510(k) · Cleared
Recalled · Class II · 2016Synthes Locking Distal Radius Plating System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Synthes (Usa) · HRS Plate, Fixation, Bone
Cleared on Sep 28, 2001 after an 84-day review. Our reading found no predicate number; beside “510(k)” the summary reads “000005”, which our checks could not confirm as its predicate. 205 later devices descend from it. It has been recalled 3 times.
- Decided
- Sep 28, 2001
- Review
- 84 days
- Class
- Class II
- Product code
- HRS
- Recalls
- 3
- Predicates
- 0
205 later devices descend from it, 16 citing it directly.
- This deviceK012114 Synthes Locking Distal Radius Plating SystemSynthes (Usa) · 2001
- Descendant · cites it directly · 162 of its ownK040593 Periarticular Locking Plates and Screws, 2358, 2359 SeriesZimmer, Inc. · 2004 · Recalled
- Descendant · 2 generations on · 80 of its ownK100271 Vari AX Distal Radius Line Extensions of XXL PlatesHowmedica Osteonics Corp. · 2010
- Descendant · 3 generations on · 56 of its ownK112455 Aiming Blocks for Variax Distal Radius PlatesHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2011
- Descendant · cites it directly · 50 of its ownK102694 2.4MM VA-LCP Dorsal Distal Radius PlatesSynthes (Usa) · 2010 · Recalled
- Descendant · 4 generations on · 40 of its ownK113760 Variax Superior and Anterior, Lateral and Midshaft Clavicle Plates…Howmedica Osteonics Corp., Dba Stryker Orthopaedics · 2012
Showing 5 of 205 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 205 · See all- K040593Periarticular Locking Plates and Screws, 2358, 2359 SeriesZimmer, Inc. · 2004 · cites it directly162descendants
- K100271Vari AX Distal Radius Line Extensions of XXL PlatesHowmedica Osteonics Corp. · 2010 · 2 generations on80descendants
- K112455Aiming Blocks for Variax Distal Radius PlatesHowmedica Osteonics Corp., Dba Stryker Orthopaedics · 2011 · 3 generations on56descendants
- K1026942.4MM VA-LCP Dorsal Distal Radius PlatesSynthes (Usa) · 2010 · cites it directly50descendants
- K113760Variax Superior and Anterior, Lateral and Midshaft Clavicle Plates Variax Aiming Blocks…Howmedica Osteonics Corp., Dba Stryker Orthopaedics · 2012 · 4 generations on40descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K012114 | Synthes Locking Distal Radius Plating System | Synthes (Usa) | 2001 | — |
| +1 | K040593 | Periarticular Locking Plates and Screws, 2358, 2359 Series | Zimmer, Inc. | 2004 | 2 |
| +2 | K100271 | Vari AX Distal Radius Line Extensions of XXL Plates | Howmedica Osteonics Corp. | 2010 | — |
| +3 | K112455 | Aiming Blocks for Variax Distal Radius Plates | Howmedica Osteonics Corp., Dba Stryker Orthopaedics | 2011 | — |
| +1 | K102694 | 2.4MM VA-LCP Dorsal Distal Radius Plates | Synthes (Usa) | 2010 | 3 |
| +4 | K113760 | Variax Superior and Anterior, Lateral and Midshaft Clavicle Plates Variax… | Howmedica Osteonics Corp., Dba Stryker Orthopaedics | 2012 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 3 | 0 |
| 2005 | 3 | 0 |
| 2006 | 2 | 0 |
| 2007 | 3 | 0 |
| 2008 | 2 | 0 |
| 2010 | 4 | 0 |
| 2011 | 5 | 0 |
| 2012 | 3 | 0 |
| 2013 | 12 | 0 |
| 2014 | 14 | 0 |
| 2015 | 7 | 0 |
| 2016 | 9 | 0 |
| 2017 | 17 | 0 |
| 2018 | 12 | 0 |
| 2019 | 11 | 0 |
| 2020 | 11 | 0 |
| 2021 | 10 | 0 |
| 2022 | 13 | 0 |
| 2023 | 11 | 0 |
| 2024 | 18 | 0 |
| 2025 | 16 | 0 |
| 2026 | 19 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HRS device
- #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Synthes Distal Radius Plate System Named without a submission number |
What descends from it? (205)
See all 20516 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K040593 | Periarticular Locking Plates and Screws, 2358, 2359 Series | 2004 | 162 |
| K102694 | 2.4MM VA-LCP Dorsal Distal Radius Plates | 2010 | 50 |
| K123875 | Aap Radius Set 2.5 | 2013 | 21 |
| K153716 | GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System. | 2016 | 15 |
| K142419 | Mini and Micro Fragments Reconstruction System-NEOFIX | 2014 | 8 |
| K061917 | Distal Radius Locking Plating System | 2006 | 6 |
| K161292 | Distal Radius Plating System | 2016 | 4 |
| K163383 | GREENS BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, GREENS BRAND… | 2017 | 3 |
| K182650 | Arzzt Distal Radius and Ulna System | 2019 | 1 |
| K153374 | CARE BRAND Locking Bone Plates and Screws Osteosynthesis Plating System, CARE Brand of… | 2016 | 1 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 28, 2016 | Class II | Z-2389-2016 | Nonconforming Material/Component | It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature… |
| Nov 2, 2015 | Class II | Z-0323-2016 | Process control | It was discovered that the above part number and lot of 2.4mm Locking Screws may contain the incorrect package insert (GP2613, VA-LCP Distal Radius System). The correct package insert for the above part is GP2615 (Modular Mini Fragment… |
| Dec 5, 2006 | Not classified | Z-0489-2007 | Other | Hole is malformed, leading to reduced construct strength. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K040593 (Zimmer, Inc., 2004) has gone furthest, with 162 descendants of its own.
How does its category compare?
All HRS devicesHRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · HRS
54in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HRS
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HRS
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HRS
2,306in 2025
MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 63 | 2 | 339 |
| 2008 | 35 | 87 | 11 | 311 |
| 2009 | 47 | 85 | 3 | 435 |
| 2010 | 49 | 100 | 16 | 491 |
| 2011 | 43 | 88 | 5 | 773 |
| 2012 | 59 | 106 | 4 | 1,663 |
| 2013 | 56 | 89 | 11 | 2,643 |
| 2014 | 76 | 57 | 0 | 2,413 |
| 2015 | 56 | 97 | 7 | 3,095 |
| 2016 | 44 | 109 | 5 | 2,276 |
| 2017 | 57 | 122 | 3 | 2,220 |
| 2018 | 40 | 116 | 2 | 2,663 |
| 2019 | 46 | 89 | 25 | 3,359 |
| 2020 | 48 | 88 | 1 | 2,790 |
| 2021 | 40 | 87 | 9 | 2,442 |
| 2022 | 47 | 144 | 3 | 1,700 |
| 2023 | 40 | 120 | 1 | 2,049 |
| 2024 | 54 | 96 | 0 | 1,849 |
| 2025 | 54 | 90 | 1 | 2,306 |
| 2026 | 51 | 82 | 3 | — |
What is it indicated for?
The Synthes Locking Distal Radius Plating System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K012114 · printed Nov 6, 2026
Cite this page
510(k) Report. “K012114: Synthes Locking Distal Radius Plating System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K012114
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026