510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Lifecare Pca Infusion System with Hospira Mednet Software

K042800510(k) TraditionalCleared Oct 18, 200410-day reviewClass II21 CFR 880.5725Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hospira, Inc. · MEA Pump, Infusion, Pca

Cleared on Oct 18, 2004 as substantially equivalent to K022203 (Abbott Laboratories, 2002) and 1 other predicate after a 10-day review. Its family tree reaches back 3 generations to 2001. 3 of the 10 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 5 times.

Decided
Oct 18, 2004
Review
10 days
Class
Class II
Product code
MEA
Recalls
5
Predicates
2

Its longest line of descent runs back 3 generations to K011442 (2001). 6 ancestors in all. 3 later devices descend from it, 3 citing it directly. 3 of the 10 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 3 other ancestors2016: 1 descendant cleared2008: 1 descendant cleared, 1 with a Class I recall12004: 1 descendant clearedK011442 · Abbott Plum a+Infusion Pump, Model 11971 · Abbott Laboratories · 2001 · Class I recallK031185 · Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module · Abbott Laboratories · 2003 · recalledK042081 · Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet Software · Hospira, Inc. · 2004 · Class I recall200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 3 descendants, 1 with a Class I recall1 K042081 · Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet Software · Hospira, Inc. · 2004 · Class I recallK031185 · Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module · Abbott Laboratories · 2003 · recalledK011442 · Abbott Plum a+Infusion Pump, Model 11971 · Abbott Laboratories · 2001 · Class I recall0
  1. Root · 3 generations backK011442 Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · Recalled (Class I)
  2. 2 generations backK031185 Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories · 2003 · Recalled
  3. 1 generation backK042081 Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3…Hospira, Inc. · 2004 · Recalled (Class I)
  4. This deviceK042800 Lifecare Pca Infusion System with Hospira Mednet SoftwareHospira, Inc. · 2004
  5. Descendant · cites it directly · 1 of its ownK043256 Hospira Lifecare Infusion System, Model 12384-04-07; Lifecare Pca…Hospira, Inc. · 2004 · Recalled
  6. Descendant · cites it directly · 0 of its ownK143612 LifeCare PCA Infusion SystemHospira, Inc. · 2016
  7. Descendant · cites it directly · 0 of its ownK083019 Gemstar SP Infusion System with Gemstar SP Infusion Suite SoftwareHospira, Inc. · 2008 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K043256Hospira Lifecare Infusion System, Model 12384-04-07; Lifecare Pca Infusion System…Hospira, Inc. · 2004 · cites it directly1descendant
  • K143612LifeCare PCA Infusion SystemHospira, Inc. · 2016 · cites it directly0descendants
  • K083019Gemstar SP Infusion System with Gemstar SP Infusion Suite SoftwareHospira, Inc. · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories2001Class I
−2K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories20034
−1K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion…Hospira, Inc.2004Class I
0 (this device)K042800Lifecare Pca Infusion System with Hospira Mednet SoftwareHospira, Inc.2004—
+1K043256Hospira Lifecare Infusion System, Model 12384-04-07; Lifecare Pca Infusion…Hospira, Inc.20044
+1K143612LifeCare PCA Infusion SystemHospira, Inc.2016—
+1K083019Gemstar SP Infusion System with Gemstar SP Infusion Suite SoftwareHospira, Inc.2008Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200230

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200811
201610
FIG. 1 — Lineage of K042800, 2001–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022203Abbott Lifecare Pca 3 Infuser, Model 12384
Summary page 1
…LifeCare PCA® Infusion System Proposed Devices: with Hospira MedNet Tm Software 5. Predicate Devices: Abbott LifeCare® PCA 3 Infuser (K022203) and Hospira Plum A+® Infusion System with Hospira MedNet TM…
Abbott Laboratories20022
*K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with…
Summary page 1
…eCare® PCA 3 Infuser (K022203) and Hospira Plum A+® Infusion System with Hospira MedNet TM Software (K042081) 6. Proposed Device Description: The LifeCare PCA® Infusion System with Hospira MedNetTM Software is an electr…
Hospira, Inc.2004Class I

Also cleared against K022203: 2 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K043256Hospira Lifecare Infusion System, Model 12384-04-07; Lifecare Pca Infusion System…Hospira, Inc.20041
K143612LifeCare PCA Infusion SystemHospira, Inc.20160
K083019Gemstar SP Infusion System with Gemstar SP Infusion Suite Software Class IHospira, Inc.20080

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 8, 2013Class IIZ-2141-2016Component design/selectionHospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw)…
Mar 8, 2013Class IIZ-2873-2016Nonconforming Material/ComponentScrew rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and…
Mar 8, 2013Class IIZ-2874-2016Nonconforming Material/ComponentScrew rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and…
Feb 27, 2013Class IIZ-1518-2015Device DesignThe infusion pump has a clear plastic door that, due to its design, allows for excessive flexing. This in turn permits unauthorized access to the vial of medication.
Aug 11, 2009Not classifiedZ-0018-2010Component design/selectionFire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet…Hospira, Inc. · 2004 · Direct predicate · ◆ on the line of descent in FIG. 14 Class I recalls
  • K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K083019Gemstar SP Infusion System with Gemstar SP Infusion Suite SoftwareHospira, Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 3 years and passes through 2 companies; a new generation was cleared every 1.3 years on median.
  • 5 of the 6 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K043256 (Hospira, Inc., 2004) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K022203.

How does its category compare?

All MEA devices

MEA Pump, Infusion, Pca: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MEA

2in 2017

2008: 12009: 12013: 12016: 12017: 2
2007high 2 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEA

Too few MEA 510(k)s in any year since 2007 to give a median.

Recalls per year · MEA

1in 2024

2008: 12010: 42011: 22014: 12017: 22022: 12024: 1
2007high 4 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MEA

3,278in 2025

2007: 812008: 2372009: 1532010: 2192011: 2452012: 4462013: 4532014: 3822015: 992016: 572017: 442018: 2562019: 1322020: 2,3312021: 3,7762022: 6,4432023: 3,9842024: 4,5592025: 3,278
2007high 6,443 (2022) · low 44 (2017)2025

MAUDE reports naming product code MEA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—081
20081—1237
20091—0153
20100—4219
20110—2245
20120—0446
20131—0453
20140—1382
20150—099
20161—057
20172—244
20180—0256
20190—0132
20200—02,331
20210—03,776
20220—16,443
20230—03,984
20240—14,559
20250—03,278
20260—0—

What is it indicated for?

The LifeCare PCA ® Infusion System with Hospira MedNetTM Software is indicated for accurate, volumetric, infusion of analgesic drugs by continuous or patient-demanded (PCA) intravenous administration. It is also indicated for short-term (less than 96 hours) continuous administration of analgesic drugs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042800: Lifecare Pca Infusion System with Hospira Mednet Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K042800

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026