Isopure Complete Water System for Hemodialysis, MD 400 and MD 600 Series

K041163510(k) TraditionalCleared Oct 7, 2004157-day reviewClass II21 CFR 876.5665Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Isopure, Corp. · FIP Subsystem, Water Purification

Cleared on Oct 7, 2004 after a 157-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 4 later devices descend from it. It has been recalled once.

Decided
Oct 7, 2004
Review
157 days
Class
Class II
Product code
FIP
Recalls
1
Predicates
0

4 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2015: 1 descendant cleared2016: 1 descendant cleared2025: 1 descendant cleared2005: 1 descendant cleared200020052010201520202025
  1. This deviceK041163 Isopure Complete Water System for Hemodialysis, MD 400 and MD 600…Isopure, Corp. · 2004
  2. Descendant · cites it directly · 3 of its ownK051620 Matrix Digital Reverse Osmosis Water Treatment System, Model Eqrmd…Better Water, Inc. · 2005
  3. Descendant · cites it directly · 1 of its ownK151637 UPT Series Medical RO Water SystemSpecialty Water Technologies · 2015
  4. Descendant · cites it directly · 0 of its ownK250514 UPT Series Medical RO Water Treatment SystemSpecialty Water Technologies, Inc. · 2025
  5. Descendant · cites it directly · 0 of its ownK161844 Matrix Pre-Treatment ControllerBetter Water, LLC · 2016

Most influential descendants

Showing 4 of 4 · See all
  • K051620Matrix Digital Reverse Osmosis Water Treatment System, Model Eqrmd SeriesBetter Water, Inc. · 2005 · cites it directly3descendants
  • K151637UPT Series Medical RO Water SystemSpecialty Water Technologies · 2015 · cites it directly1descendant
  • K250514UPT Series Medical RO Water Treatment SystemSpecialty Water Technologies, Inc. · 2025 · cites it directly0descendants
  • K161844Matrix Pre-Treatment ControllerBetter Water, LLC · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K041163Isopure Complete Water System for Hemodialysis, MD 400 and MD 600 SeriesIsopure, Corp.2004—
+1K051620Matrix Digital Reverse Osmosis Water Treatment System, Model Eqrmd SeriesBetter Water, Inc.2005—
+1K151637UPT Series Medical RO Water SystemSpecialty Water Technologies2015—
+1K250514UPT Series Medical RO Water Treatment SystemSpecialty Water Technologies, Inc.2025—
+1K161844Matrix Pre-Treatment ControllerBetter Water, LLC2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
201510
201610
202510
FIG. 1 — Lineage of K041163, 2004–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FIP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FIP device

  1. #1 K251851 DIASAFEplusUS (F00013010)score 14.0 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (4)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K051620Matrix Digital Reverse Osmosis Water Treatment System, Model Eqrmd SeriesBetter Water, Inc.20053
K151637UPT Series Medical RO Water SystemSpecialty Water Technologies20151
K250514UPT Series Medical RO Water Treatment SystemSpecialty Water Technologies, Inc.20250
K161844Matrix Pre-Treatment ControllerBetter Water, LLC20160

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 24, 2011Not classifiedZ-0190-2012Employee errorThe cleaning solution Step 2, 3000-22000 GPD labeled bottles (NaOH) were filled with Step 1, 3000-22000 GPD chemical (HCl).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K051620 (Better Water, Inc., 2005) has gone furthest, with 3 descendants of its own.

How does its category compare?

All FIP devices

FIP Subsystem, Water Purification: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FIP

4in 2025

2007: 12008: 32009: 32010: 22011: 42012: 22013: 42014: 42015: 22016: 22017: 12018: 12019: 22020: 12021: 12022: 12023: 32024: 12025: 4
2007high 4 (2011) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FIP

88in 2025

2008: 202 days median2009: 167 days median2011: 116 days median2013: 125 days median2014: 126 days median2023: 69 days median2025: 88 days median
2007high 202 (2008) · low 69 (2023)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FIP

1in 2020

2020: 1
2007high 1 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FIP

77in 2025

2007: 12008: 52010: 122011: 42012: 42013: 412014: 342015: 122016: 32017: 62018: 32019: 132020: 122021: 352022: 572023: 1402024: 1742025: 77
2007high 174 (2024) · low 0 (2009)2025

MAUDE reports naming product code FIP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008320205
2009316700
20102—012
2011411604
20122—04
20134125041
20144126034
20152—012
20162—03
20171—06
20181—03
20192—013
20201—112
20211—035
20221—057
20233690140
20241—0174
2025488077
20260—0—

What is it indicated for?

The Isopure Complete Water Purification System each device is intended for use in Hemodialysis systems and is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. These systems are to be use in Hospitals and/or Hemodialysis facilities. Each indirect feed model has the following optional configuration: MD61 0 - Single distribution pump with a series of…

Read the full statement (195 words)

The Isopure Complete Water Purification System each device is intended for use in Hemodialysis systems and is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. These systems are to be use in Hospitals and/or Hemodialysis facilities.

Each indirect feed model has the following optional configuration:

MD61 0 - Single distribution pump with a series of ultra-filters based on the required flow of the facility. The MD610 has no Deionization capabilities.

MD620 - Dual multi-stage distribution pumps with a series of ultra-filters based on the required flow of the facility. The MD620 has no Deionization capabilities.

MD630 - Single multi-stage distribution pump with a series of ultra-filters based on the required flow of the facility. The MD630 is equipped to process a single bank of medical grade Deionization exchange tanks.

MD640 - Dual multistage distribution pumps with automatic pump alternation with a series of ultra-filters based on the required flow of the facility. The MD640 is equipped to process dual banks of medical grade Deionization exchange tanks. The MD640 is capable of operating continuously by automatically alternating between the two banks of exchange tanks.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K041163: Isopure Complete Water System for Hemodialysis, MD 400 and MD 600 Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041163

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026