510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety Software and Supported Syringes (PSS)

K111386510(k) TraditionalCleared Aug 29, 2011103-day reviewClass II21 CFR 880.5725Recalls 16 (6 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smiths Medical MD, Inc. · FRN Pump, Infusion

Cleared on Aug 29, 2011 as substantially equivalent to K040899 (Medex, Inc., 2004) and 3 other predicates after a 103-day review. Its family tree reaches back 3 generations to 1994. 5 of the 16 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 16 times, including 6 Class I recalls.

Decided
Aug 29, 2011
Review
103 days
Class
Class II
Product code
FRN
Recalls
16 (6 Class I)
Predicates
4

Its longest line of descent runs back 3 generations to K940147 (1994). 11 ancestors in all, the oldest K890120 (1989). 4 later devices descend from it, 2 citing it directly. 5 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2004: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor1999: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor2016: 2 descendants cleared2022: 1 descendant cleared, 1 with a Class I recall12024: 1 descendant cleared, 1 with a Class I recall1K940147 · Auto Syringe AS40A Infusion Pump · Baxter Corp. · 1994 · recalledK974332 · Variable Pressure Syringe Pump · Alaris Medical Systems, Inc. · 1998K023264 · Medley Syringe Pump Module System, Model 8110 · Alaris Medical Systems, Inc. · 2002 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors, 1 with a Class I recall12 generations back: 1 other ancestor3 generations back: 4 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 2 descendants, 1 with a Class I recall1 K023264 · Medley Syringe Pump Module System, Model 8110 · Alaris Medical Systems, Inc. · 2002 · Class I recallK974332 · Variable Pressure Syringe Pump · Alaris Medical Systems, Inc. · 1998K940147 · Auto Syringe AS40A Infusion Pump · Baxter Corp. · 1994 · recalled0
  1. Root · 3 generations backK940147 Auto Syringe AS40A Infusion PumpBaxter Corp. · 1994 · Recalled
  2. 2 generations backK974332 Variable Pressure Syringe PumpAlaris Medical Systems, Inc. · 1998
  3. 1 generation backK023264 Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Recalled (Class I)
  4. This deviceK111386 Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2…Smiths Medical MD, Inc. · 2011
  5. Descendant · cites it directly · 1 of its ownK211125 Novum IQ Syringe PumpBaxter Healthcare Corporation · 2022 · Recalled (Class I)
  6. Descendant · cites it directly · 1 of its ownK160482 t:slim Insulin Delivery System, t:flex Insulin Delivery System…Tandem Diabetes Care, Inc. · 2016
  7. Descendant · 2 generations on · 0 of its ownK242390 Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation · 2024 · Recalled (Class I)
  8. Descendant · 2 generations on · 0 of its ownK162080 t:slim Insulin Delivery System, t:flex Insulin Delivery System…Tandem Diabetes Care, Inc. · 2016

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K211125Novum IQ Syringe PumpBaxter Healthcare Corporation · 2022 · cites it directly1descendant
  • K160482t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device UpdaterTandem Diabetes Care, Inc. · 2016 · cites it directly1descendant
  • K242390Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation · 2024 · 2 generations on0descendants
  • K162080t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device UpdaterTandem Diabetes Care, Inc. · 2016 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K940147Auto Syringe AS40A Infusion PumpBaxter Corp.19941
−2K974332Variable Pressure Syringe PumpAlaris Medical Systems, Inc.1998—
−1K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc.2002Class I
0 (this device)K111386Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication…Smiths Medical MD, Inc.2011—
+1K211125Novum IQ Syringe PumpBaxter Healthcare Corporation2022Class I
+1K160482t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device…Tandem Diabetes Care, Inc.2016—
+2K242390Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation2024Class I
+2K162080t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device…Tandem Diabetes Care, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010
199510
199610
199910
200310
200411
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201620
202211
202411
FIG. 1 — Lineage of K111386, 1994–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040899Medex 3000 Series MR Syringe Infusion Pump
Summary page 1
…Product Code: FRN, MRZ Regulation: 21 CFR §880.5725 Predicate Device(s) The predicate devices are the currently marketed Medfusion® Model 3500 Syringe Infusion Pump (K040899), PharmGuard® Toolbox (K042432), ALARIS Medical Systems, Inc. Medley ™ Syringe Pump Module (K023264), and the ALARIS Medical Syst…
Medex, Inc.2004Class I
*K042432Pharmguard Toolbox, Model G6000782
Summary page 1
…tion: 21 CFR §880.5725 Predicate Device(s) The predicate devices are the currently marketed Medfusion® Model 3500 Syringe Infusion Pump (K040899), PharmGuard® Toolbox (K042432), ALARIS Medical Systems, Inc. Medley ™ Syringe Pump Module (K023264), and the ALARIS Medical Systems Inc. Medley™ System with Me…
Medex, Inc.2005—
*K023264Medley Syringe Pump Module System, Model 8110
Summary page 1
…e the currently marketed Medfusion® Model 3500 Syringe Infusion Pump (K040899), PharmGuard® Toolbox (K042432), ALARIS Medical Systems, Inc. Medley ™ Syringe Pump Module (K023264), and the ALARIS Medical Systems Inc. Medley™ System with Medication Management System (K030459). Device Description The Medfusi…
Alaris Medical Systems, Inc.2002Class I
*K030459Medley System with Medication Management System (MMS)
Summary page 1
…box (K042432), ALARIS Medical Systems, Inc. Medley ™ Syringe Pump Module (K023264), and the ALARIS Medical Systems Inc. Medley™ System with Medication Management System (K030459). Device Description The Medfusion® Model 4000 Syringe Infusion Pump software version 1.1 is a wireless capable electro-mechani…
Alaris Medical Systems, Inc.2003—

Also cleared against K040899: 2 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211125Novum IQ Syringe Pump Class IBaxter Healthcare Corporation20221
K160482t:slim Insulin Delivery System, t:flex Insulin Delivery System, Tandem Device UpdaterTandem Diabetes Care, Inc.20161

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 19, 2023Class IZ-0864-2024Software Design ChangeMedfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose…
Dec 19, 2023Class IIZ-1541-2024Process controlMedfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the…
Dec 19, 2023Class IIZ-1543-2024Process controlMedfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the…
Aug 16, 2023Class IIZ-0079-2024Under Investigation by firmA force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will…
Oct 10, 2022Class IIZ-0272-2023Under Investigation by firmPotential for delay in sending Infusion Start Messages to third-party Electronic Health Record systems which may affect infusion auto-programming on the pump and documentation
Apr 18, 2022Class IZ-1271-2022Device DesignMultiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues…
Jun 26, 2020Class IZ-2734-2020Software designInaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
Jun 26, 2020Class IZ-2735-2020Software designInaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
Oct 28, 2019Class IZ-0610-2020Software designThere is an anomaly in Medfusion 4000 Syringe Pump Firmware version 1.7.0 that could potentially result in loss of primary audible and visual battery alarm functionality, and interruption of therapy.
Oct 15, 2018Class IIZ-2554-2019Process controlCustomers who utilize PharmGuard¿ Server Software (PGS) with Medfusion¿ 4000 pumps manufactured prior to November 10, 2017, may experience wireless network connectivity issues on certain pumps due to a potential Radio Module issue on some…
Apr 13, 2018Class IIZ-2662-2018Process change controlCertain Barrel Clamp Guides may contain a ridge that could potentially lead to spring slippage, resulting in the inability of the pump to recognize a loaded syringe or the pump, misidentifying the size of a loaded syringe.
Nov 13, 2017Class IIZ-1134-2018Process change controlCertain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump)…
Aug 22, 2017Class IIZ-0313-2018Component design/selectionPotential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.
Dec 3, 2015Class IIZ-0653-2022Component change controlThere have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall…
Nov 22, 2013Class IIZ-0540-2014Device DesignSmiths Medical is conducting a recall involving all Medfusion 4000 Syringe Infusion Pumps. Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other…
Feb 14, 2012Class IZ-2107-2012Software designSmiths Medical has identified a software anomaly in the Medfusion 4000 Syringe Infusion Pump that causes a device history log corruption and triggers a Watchdog Fail-Safe alarm. When the device exhibits this failure mode, visual and…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K040899Medex 3000 Series MR Syringe Infusion PumpMedex, Inc. · 2004 · Direct predicate · ◆ in the 2004 ancestors bar of FIG. 14 Class I recalls
  • K023264Medley Syringe Pump Module System, Model 8110Alaris Medical Systems, Inc. · 2002 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K211125Novum IQ Syringe PumpBaxter Healthcare Corporation · 2022 · Cites it directly · ◆ in the 2022 descendants bar of FIG. 11 Class I recall
  • K242390Novum IQ Syringe Pump (40800BAXUS)Baxter Healthcare Corporation · 2024 · 2 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 4.7 years on median.
  • 5 of the 11 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K211125 (Baxter Healthcare Corporation, 2022) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K040899.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The PharmGuard® Supported Syringes is an accessory to the PharmGuard® Toolbox 2 Medication Safety Software that expands the available supported syringes that can be selected when creating user defined configuration parameters using the PharmGuard® Toolbox 2 Medication Safety Software. The Medfusion® Model 4000 Syringe Infusion Pump is indicated for the following uses: In the administration of fluids requiring precisely controlled infusion rates including…

Read the full statement (253 words)

The PharmGuard® Supported Syringes is an accessory to the PharmGuard® Toolbox 2 Medication Safety Software that expands the available supported syringes that can be selected when creating user defined configuration parameters using the PharmGuard® Toolbox 2 Medication Safety Software.

The Medfusion® Model 4000 Syringe Infusion Pump is indicated for the following uses:

In the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, Enteral solutions and other therapeutic fluids.

By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral,

By the following delivery modes: continuous, volume/time, mass, body weight, custom dilution, intermittent and bolus.

In critical care, anesthesia, neonatal and pediatric applications or other healthcare settings where the use of the syringe infusion pump can be monitored or supervised by a clinician.

Inside the MRI room mounted outside the 150 Gauss line and with shielded magnets of field strength of 1.5 Tesla.

The PharmGuard® Toolbox 2 Medication Safety Software is indicated to provide a vehicle to create and securely upload user-defined configuration parameters a library of drug names and associated user-defined infusion parameters to Medfusion® Model 4000 Syringe Infusion Pump with software version V1.0 or higher. The PharmGuard® Toolbox 2 Medication Safety Software collects and downloads operational, infusion and alarm history events, library usage counts, and PharmGuard® safety events from the Medfusion® Pump. It is intended to provide a dataset users can collect and analyze to improve overall processes and lessen the likelihood of operator error when entering infusion parameters into the Medfusion® Pump.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K111386: Medfusion Model 4000 Syringe Infusion Pump, Pharmguard Toolbox 2 Medication Safety Software and Supported Syringes (PSS).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111386

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.