510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Robocouch Patient Support System

K042146510(k) TraditionalCleared Sep 17, 200439-day reviewClass II21 CFR 892.5770Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Accuray Incorporated · JAI Couch, Radiation Therapy, Powered

Cleared on Sep 17, 2004 after a 39-day review. The FDA summary names its predicate without a number. 7 later devices descend from it. It has been recalled 3 times.

Decided
Sep 17, 2004
Review
39 days
Class
Class II
Product code
JAI
Recalls
3
Predicates
0

7 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2012: 1 descendant cleared2016: 1 descendant cleared2022: 1 descendant cleared2019: 1 descendant cleared2024: 1 descendant cleared2023: 1 descendant cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK042146 Robocouch Patient Support SystemAccuray Incorporated · 2004
  2. Descendant · cites it directly · 6 of its ownK122413 Patient Positioning SystemForte Automation Systems, Inc. · 2012 · Recalled
  3. Descendant · 2 generations on · 5 of its ownK160518 LEONI ORION SystemLeoni Cia Cable System · 2016 · Recalled
  4. Descendant · 3 generations on · 2 of its ownK221791 ProBeam 360° Proton Therapy SystemVarian Medical Systems, Inc. · 2022
  5. Descendant · 4 generations on · 1 of its ownK231863 ProBeam 360° Proton Therapy System v2.0 (Multiroom)Varian Medical Systems, Inc. · 2023
  6. Descendant · 4 generations on · 0 of its ownK252815 ProBeam 360 Proton Therapy System v3.0Varian Medical Systems, Inc. · 2026

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2012: 12016: 12019: 12022: 12023: 12024: 12026: 1
2012high 1 (2012) · low 0 (2013)2026

Most influential descendants

Showing 5 of 7 · See all
  • K122413Patient Positioning SystemForte Automation Systems, Inc. · 2012 · cites it directly6descendants
  • K160518LEONI ORION SystemLeoni Cia Cable System · 2016 · 2 generations on5descendants
  • K221791ProBeam 360° Proton Therapy SystemVarian Medical Systems, Inc. · 2022 · 3 generations on2descendants
  • K231863ProBeam 360° Proton Therapy System v2.0 (Multiroom)Varian Medical Systems, Inc. · 2023 · 4 generations on1descendant
  • K252815ProBeam 360 Proton Therapy System v3.0Varian Medical Systems, Inc. · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K042146Robocouch Patient Support SystemAccuray Incorporated2004—
+1K122413Patient Positioning SystemForte Automation Systems, Inc.20121
+2K160518LEONI ORION SystemLeoni Cia Cable System20162
+3K221791ProBeam 360° Proton Therapy SystemVarian Medical Systems, Inc.2022—
+4K231863ProBeam 360° Proton Therapy System v2.0 (Multiroom)Varian Medical Systems, Inc.2023—
+4K252815ProBeam 360 Proton Therapy System v3.0Varian Medical Systems, Inc.2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201610
201910
202210
202310
202410
202610
FIG. 1 — Lineage of K042146, 2004–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAI device

  1. #1 K231612 Eve Patient Positioner Systemscore 7.9 · 3 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—CyberKnife® System Patient Support Subsystem
Named without a submission number
—
*—Elekta Precise Table
Named without a submission number
—

What descends from it? (7)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122413Patient Positioning SystemForte Automation Systems, Inc.20126

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 27, 2013Class IIZ-1155-2013Nonconforming Material/ComponentA5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.
Jul 21, 2009Not classifiedZ-0853-2011Device DesignMechanical Failure-- Patient Support System's table top may unexpectedly descend several inches.
Oct 30, 2008Not classifiedZ-0856-2009Process controlProduct may not be tensioned properly, potentially causing unexpected rotation or descent.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K122413 (Forte Automation Systems, Inc., 2012) has gone furthest, with 6 descendants of its own.

How does its category compare?

All JAI devices

JAI Couch, Radiation Therapy, Powered: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAI

1in 2024

2007: 22012: 42013: 12016: 12017: 12019: 12024: 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAI

61in 2012

2012: 61 days median
2007high 61 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAI

2in 2023

2007: 12008: 22009: 12010: 22013: 22017: 12023: 2
2007high 2 (2008) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAI

1in 2024

2007: 42008: 22009: 32012: 22013: 52014: 22015: 12016: 22017: 32019: 12024: 1
2007high 5 (2013) · low 0 (2010)2025

MAUDE reports naming product code JAI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—14
20080—22
20090—13
20100—20
20110—00
201246102
20131—25
20140—02
20150—01
20161—02
20171—13
20180—00
20191—01
20200—00
20210—00
20220—00
20230—20
20241—01
20250—00
20260—0—

What is it indicated for?

The RoboCouch Patient Support System is indicated for the support and

positioning of a patient during radiosurgery and radiotherapy procedures and

other medical procedures when precise positioning is required.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042146: Robocouch Patient Support System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K042146

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026