Intra-Aortic Balloon (Iab)

K040801510(k) SpecialCleared May 6, 200438-day reviewClass II21 CFR 870.3535Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arrow Intl., Inc. · DSP System, Balloon, Intra-Aortic And Control

Cleared on May 6, 2004 as substantially equivalent to K970689 (Arrow Intl., Inc., 1998) and 4 other predicates after a 38-day review. Its family tree reaches back 3 generations to 1989. 9 of the 18 devices in its family, counting itself, have a Class I recall. 9 later devices descend from it. It has been recalled 13 times.

Decided
May 6, 2004
Review
38 days
Class
Class II
Product code
DSP
Recalls
13
Predicates
5

Its longest line of descent runs back 3 generations to K892799 (1989). 8 ancestors in all. 9 later devices descend from it, 1 citing it directly. 9 of the 18 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors, 2 with a Class I recall21998: 2 other ancestors, 1 with a Class I recall11996: 1 other ancestor2007: 1 descendant cleared, 1 with a Class I recall12017: 2 descendants cleared, 1 with a Class I recall12011: 2 descendants cleared, 1 with a Class I recall12012: 1 descendant cleared2013: 1 descendant cleared2018: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall1K892799 · 8 FR Percutan Double Lumen Intra-Aortic Balloon · Kontron Instruments, Inc. · 1989K963920 · 8 FR.-30CC Narrowflex Intra-Aortic Balloon Catheter, · Arrow Intl., Inc. · 1997K021462 · Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement System, Models Iab-05830-F,IAB5840 · Arrow Intl., Inc. · 2002 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors, 2 with a Class I recall22 generations back: 2 other ancestors, 1 with a Class I recall1 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 3 descendants, 1 with a Class I recall13 generations forward: 4 descendants, 2 with a Class I recall24 generations forward: 1 descendant, 1 with a Class I recall1 K021462 · Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement System, Models Iab-05830-F,IAB5840 · Arrow Intl., Inc. · 2002 · Class I recallK963920 · 8 FR.-30CC Narrowflex Intra-Aortic Balloon Catheter, · Arrow Intl., Inc. · 1997K892799 · 8 FR Percutan Double Lumen Intra-Aortic Balloon · Kontron Instruments, Inc. · 19890
  1. Root · 3 generations backK892799 8 FR Percutan Double Lumen Intra-Aortic BalloonKontron Instruments, Inc. · 1989
  2. 2 generations backK963920 8 FR.-30CC Narrowflex Intra-Aortic Balloon Catheter,Arrow Intl., Inc. · 1997
  3. 1 generation backK021462 Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor…Arrow Intl., Inc. · 2002 · Recalled (Class I)
  4. This deviceK040801 Intra-Aortic Balloon (Iab)Arrow Intl., Inc. · 2004
  5. Descendant · cites it directly · 8 of its ownK063525 Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter…Datascope Corp. · 2007 · Recalled (Class I)
  6. Descendant · 2 generations on · 3 of its ownK112372 Cardiosave Hybrid Intra-Aortic Balloon PumpCardiac Assist, Maquet Cardiovascular, LLC · 2011 · Recalled (Class I)
  7. Descendant · 2 generations on · 2 of its ownK112327 Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and…Cardiac Assist, Maquet Cardiovascular, LLC · 2011 · Recalled
  8. Descendant · 3 generations on · 1 of its ownK151254 CARDIOSAVE Hybrid Intra-Aortic Balloon PumpDatascope Corp. · 2015 · Recalled (Class I)
  9. Descendant · 3 generations on · 0 of its ownK181122 CARDIOSAVE Intra-Aortic Balloon PumpDatascope Corp. · 2018 · Recalled (Class I)

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2007: 12011: 22012: 12013: 12015: 12017: 22018: 1
2007high 2 (2011) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 9 · See all
  • K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300…Datascope Corp. · 2007 · cites it directly8descendants
  • K112372Cardiosave Hybrid Intra-Aortic Balloon PumpCardiac Assist, Maquet Cardiovascular, LLC · 2011 · 2 generations on3descendants
  • K112327Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and AccessoriesCardiac Assist, Maquet Cardiovascular, LLC · 2011 · 2 generations on2descendants
  • K151254CARDIOSAVE Hybrid Intra-Aortic Balloon PumpDatascope Corp. · 2015 · 3 generations on1descendant
  • K181122CARDIOSAVE Intra-Aortic Balloon PumpDatascope Corp. · 2018 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K8927998 FR Percutan Double Lumen Intra-Aortic BalloonKontron Instruments, Inc.1989—
−2K9639208 FR.-30CC Narrowflex Intra-Aortic Balloon Catheter,Arrow Intl., Inc.1997—
−1K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a…Arrow Intl., Inc.2002Class I
0 (this device)K040801Intra-Aortic Balloon (Iab)Arrow Intl., Inc.2004—
+1K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300…Datascope Corp.2007Class I
+2K112372Cardiosave Hybrid Intra-Aortic Balloon PumpCardiac Assist, Maquet Cardiovascular, LLC2011Class I
+2K112327Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and AccessoriesCardiac Assist, Maquet Cardiovascular, LLC20111
+3K151254CARDIOSAVE Hybrid Intra-Aortic Balloon PumpDatascope Corp.2015Class I
+3K181122CARDIOSAVE Intra-Aortic Balloon PumpDatascope Corp.2018Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199821
200022

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200711
201121
201210
201310
201511
201721
201811
FIG. 1 — Lineage of K040801, 1989–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSP device

  1. #1 K250542 AC3™ Range™ Intra-Aortic Balloon Pumpscore 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K970689Iab-04840-U;Iab-04250-U;Iab-04240-U;Iab-04230-U
Summary page 4
…s substantial equivalence. Table 1: Predicate Devices K040801 Teleflex: Arrow International, Inc vavaw teieflex com 510(k) Intra-Aortic Balloon Description Catalog K970689 | IAB 8Fr, 30/40cc and, 10 Fr, 50cc with Peel-Away Hemostasis IAB-042XX- U K963920 | |AB 8Fr, 30cc Arrow NarrowFlex Universal IAB…
Arrow Intl., Inc.1998—
*K9639208 FR.-30CC Narrowflex Intra-Aortic Balloon Catheter,
Summary page 4
…nternational, Inc vavaw teieflex com 510(k) Intra-Aortic Balloon Description Catalog K970689 | IAB 8Fr, 30/40cc and, 10 Fr, 50cc with Peel-Away Hemostasis IAB-042XX- U K963920 | |AB 8Fr, 30cc Arrow NarrowFlex Universal IAB-04830-U K993966 | IAB 8Fr, 40cc Arrow NarrowFlex Universal IAB-04840-U K000729 | |…
Arrow Intl., Inc.1997—
*K9939668FR. Narrowflex Universal Intra-Aortic Balloon Catheter
Summary page 4
…oon Description Catalog K970689 | IAB 8Fr, 30/40cc and, 10 Fr, 50cc with Peel-Away Hemostasis IAB-042XX- U K963920 | |AB 8Fr, 30cc Arrow NarrowFlex Universal IAB-04830-U K993966 | IAB 8Fr, 40cc Arrow NarrowFlex Universal IAB-04840-U K000729 | |AB 8Fr, 30/40cc Arrow Ultra Series IAB-058XX-U / IAB-68XX-U K0…
Arrow Intl., Inc.2000Class I
*K000729Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC Universal
Summary page 4
…50cc with Peel-Away Hemostasis IAB-042XX- U K963920 | |AB 8Fr, 30cc Arrow NarrowFlex Universal IAB-04830-U K993966 | IAB 8Fr, 40cc Arrow NarrowFlex Universal IAB-04840-U K000729 | |AB 8Fr, 30/40cc Arrow Ultra Series IAB-058XX-U / IAB-68XX-U K021462 | |AB 8Fr, 30/40cc Arrow Ultra Series with Light Wave Sen…
Arrow Intl., Inc.2000Class I
*K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement…
Summary page 4
…w NarrowFlex Universal IAB-04830-U K993966 | IAB 8Fr, 40cc Arrow NarrowFlex Universal IAB-04840-U K000729 | |AB 8Fr, 30/40cc Arrow Ultra Series IAB-058XX-U / IAB-68XX-U K021462 | |AB 8Fr, 30/40cc Arrow Ultra Series with Light Wave Sensor IAB-058XX-LWS E. Description of Device IAB's are designed to provi…
Arrow Intl., Inc.2002Class I

What descends from it? (9)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300… Class IDatascope Corp.20078

Has anything in its family been recalled? (9 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 2, 2009Not classifiedZ-1216-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1214-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1219-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1221-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1218-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1210-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1215-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1220-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1217-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1211-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1212-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1222-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1213-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement System, Models…Arrow Intl., Inc. · 2002 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K000729Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC UniversalArrow Intl., Inc. · 2000 · Direct predicate · ◆ in the 2000 ancestors bar of FIG. 17 Class I recalls
  • K9939668FR. Narrowflex Universal Intra-Aortic Balloon CatheterArrow Intl., Inc. · 2000 · Direct predicate · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K981660Bard 8 Fr., 40cc. & 7Fr., 30cc Intra-Aortic Balloon CathetersC.R. Bard, Inc. · 1998 · 2 generations back · ◆ in the 1998 ancestors bar of FIG. 12 Class I recalls
  • K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300 Intra-Aortic Balloon Pump…Datascope Corp. · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K112372Cardiosave Hybrid Intra-Aortic Balloon PumpCardiac Assist, Maquet Cardiovascular, LLC · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 110 Class I recalls
  • K181122CARDIOSAVE Intra-Aortic Balloon PumpDatascope Corp. · 2018 · 3 generations on · ◆ in the 2018 descendants bar of FIG. 127 Class I recalls
  • K151254CARDIOSAVE Hybrid Intra-Aortic Balloon PumpDatascope Corp. · 2015 · 3 generations on · ◆ in the 2015 descendants bar of FIG. 112 Class I recalls
  • K163542CARDIOSAVE Hybrid and Rescue Intra-Aortic Balloon Pump, CARDIOSAVE Li-Ion Battery Transport and Storage CaseDatascope Corp. · 2017 · 4 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls

What does this lineage say?

  • The line of descent spans 15 years and passes through 2 companies; a new generation was cleared every 5.0 years on median.
  • 4 of the 8 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K063525 (Datascope Corp., 2007) has gone furthest, with 8 descendants of its own.

How does its category compare?

All DSP devices

DSP System, Balloon, Intra-Aortic And Control: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DSP

1in 2025

2007: 22009: 22011: 22012: 22013: 12015: 12017: 32018: 12019: 32020: 22023: 12025: 1
2007high 3 (2017) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSP

140in 2019

2017: 73 days median2019: 140 days median
2007high 140 (2019) · low 73 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSP

4in 2025

2007: 32009: 142010: 82012: 32014: 12016: 9, 9 Class I2018: 2, 2 Class I2019: 32021: 5, 3 Class I2022: 33, 33 Class I2024: 12025: 4
2007high 33 (2022) · low 0 (2008)2026

Class I (47 since 2007) Class II/III not classified

Adverse event reports per year · DSP

7,422in 2025

2007: 4612008: 6172009: 6562010: 1,0192011: 5902012: 4902013: 4952014: 1,7382015: 4042016: 4192017: 2,1642018: 3,3312019: 3,5522020: 3,3932021: 4,3452022: 5,1132023: 6,9802024: 7,3172025: 7,422
2007high 7,422 (2025) · low 404 (2015)2025

MAUDE reports naming product code DSP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—3461
20080—0617
20092—14656
20100—81,019
20112—0590
20122—3490
20131—0495
20140—11,738
20151—0404
20160—9 (9)419
201737302,164
20181—2 (2)3,331
2019314033,552
20202—03,393
20210—5 (3)4,345
20220—33 (33)5,113
20231—06,980
20240—17,317
20251—47,422
20260—0—

What is it indicated for?

The Intra-Aortic Balloon (IAB) is clinically indicated for the following conditions:

a. Acute Coronary Syndrome

b. Cardiac and Non-Cardiac Surgery

c. Complications of Heart Failure

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K040801: Intra-Aortic Balloon (Iab).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K040801

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026