510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Hammerloc TM Tibial Fixation Device

K103145510(k) TraditionalCleared Feb 9, 2011107-day reviewClass II21 CFR 888.3040Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biomet Sports Medicine · MBI Fastener, Fixation, Nondegradable, Soft Tissue

Cleared on Feb 9, 2011 as substantially equivalent to K041274 (Arthrotek, Inc., 2004) and 3 other predicates after a 107-day review. Its family tree reaches back 3 generations to 1995. 1 later device descends from it. It has been recalled 2 times.

Decided
Feb 9, 2011
Review
107 days
Class
Class II
Product code
MBI
Recalls
2
Predicates
4

Its longest line of descent runs back 3 generations to K943806 (1995). 13 ancestors in all, the oldest K931519 (1994). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor1999: 1 other ancestor1998: 2 other ancestors2001: 3 other ancestors1996: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2020: 1 descendant clearedK943806 · Harpoon Suture Anchors · Biomet, Inc. · 1995K012872 · Resorbable Screw Anchor · Arthrotek, Inc. · 2001K041274 · Resorbable Interference Screw · Arthrotek, Inc. · 2004 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 1 other ancestor3 generations back: 6 other ancestors 1 generation forward: 1 descendant K041274 · Resorbable Interference Screw · Arthrotek, Inc. · 2004 · recalledK012872 · Resorbable Screw Anchor · Arthrotek, Inc. · 2001K943806 · Harpoon Suture Anchors · Biomet, Inc. · 19950
  1. Root · 3 generations backK943806 Harpoon Suture AnchorsBiomet, Inc. · 1995
  2. 2 generations backK012872 Resorbable Screw AnchorArthrotek, Inc. · 2001
  3. 1 generation backK041274 Resorbable Interference ScrewArthrotek, Inc. · 2004 · Recalled
  4. This deviceK103145 Hammerloc TM Tibial Fixation DeviceBiomet Sports Medicine · 2011
  5. Descendant · cites it directly · 0 of its ownK193092 TunneLoc Tibial Fixation DeviceBiomet, Inc. · 2020

Most influential descendants

Showing 1 of 1 · See all
  • K193092TunneLoc Tibial Fixation DeviceBiomet, Inc. · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K943806Harpoon Suture AnchorsBiomet, Inc.1995—
−2K012872Resorbable Screw AnchorArthrotek, Inc.2001—
−1K041274Resorbable Interference ScrewArthrotek, Inc.20045
0 (this device)K103145Hammerloc TM Tibial Fixation DeviceBiomet Sports Medicine2011—
+1K193092TunneLoc Tibial Fixation DeviceBiomet, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
199610
199820
199910
200130
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
FIG. 1 — Lineage of K103145, 1995–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBI device

  1. #1 K203268 Arthrex FiberTak Suture Anchorscore 21.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041274Resorbable Interference Screw
Summary page 1
…0) MBI Screw, Fixation, Bone (21CFR §888.3040) HWC Legally Marketed Devices To Which Substantial Equivalence Is Claimed: K041274 Resorbable Interference Screw K982497 Arthrotek Interference Screw K083607 AperFix Tm…
Arthrotek, Inc.20045
*K982497Arthrotek Interference Screw
Summary page 1
…ion, Bone (21CFR §888.3040) HWC Legally Marketed Devices To Which Substantial Equivalence Is Claimed: K041274 Resorbable Interference Screw K982497 Arthrotek Interference Screw K083607 AperFix Tm Tibial Implant with Inserter K983560 I…
Biomet, Inc.19985
*K083607Aperfix Implant with Inserter, 8, 9, 10, 11, MM X30 MM
Summary page 1
…es To Which Substantial Equivalence Is Claimed: K041274 Resorbable Interference Screw K982497 Arthrotek Interference Screw K083607 AperFix Tm Tibial Implant with Inserter K983560 Intratunnel Tibial Fixation Fastener Device Description…
Cayenne Medical, Inc.2009—
*K983560Innovasive Devices Intratunnel Tibial Fixation Fastener
Summary page 1
…Resorbable Interference Screw K982497 Arthrotek Interference Screw K083607 AperFix Tm Tibial Implant with Inserter K983560 Intratunnel Tibial Fixation Fastener Device Description: The Biomet Sports Medicine TunneLoc Tm Tibial Fixation Device is…
Innovasive Devices, Inc.1999—

Also cleared against K041274: 5 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K193092TunneLoc Tibial Fixation DeviceBiomet, Inc.20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 11, 2017Class IIZ-1071-2018Vendor change controlGamma sterilization validation testing indicates the sterilization dose for the product may not be adequate.
Oct 11, 2017Class IIZ-1072-2018Vendor change controlGamma sterilization validation testing indicates the sterilization dose for the product may not be adequate.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 6.4 years on median.
  • 2 of the 13 ancestors have a recall of any class.
  • Of its descendants, K193092 (Biomet, Inc., 2020) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K041274.

How does its category compare?

All MBI devices

MBI Fastener, Fixation, Nondegradable, Soft Tissue: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · MBI

27in 2025

2007: 162008: 102009: 192010: 172011: 302012: 242013: 312014: 232015: 232016: 172017: 172018: 162019: 312020: 352021: 182022: 192023: 522024: 222025: 272026 (partial year): 30
2007high 52 (2023) · low 10 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBI

58in 2025

2007: 80 days median2008: 69 days median2009: 82 days median2010: 76 days median2011: 75 days median2012: 94 days median2013: 100 days median2014: 101 days median2015: 85 days median2016: 94 days median2017: 70 days median2018: 91 days median2019: 90 days median2020: 83 days median2021: 107 days median2022: 64 days median2023: 57 days median2024: 30 days median2025: 58 days median2026 (partial year): 63 days median
2007high 107 (2021) · low 30 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBI

1in 2025

2007: 242008: 62009: 22010: 32011: 232012: 62013: 22017: 42018: 22020: 12023: 12024: 12025: 1
2007high 24 (2007) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MBI

3,826in 2025

2007: 2402008: 3382009: 3552010: 2912011: 3282012: 3712013: 4422014: 5042015: 5072016: 7952017: 6752018: 1,0722019: 1,3032020: 1,4542021: 1,8902022: 1,5912023: 2,5032024: 3,7792025: 3,826
2007high 3,826 (2025) · low 240 (2007)2025

MAUDE reports naming product code MBI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007168024240
200810696338
200919822355
201017763291
2011307523328
201224946371
2013311002442
2014231010504
201523850507
201617940795
201717704675
2018169121,072
2019319001,303
2020358311,454
20211810701,890
2022196401,591
2023525712,503
2024223013,779
2025275813,826
202630630—

What is it indicated for?

To provide fixation of soft-tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103145: Hammerloc TM Tibial Fixation Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103145

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026