510(k) Report

510(k) · Cleared

Recalled · Class III · 2015

BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473; (4-2-2), Model 338962

K062087510(k) SpecialCleared Sep 14, 200652-day reviewClass II21 CFR 864.5220Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bd Biosciences · GKZ Counter, Differential Cell

Cleared on Sep 14, 2006 as substantially equivalent to K041074 (Bd Biosciences, 2004) and 1 other predicate after a 52-day review. Its family tree reaches back 4 generations to 1994. 2 later devices descend from it. It has been recalled 3 times.

Decided
Sep 14, 2006
Review
52 days
Class
Class II
Product code
GKZ
Recalls
3
Predicates
2

Its longest line of descent runs back 4 generations to K925274 (1994). 5 ancestors in all. 2 later devices descend from it, 1 citing it directly.

  1. Root · 4 generations backK925274 Autocomp (TM) Software and Calibrite (TM) BeadsBecton Dickinson Immunocytometry Systems · 1994
  2. 3 generations backK961623 Facscomp Software & Calibrite 3 Color Beads Kit/Calibrite Percp BeadsBecton Dickinson Monoclonal Center · 1996
  3. 2 generations backK973483 Facscomp Software and Calibrite Beads for Use in Flow Cytometer…Becton Dickinson Immunocytometry Systems · 1998 · Recalled
  4. 1 generation backK040725 BD Facscanto SystemBecton Dickinson Immunocytometry Systems · 2004 · Recalled
  5. This deviceK062087 BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473…Bd Biosciences · 2006
  6. Descendant · cites it directly · 1 of its ownK170974 BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical…Becton, Dickinson and Company · 2017 · Recalled
  7. Descendant · 2 generations on · 0 of its ownK201814 BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2, 4-3-3 and 4-3-5…Becton, Dickinson and Company · 2020 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K170974BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical configurations), BD…Becton, Dickinson and Company · 2017 · cites it directly1descendant
  • K201814BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2, 4-3-3 and 4-3-5 optical configurations)…Becton, Dickinson and Company · 2020 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K925274Autocomp (TM) Software and Calibrite (TM) BeadsBecton Dickinson Immunocytometry Systems1994—
−3K961623Facscomp Software & Calibrite 3 Color Beads Kit/Calibrite Percp BeadsBecton Dickinson Monoclonal Center1996—
−2K973483Facscomp Software and Calibrite Beads for Use in Flow Cytometer Instrument SetBecton Dickinson Immunocytometry Systems19981
−1K040725BD Facscanto SystemBecton Dickinson Immunocytometry Systems20042
0 (this device)K062087BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473; (4-2-2)…Bd Biosciences2006—
+1K170974BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical…Becton, Dickinson and Company201717
+2K201814BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2, 4-3-3 and 4-3-5 optical…Becton, Dickinson and Company20205

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202010
FIG. 1 — Lineage of K062087, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041074BD Facscanto with BD Facsanto Software
Summary page 1
…e The BD FACSCanto II system is substantially equivalent to the BD FACSCanto system with BD FACSCanto clinical software and BD FACSDiva software (premarket notifications K041074 and K040725). Device Description The BD FACSCanto II system is comprised of a flow cytometer, a fluidics cart, an…
Bd Biosciences20041
*K040725BD Facscanto System
Summary page 1
…SCanto II system is substantially equivalent to the BD FACSCanto system with BD FACSCanto clinical software and BD FACSDiva software (premarket notifications K041074 and K040725). Device Description The BD FACSCanto II system is comprised of a flow cytometer, a fluidics cart, and a computer…
Becton Dickinson Immunocytometry Systems20042

Also cleared against K041074: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170974BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical configurations), BD…Becton, Dickinson and Company20171

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 15, 2015Class IIIZ-2090-2015Labeling Change ControlInstruments installed between June 2007 and September 2011 do not contain labeling pursuant to the Food and Drug Administration's (FDA's) Guidance on Laser Products- Conformance with IEC 60825-1 and IEC 60601-2-22 (Laser Notice No. 50).
Dec 21, 2007Not classifiedZ-1530-2008Software designValues might not update in statistics views after certain functions are performed in a worksheet.
Apr 5, 2007Not classifiedZ-1525-2008Software designWhen data file containing one or no fluoresence parameters is exported, the software will automatically apply compensation to this file and all subsequently exported files.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 2.0 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • Of its descendants, K170974 (Becton, Dickinson and Company, 2017) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K041074.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

Immunophenotyping in clinical laboratories, using previously cleared in vitro diagnostic assays for flow cytometry.

Identification and enumeration of lymphocyte subsets in human cells in suspension.

For in vitro diagnostic use.

For use with or without the BD FACS Sample Prep Assistant II.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062087: BD Facscanto II Flow Cytometer, Model 338960; (5-3), Model 339473; (4-2-2), Model 338962.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K062087

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026