510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Resorbable Interference Screw

K041274510(k) TraditionalCleared Jul 28, 200477-day reviewClass II21 CFR 888.3040Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arthrotek, Inc. · HWC Screw, Fixation, Bone

Cleared on Jul 28, 2004 as substantially equivalent to K982497 (Biomet, Inc., 1998) and 2 other predicates after a 77-day review. Its family tree reaches back 2 generations to 1995. 26 later devices descend from it. It has been recalled 5 times.

Decided
Jul 28, 2004
Review
77 days
Class
Class II
Product code
HWC
Recalls
5
Predicates
3

Its longest line of descent runs back 2 generations to K943806 (1995). 10 ancestors in all, the oldest K931519 (1994). 26 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 3 other ancestors1998: 2 other ancestors1996: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2011: 6 descendants cleared2016: 2 descendants cleared2006: 1 descendant cleared2007: 2 descendants cleared2004: 1 descendant cleared2014: 2 descendants cleared2008: 1 descendant cleared2013: 4 descendants cleared2020: 4 descendants cleared2015: 1 descendant cleared2023: 1 descendant cleared2018: 1 descendant clearedK943806 · Harpoon Suture Anchors · Biomet, Inc. · 1995K012872 · Resorbable Screw Anchor · Arthrotek, Inc. · 20011990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 6 other ancestors 1 generation forward: 6 descendants2 generations forward: 6 descendants3 generations forward: 8 descendants4 generations forward: 5 descendants5 generations forward: 1 descendant K012872 · Resorbable Screw Anchor · Arthrotek, Inc. · 2001K943806 · Harpoon Suture Anchors · Biomet, Inc. · 19950+5
  1. Root · 2 generations backK943806 Harpoon Suture AnchorsBiomet, Inc. · 1995
  2. 1 generation backK012872 Resorbable Screw AnchorArthrotek, Inc. · 2001
  3. This deviceK041274 Resorbable Interference ScrewArthrotek, Inc. · 2004
  4. Descendant · cites it directly · 9 of its ownK070389 Peek Knotless AnchorsBiomet Sports Medicine, Inc. · 2007 · Recalled
  5. Descendant · 2 generations on · 7 of its ownK110230 Rog Knotless Suture AnchorRog Sports Medicine, Inc. · 2011
  6. Descendant · cites it directly · 7 of its ownK061801 2.8 & 3.5MM Lactoscrew Suture AnchorsArthrotek, Inc. · 2006
  7. Descendant · 2 generations on · 6 of its ownK080088 Biomet Sports Medicine Anchor Devices and Ziploop ConstructsBiomet Manufacturing, Inc. · 2008 · Recalled
  8. Descendant · 3 generations on · 5 of its ownK111590 Rog Suture Anchor, Rog Knotless Suture AnchorRog Sports Medicine, Inc. · 2011

Showing 5 of 26 descendants. See all descendants

Descendants cleared per year

2004: 12006: 12007: 22008: 12011: 62013: 42014: 22015: 12016: 22018: 12020: 42023: 1
2004high 6 (2011) · low 0 (2005)2026

Most influential descendants

Showing 5 of 26 · See all
  • K070389Peek Knotless AnchorsBiomet Sports Medicine, Inc. · 2007 · cites it directly9descendants
  • K110230Rog Knotless Suture AnchorRog Sports Medicine, Inc. · 2011 · 2 generations on7descendants
  • K0618012.8 & 3.5MM Lactoscrew Suture AnchorsArthrotek, Inc. · 2006 · cites it directly7descendants
  • K080088Biomet Sports Medicine Anchor Devices and Ziploop ConstructsBiomet Manufacturing, Inc. · 2008 · 2 generations on6descendants
  • K111590Rog Suture Anchor, Rog Knotless Suture AnchorRog Sports Medicine, Inc. · 2011 · 3 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K943806Harpoon Suture AnchorsBiomet, Inc.1995—
−1K012872Resorbable Screw AnchorArthrotek, Inc.2001—
0 (this device)K041274Resorbable Interference ScrewArthrotek, Inc.2004—
+1K070389Peek Knotless AnchorsBiomet Sports Medicine, Inc.20071
+2K110230Rog Knotless Suture AnchorRog Sports Medicine, Inc.2011—
+1K0618012.8 & 3.5MM Lactoscrew Suture AnchorsArthrotek, Inc.2006—
+2K080088Biomet Sports Medicine Anchor Devices and Ziploop ConstructsBiomet Manufacturing, Inc.20088
+3K111590Rog Suture Anchor, Rog Knotless Suture AnchorRog Sports Medicine, Inc.2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
199610
199820
200130

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200610
200720
200810
201160
201340
201420
201510
201620
201810
202040
202310
FIG. 1 — Lineage of K041274, 1995–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982497Arthrotek Interference Screw
Summary page 2
…nce Screw is made of the same materials, using the same manufacturing processes, and conforming to the same standards as the predicated Artbrotek interference Screw (K982497). More sizes have been added. Non-Clinical Testing-. Mechanical testing was done on the, 3.0 wim Resorbable. Interference Sc…
Biomet, Inc.19985
*K012572Soft Tissue Screw and Washer
Summary page 1
…ivalence Is Claimed: Arthrotek Interference Screw - Arthrotek Inc. (K(982497) Soft Tissue Screw and Washer - Arthrotek Inc. (K(012572) Resorbable Screw Anchor - Arthrotek Inc, ((01 2872) Device Description: The 3 mm, 4 mm, 5 mm, 6 mm, 10…
Biomet, Inc.2001—
*K012872Resorbable Screw AnchorArthrotek, Inc.2001—

Also cleared against K982497: 5 other devices →

What descends from it? (26)

See all 26

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K070389Peek Knotless AnchorsBiomet Sports Medicine, Inc.20079
K0618012.8 & 3.5MM Lactoscrew Suture AnchorsArthrotek, Inc.20067
K042552Bio-Core Interference ScrewBiomet, Inc.20044
K103145Hammerloc TM Tibial Fixation DeviceBiomet Sports Medicine20111
K071464Inion Hexalon Biodegradable Acl/PCL ScrewInion OY20071
K151701Tenodesis screw systemNormed Medizin-Technik GmbH20160

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 28, 2018Class IIZ-1737-2018Nonconforming Material/ComponentSpecific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.
Mar 28, 2018Class IIZ-1738-2018Nonconforming Material/ComponentSpecific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.
Mar 28, 2018Class IIZ-1736-2018Nonconforming Material/ComponentSpecific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.
Mar 28, 2018Class IIZ-1735-2018Nonconforming Material/ComponentSpecific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.
Jun 3, 2015Class IIZ-1944-2015Mixed-up of materials/componentsPart Number: 905620 Lot: 824610 Gentle Threads Interference Screw 10x25mm Round Head is mixed with the Part Number: 905627 Lot: 753410 Gentle Threads Interference Screw 9x35mm Full Thread. If the incorrect screw is used various failures…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 6.4 years on median.
  • 1 of the 10 ancestors has a recall of any class.
  • Of its descendants, K070389 (Biomet Sports Medicine, Inc., 2007) has gone furthest, with 9 descendants of its own.
  • 5 other devices also cite its predicate K982497.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—

What is it indicated for?

Indications for the Resorbable Interference Screw includes use in soft tissue reattachment procedures in the shoulder, wrist/hand, snkle/foot, elbow, and knee. Specific indications include the following: Shoulder: Bankart repair, SLAP lesion repair, acromio-clavieular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair. Wrist/Hand: Scapholunate ligament reconstruction, ulnar/radial collateral…

Read the full statement (186 words)

Indications for the Resorbable Interference Screw includes use in soft tissue reattachment procedures in the shoulder, wrist/hand, snkle/foot, elbow, and knee. Specific indications include the following:

Shoulder: Bankart repair, SLAP lesion repair, acromio-clavieular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair.

Wrist/Hand: Scapholunate ligament reconstruction, ulnar/radial collateral ligament reconstruction.

Ankle/Foot: Lateral stabilization, medial stabilization, Achilles tendon repair/reconstruction, hallux valgus reconstruction, mid- and forefoot reconstruction.

Elbow: Tennis elbow repair, ulnar or radial collateral ligament reconstruction, biceps tendon reconstruction.

Knee: Extra-capsular repair, mediac collateral ligament (MCL) repair, lateral collateral ligament (LCL) repair, posterior oblique ligament repair, joint capsule closure, iliotibial band tenodesis reconstruction, patellar ligament/tendon repair, and vastus medialis obliquns (VMO) muscle advancement.

In addition to the above indications, the 7.0, 8.0, 9.0, 10.0, 11.0, and 12.0 mnm screws are indicated for the following uses:

1. To provide interference fixation of patellar bone-tendon-bone grafts in anterior cruciate ligament (ACL) reconstruction.

2. To provide interference fixation during femoral and/or tibial fixation in anterior cruciate ligament reconstruction using a soft tissue graft (senitendinosus, gracilis).

3. To provide interference fixation during posterior cmeciate ligament (PCL) reconstruction.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041274: Resorbable Interference Screw.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041274

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026