Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC Universal

K000729510(k) SpecialCleared May 19, 200074-day reviewClass II21 CFR 870.3535Recalls 27 (7 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arrow Intl., Inc. · DSP System, Balloon, Intra-Aortic And Control

Cleared on May 19, 2000 after a 74-day review. The FDA summary names its predicate without a number. 11 of the 22 devices in its family, counting itself, have a Class I recall. 21 later devices descend from it. It has been recalled 27 times, including 7 Class I recalls.

Decided
May 19, 2000
Review
74 days
Class
Class II
Product code
DSP
Recalls
27 (7 Class I)
Predicates
0

21 later devices descend from it, 3 citing it directly. 11 of the 22 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared2002: 1 descendant cleared, 1 with a Class I recall12006: 2 descendants cleared, 1 with a Class I recall12020: 2 descendants cleared, 2 with a Class I recall22019: 2 descendants cleared2007: 1 descendant cleared, 1 with a Class I recall12009: 1 descendant cleared2017: 3 descendants cleared, 2 with a Class I recall22011: 2 descendants cleared, 1 with a Class I recall12012: 1 descendant cleared2013: 1 descendant cleared2018: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared2025: 1 descendant cleared200020052010201520202025
  1. This deviceK000729 Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and…Arrow Intl., Inc. · 2000
  2. Descendant · cites it directly · 18 of its ownK021462 Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor…Arrow Intl., Inc. · 2002 · Recalled (Class I)
  3. Descendant · 2 generations on · 14 of its ownK060309 Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · Recalled (Class I)
  4. Descendant · cites it directly · 9 of its ownK040801 Intra-Aortic Balloon (Iab)Arrow Intl., Inc. · 2004 · Recalled
  5. Descendant · 2 generations on · 8 of its ownK063525 Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter…Datascope Corp. · 2007 · Recalled (Class I)
  6. Descendant · 3 generations on · 4 of its ownK162820 AC3 Series IABP SystemArrow International, Inc. · 2017 · Recalled (Class I)

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

2002: 12004: 12006: 22007: 12009: 12011: 22012: 12013: 12015: 12017: 32018: 12019: 22020: 22023: 12025: 1
2002high 3 (2017) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 21 · See all
  • K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement…Arrow Intl., Inc. · 2002 · cites it directly18descendants
  • K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · 2 generations on14descendants
  • K040801Intra-Aortic Balloon (Iab)Arrow Intl., Inc. · 2004 · cites it directly9descendants
  • K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300…Datascope Corp. · 2007 · 2 generations on8descendants
  • K162820AC3 Series IABP SystemArrow International, Inc. · 2017 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K000729Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC UniversalArrow Intl., Inc.2000—
+1K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a…Arrow Intl., Inc.2002Class I
+2K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc.2006Class I
+1K040801Intra-Aortic Balloon (Iab)Arrow Intl., Inc.200413
+2K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300…Datascope Corp.2007Class I
+3K162820AC3 Series IABP SystemArrow International, Inc.2017Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200211
200410
200621
200711
200910
201121
201210
201310
201511
201732
201811
201920
202022
202310
202510
FIG. 1 — Lineage of K000729, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

11 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSP device

  1. #1 K250542 AC3™ Range™ Intra-Aortic Balloon Pumpscore 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Arrow 8Fr. 30cc and 40cc Universal Intra-Aortic Balloon Catheter Kits
Named without a submission number
—

What descends from it? (21)

See all 21

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement… Class IArrow Intl., Inc.200218
K040801Intra-Aortic Balloon (Iab)Arrow Intl., Inc.20049
K062582Supracor Balloon CatheterAbiomed, Inc.20061

Has anything in its family been recalled? (11 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 29, 2024Class IZ-1879-2024Nonconforming Material/ComponentTeleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely…
Apr 29, 2024Class IZ-1880-2024Nonconforming Material/ComponentTeleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely…
Feb 10, 2016Class IZ-1058-2016Process controlThe sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
Feb 10, 2016Class IZ-1059-2016Process controlThe sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
Feb 10, 2016Class IZ-1056-2016Process controlThe sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
Feb 10, 2016Class IZ-1060-2016Process controlThe sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
Feb 10, 2016Class IZ-1062-2016Process controlThe sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
Nov 27, 2012Class IIZ-1064-2013Process controlPotential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
Nov 27, 2012Class IIZ-1065-2013Process controlPotential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
Oct 11, 2010Not classifiedZ-0790-2011Device DesignSuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
Oct 11, 2010Not classifiedZ-0791-2011Device DesignSuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
Oct 11, 2010Not classifiedZ-0792-2011Device DesignSuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
Oct 11, 2010Not classifiedZ-0793-2011Device DesignSuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
Oct 11, 2010Not classifiedZ-0969-2011Device DesignSuperArrow Flex IAB catheter becomes stuck in the sheath, unable to move the IAB catheter forward or backward, causing a delay in therapy, bleeding or arterial injury.
Feb 2, 2009Not classifiedZ-1216-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1214-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1219-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1221-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1218-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1210-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1215-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1220-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1217-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1211-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1212-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1222-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Feb 2, 2009Not classifiedZ-1213-2009Process designA fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement System, Models…Arrow Intl., Inc. · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 12 Class I recalls
  • K200634Arrow FiberOptix Intra-Aortic Balloon Catheter KitArrow International, Inc. · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300 Intra-Aortic Balloon Pump…Datascope Corp. · 2007 · 2 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 128 Class I recalls
  • K201112AC3 IABP, AC3 Optimus IABP, AutoCAT2 Wave IABP, AutoCAT2 IABPArrow International, Inc. · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 111 Class I recalls
  • K162820AC3 Series IABP SystemArrow International, Inc. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls
  • K112372Cardiosave Hybrid Intra-Aortic Balloon PumpCardiac Assist, Maquet Cardiovascular, LLC · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 110 Class I recalls
  • K181122CARDIOSAVE Intra-Aortic Balloon PumpDatascope Corp. · 2018 · 4 generations on · ◆ in the 2018 descendants bar of FIG. 127 Class I recalls
  • K151254CARDIOSAVE Hybrid Intra-Aortic Balloon PumpDatascope Corp. · 2015 · 4 generations on · ◆ in the 2015 descendants bar of FIG. 112 Class I recalls
  • K163542CARDIOSAVE Hybrid and Rescue Intra-Aortic Balloon Pump, CARDIOSAVE Li-Ion Battery Transport and Storage CaseDatascope Corp. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls

What does this lineage say?

  • Of its descendants, K021462 (Arrow Intl., Inc., 2002) has gone furthest, with 18 descendants of its own.

How does its category compare?

All DSP devices

DSP System, Balloon, Intra-Aortic And Control: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DSP

1in 2025

2007: 22009: 22011: 22012: 22013: 12015: 12017: 32018: 12019: 32020: 22023: 12025: 1
2007high 3 (2017) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSP

140in 2019

2017: 73 days median2019: 140 days median
2007high 140 (2019) · low 73 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSP

4in 2025

2007: 32009: 142010: 82012: 32014: 12016: 9, 9 Class I2018: 2, 2 Class I2019: 32021: 5, 3 Class I2022: 33, 33 Class I2024: 12025: 4
2007high 33 (2022) · low 0 (2008)2026

Class I (47 since 2007) Class II/III not classified

Adverse event reports per year · DSP

7,422in 2025

2007: 4612008: 6172009: 6562010: 1,0192011: 5902012: 4902013: 4952014: 1,7382015: 4042016: 4192017: 2,1642018: 3,3312019: 3,5522020: 3,3932021: 4,3452022: 5,1132023: 6,9802024: 7,3172025: 7,422
2007high 7,422 (2025) · low 404 (2015)2025

MAUDE reports naming product code DSP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—3461
20080—0617
20092—14656
20100—81,019
20112—0590
20122—3490
20131—0495
20140—11,738
20151—0404
20160—9 (9)419
201737302,164
20181—2 (2)3,331
2019314033,552
20202—03,393
20210—5 (3)4,345
20220—33 (33)5,113
20231—06,980
20240—17,317
20251—47,422
20260—0—

What is it indicated for?

The Arrow Ultra 8 Intra-Aortic Balloon Catheters (IAB) 8 Fr 30cc and 40cc Universal is clinically indicated for the following conditions:

a. Acute Coronary Syndrome

b. Cardiac and Non-Cardiac Surgery

c. Complications of Heart Failure

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K000729: Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC Universal.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K000729

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026