510(k) Report

510(k) · Cleared

Recalled · Class I · 2025

Cadd Administration Set with Flow Stop, Medication Cassette Reservoir with Flow Stop and Extension Set

K040636510(k) TraditionalCleared May 19, 200470-day reviewClass II21 CFR 880.5440Recalls 6 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smiths Medical MD, Inc. · FPA Set, Administration, Intravascular

Cleared on May 19, 2004 after a 70-day review. The FDA summary names its predicate without a number. 4 of the 8 devices in its family, counting itself, have a Class I recall. 7 later devices descend from it. It has been recalled 6 times, including a Class I recall.

Decided
May 19, 2004
Review
70 days
Class
Class II
Product code
FPA
Recalls
6 (1 Class I)
Predicates
0

7 later devices descend from it, 3 citing it directly. 4 of the 8 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2017: 2 descendants cleared, 2 with a Class I recall22008: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2023: 1 descendant cleared2024: 2 descendants cleared200020052010201520202025
  1. This deviceK040636 Cadd Administration Set with Flow Stop, Medication Cassette…Smiths Medical MD, Inc. · 2004
  2. Descendant · cites it directly · 3 of its ownK170982 CADD®- Solis Ambulatory Infusion Pump with Wireless Communication…Smiths Medical Ads, Inc. · 2017 · Recalled (Class I)
  3. Descendant · cites it directly · 2 of its ownK081156 Cadd Medication Cassette ReservoirSmiths Medical MD, Inc. · 2008 · Recalled (Class I)
  4. Descendant · 2 generations on · 1 of its ownK232793 Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES)…Zevex, Inc. · 2024
  5. Descendant · cites it directly · 1 of its ownK162219 CADD® Infusion Disposables Portfolio with NRFit™ connectorsSmiths Medical Asd, Inc. · 2017 · Recalled (Class I)
  6. Descendant · 3 generations on · 0 of its ownK242660 Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES)…Zevex, Inc. · 2024

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2008: 12017: 22021: 12023: 12024: 2
2008high 2 (2017) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 7 · See all
  • K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion…Smiths Medical Ads, Inc. · 2017 · cites it directly3descendants
  • K081156Cadd Medication Cassette ReservoirSmiths Medical MD, Inc. · 2008 · cites it directly2descendants
  • K232793Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin…Zevex, Inc. · 2024 · 2 generations on1descendant
  • K162219CADD® Infusion Disposables Portfolio with NRFit™ connectorsSmiths Medical Asd, Inc. · 2017 · cites it directly1descendant
  • K242660Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin…Zevex, Inc. · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K040636Cadd Administration Set with Flow Stop, Medication Cassette Reservoir with…Smiths Medical MD, Inc.2004—
+1K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis…Smiths Medical Ads, Inc.2017Class I
+1K081156Cadd Medication Cassette ReservoirSmiths Medical MD, Inc.2008Class I
+2K232793Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software…Zevex, Inc.2024—
+1K162219CADD® Infusion Disposables Portfolio with NRFit™ connectorsSmiths Medical Asd, Inc.2017Class I
+3K242660Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software…Zevex, Inc.2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200811
201722
202110
202310
202420
FIG. 1 — Lineage of K040636, 2004–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—CADD® Medication Cassette Reservoir
Named without a submission number
—
*—CADD® Administration Set
Named without a submission number
—

What descends from it? (7)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion… Class ISmiths Medical Ads, Inc.20173
K081156Cadd Medication Cassette Reservoir Class ISmiths Medical MD, Inc.20082
K162219CADD® Infusion Disposables Portfolio with NRFit™ connectors Class ISmiths Medical Asd, Inc.20171

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 6, 2025Class IIZ-2522-2025Under Investigation by firmCertain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
Dec 9, 2022Class IZ-0963-2023Under Investigation by firmThere are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
Mar 17, 2017Class IIZ-1767-2017Packaging process controlSmiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD¿ Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the product. Although the…
Apr 22, 2016Class IIZ-1831-2016Nonconforming Material/ComponentSmiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.
Feb 2, 2015Class IIZ-1286-2015Under Investigation by firmSmiths Medical has become aware of an issue with specific lots of 50mL CADD" Medication Cassette Reservoirs (Cassette). Some Cassettes may leak at the sealing area of the pump tube and medication bag. Smiths Medical has received no…
Dec 14, 2007Not classifiedZ-0879-2008Process controlLeakage associated with the CADD Medication Cassette Reservoirs for use with the CADD Ambulatory Infusion Pumps

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion Pump, Model 2110…Smiths Medical Ads, Inc. · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 15 Class I recalls
  • K162219CADD® Infusion Disposables Portfolio with NRFit™ connectorsSmiths Medical Asd, Inc. · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K081156Cadd Medication Cassette ReservoirSmiths Medical MD, Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K170982 (Smiths Medical Ads, Inc., 2017) has gone furthest, with 3 descendants of its own.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

The CADD® Medication Cassette Reservoir with Flow Stop is designed for use with the CADD® pumps (except CADD-Micro® and CADD-TPN®) for delivery of medications and fluids.

The CADD® Administration Set with Flow Stop is designed for use with the CADD® pumps (except CADD-Microo and CADD-TPN ) for delivery of medications and fluids.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040636: Cadd Administration Set with Flow Stop, Medication Cassette Reservoir with Flow Stop and Extension Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K040636

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026