510(k) Report

510(k) · Cleared

Recalled · Class III · 2015

BD Facscanto with BD Facsanto Software

K041074510(k) TraditionalCleared Sep 17, 2004144-day reviewClass II21 CFR 864.5220Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bd Biosciences · GKZ Counter, Differential Cell

Cleared on Sep 17, 2004 as substantially equivalent to K973483 (Becton Dickinson Immunocytometry Systems, 1998) after a 144-day review. Its family tree reaches back 3 generations to 1994. 6 later devices descend from it. It has been recalled once.

Decided
Sep 17, 2004
Review
144 days
Class
Class II
Product code
GKZ
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K925274 (1994). 6 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2006: 2 descendants cleared2015: 1 descendant cleared2009: 1 descendant cleared2017: 1 descendant cleared2020: 1 descendant clearedK925274 · Autocomp (TM) Software and Calibrite (TM) Beads · Becton Dickinson Immunocytometry Systems · 1994K961623 · Facscomp Software & Calibrite 3 Color Beads Kit/Calibrite Percp Beads · Becton Dickinson Monoclonal Center · 1996K973483 · Facscomp Software and Calibrite Beads for Use in Flow Cytometer Instrument Set · Becton Dickinson Immunocytometry Systems · 1998 · recalled1990200020102020
  1. Root · 3 generations backK925274 Autocomp (TM) Software and Calibrite (TM) BeadsBecton Dickinson Immunocytometry Systems · 1994
  2. 2 generations backK961623 Facscomp Software & Calibrite 3 Color Beads Kit/Calibrite Percp BeadsBecton Dickinson Monoclonal Center · 1996
  3. 1 generation backK973483 Facscomp Software and Calibrite Beads for Use in Flow Cytometer…Becton Dickinson Immunocytometry Systems · 1998 · Recalled
  4. This deviceK041074 BD Facscanto with BD Facsanto SoftwareBd Biosciences · 2004
  5. Descendant · cites it directly · 3 of its ownK060375 BD Multitest 6-Color TBNK Reagent with BD Trucount Tubes with Models…Bd Biosciences · 2006
  6. Descendant · cites it directly · 2 of its ownK141468 BD FACSCanto II flow cytometer (4-2-2 and 5-3 configurations), BD…Bd Biosciences · 2015
  7. Descendant · 2 generations on · 2 of its ownK090967 BD Multitest 6-Color TBNK Reagent, Model 644611Bd Biosciences · 2009 · Recalled
  8. Descendant · cites it directly · 2 of its ownK062087 BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473…Bd Biosciences · 2006 · Recalled
  9. Descendant · 2 generations on · 1 of its ownK170974 BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical…Becton, Dickinson and Company · 2017 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2006: 22009: 12015: 12017: 12020: 1
2006high 2 (2006) · low 0 (2007)2026

Most influential descendants

Showing 5 of 6 · See all
  • K060375BD Multitest 6-Color TBNK Reagent with BD Trucount Tubes with Models 337 and 166Bd Biosciences · 2006 · cites it directly3descendants
  • K141468BD FACSCanto II flow cytometer (4-2-2 and 5-3 configurations), BD FACSCanto flow…Bd Biosciences · 2015 · cites it directly2descendants
  • K090967BD Multitest 6-Color TBNK Reagent, Model 644611Bd Biosciences · 2009 · 2 generations on2descendants
  • K062087BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473; (4-2-2), Model 338962Bd Biosciences · 2006 · cites it directly2descendants
  • K170974BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical configurations), BD…Becton, Dickinson and Company · 2017 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K925274Autocomp (TM) Software and Calibrite (TM) BeadsBecton Dickinson Immunocytometry Systems1994—
−2K961623Facscomp Software & Calibrite 3 Color Beads Kit/Calibrite Percp BeadsBecton Dickinson Monoclonal Center1996—
−1K973483Facscomp Software and Calibrite Beads for Use in Flow Cytometer Instrument SetBecton Dickinson Immunocytometry Systems19981
0 (this device)K041074BD Facscanto with BD Facsanto SoftwareBd Biosciences2004—
+1K060375BD Multitest 6-Color TBNK Reagent with BD Trucount Tubes with Models 337 and 166Bd Biosciences2006—
+1K141468BD FACSCanto II flow cytometer (4-2-2 and 5-3 configurations), BD FACSCanto…Bd Biosciences2015—
+2K090967BD Multitest 6-Color TBNK Reagent, Model 644611Bd Biosciences20093
+1K062087BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473; (4-2-2)…Bd Biosciences20063
+2K170974BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical…Becton, Dickinson and Company201717

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200620
200910
201510
201710
202010
FIG. 1 — Lineage of K041074, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K973483Facscomp Software and Calibrite Beads for Use in Flow Cytometer Instrument Set
Summary page 3
…tiset software. The BD FACSCalibuar 4-color instrument was cleared with FACSComnp Software and CaliBRITE Beads by the Center for Devices and Radiological Health, under K973483 on 2/17/98. The BD EACSCanto with BD FACSCanto software and BD FACSCalibur have similar intended uses, measure the same sample t…
Becton Dickinson Immunocytometry Systems19981

Also cleared against K973483: 4 other devices →

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060375BD Multitest 6-Color TBNK Reagent with BD Trucount Tubes with Models 337 and 166Bd Biosciences20063
K141468BD FACSCanto II flow cytometer (4-2-2 and 5-3 configurations), BD FACSCanto flow…Bd Biosciences20152
K062087BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473; (4-2-2), Model 338962Bd Biosciences20062

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 15, 2015Class IIIZ-2090-2015Labeling Change ControlInstruments installed between June 2007 and September 2011 do not contain labeling pursuant to the Food and Drug Administration's (FDA's) Guidance on Laser Products- Conformance with IEC 60825-1 and IEC 60601-2-22 (Laser Notice No. 50).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 1.7 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K060375 (Bd Biosciences, 2006) has gone furthest, with 3 descendants of its own.
  • 4 other devices also cite its predicate K973483.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

Immunophenotyping in clinical laboratories, using previously cleared IVD assays for flow cytometry that utilize the lyse no-wash sample preparation method.

Immunophenotyping of lymphocyte subsets including CD3+CD8+, CD3+CD4+, CD3-CD16+ and/or CD56+, CD3-CD19+, and CD3+.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K041074: BD Facscanto with BD Facsanto Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K041074

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.