jun 10, 2014Class IIZ-2131-2014Device DesignA complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pulling through plate hole.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 5.5 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K100271 (Howmedica Osteonics Corp., 2010) has gone furthest, with 80 descendants of its own.
  • 15 other devices also cite its predicate K012114.

How does its category compare?

All HRS devices

HRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HRS

54in 2025

2007: 352008: 352009: 472010: 492011: 432012: 592013: 562014: 762015: 562016: 442017: 572018: 402019: 462020: 482021: 402022: 472023: 402024: 542025: 542026 (partial year): 51
2007high 76 (2014) · low 35 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HRS

90in 2025

2007: 63 days median2008: 87 days median2009: 85 days median2010: 100 days median2011: 88 days median2012: 106 days median2013: 89 days median2014: 57 days median2015: 97 days median2016: 109 days median2017: 122 days median2018: 116 days median2019: 89 days median2020: 88 days median2021: 87 days median2022: 144 days median2023: 120 days median2024: 96 days median2025: 90 days median2026 (partial year): 82 days median
2007high 144 (2022) · low 57 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HRS

1in 2025

2007: 22008: 112009: 32010: 162011: 52012: 42013: 112015: 72016: 52017: 32018: 22019: 252020: 12021: 92022: 32023: 12025: 12026 (partial year): 3
2007high 25 (2019) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HRS

2,306in 2025

2007: 3392008: 3112009: 4352010: 4912011: 7732012: 1,6632013: 2,6432014: 2,4132015: 3,0952016: 2,2762017: 2,2202018: 2,6632019: 3,3592020: 2,7902021: 2,4422022: 1,7002023: 2,0492024: 1,8492025: 2,306
2007high 3,359 (2019) · low 311 (2008)2025

MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735632339
2008358711311
200947853435
20104910016491
201143885773
20125910641,663
20135689112,643
2014765702,413
2015569773,095
20164410952,276
20175712232,220
20184011622,663
20194689253,359
2020488812,790
2021408792,442
20224714431,700
20234012012,049
2024549601,849
2025549012,306
202651823—

What is it indicated for?

The Periarticular Locking Plate System is indicated for temporary internal fixation and stabilization of osteotomies and fractures, including:

* Comminuted fractures

* Supracondylar fractures

* Intra-articular and extra-articular condylar fractures

* Fractures in osteopenic bone

* Nonunions

* Malunions

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040593: Periarticular Locking Plates and Screws, 2358, 2359 Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K040593

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026