Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet Software

K042081510(k) TraditionalCleared Aug 24, 200421-day reviewClass II21 CFR 880.5725Recalls 19 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hospira, Inc. · FRN Pump, Infusion

Cleared on Aug 24, 2004 as substantially equivalent to K031185 (Abbott Laboratories, 2003) and 1 other predicate after a 21-day review. Its family tree reaches back 2 generations to 2001. 12 of the 32 devices in its family, counting itself, have a Class I recall. 27 later devices descend from it. It has been recalled 19 times, including 4 Class I recalls.

Decided
Aug 24, 2004
Review
21 days
Class
Class II
Product code
FRN
Recalls
19 (4 Class I)
Predicates
2

Its longest line of descent runs back 2 generations to K011442 (2001). 4 ancestors in all. 27 later devices descend from it, 6 citing it directly. 12 of the 32 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors2014: 2 descendants cleared2008: 2 descendants cleared, 2 with a Class I recall22004: 3 descendants cleared2005: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall12016: 2 descendants cleared2011: 1 descendant cleared, 1 with a Class I recall12007: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12017: 2 descendants cleared, 1 with a Class I recall12023: 3 descendants cleared, 1 with a Class I recall12022: 1 descendant cleared2026: 2 descendants cleared2019: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2025: 3 descendants clearedK011442 · Abbott Plum a+Infusion Pump, Model 11971 · Abbott Laboratories · 2001 · Class I recallK031185 · Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module · Abbott Laboratories · 2003 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 6 descendants, 3 with a Class I recall32 generations forward: 10 descendants, 5 with a Class I recall53 generations forward: 6 descendants, 2 with a Class I recall24 generations forward: 4 descendants5 generations forward: 1 descendant K031185 · Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module · Abbott Laboratories · 2003 · recalledK011442 · Abbott Plum a+Infusion Pump, Model 11971 · Abbott Laboratories · 2001 · Class I recall0+5
  1. Root · 2 generations backK011442 Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · Recalled (Class I)
  2. 1 generation backK031185 Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories · 2003 · Recalled
  3. This deviceK042081 Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3…Hospira, Inc. · 2004
  4. Descendant · cites it directly · 12 of its ownK052052 Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3…Hospira, Inc. · 2005 · Recalled (Class I)
  5. Descendant · 2 generations on · 10 of its ownK103224 Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe…Hospira, Inc. · 2011 · Recalled (Class I)
  6. Descendant · cites it directly · 9 of its ownK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  7. Descendant · 2 generations on · 5 of its ownK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  8. Descendant · cites it directly · 4 of its ownK042609 Hospira Mednet Medication Management SuiteHospira, Inc. · 2004 · Recalled

Showing 5 of 27 descendants. See all descendants

Descendants cleared per year

2004: 32005: 12007: 12008: 22011: 12014: 22015: 12016: 22017: 22019: 12020: 12021: 12022: 12023: 32025: 32026: 2
2004high 3 (2004) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 27 · See all
  • K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5Hospira, Inc. · 2005 · cites it directly12descendants
  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · 2 generations on10descendants
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for…Hospira, Inc. · 2015 · cites it directly9descendants
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 2 generations on5descendants
  • K042609Hospira Mednet Medication Management SuiteHospira, Inc. · 2004 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories2001Class I
−1K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories20034
0 (this device)K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion…Hospira, Inc.2004—
+1K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump…Hospira, Inc.2005Class I
+2K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc.2011Class I
+1K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
+2K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
+1K042609Hospira Mednet Medication Management SuiteHospira, Inc.20041

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200430
200511
200711
200822
201111
201420
201511
201620
201721
201911
202011
202110
202210
202331
202530
202620
FIG. 1 — Lineage of K042081, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

12 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play Module
Summary page 1
…Software 5. Predicate Devices: Abbott Plum A+® Infusion Pump with HPL/RS Plug-and- Play Module (K031185) Abbott Plum A+3 MT Multichannel Infusion Pump (K021350) 6. Proposed Device Description:…
Abbott Laboratories20034
*K021350Abbott Plum a+3 Multichannel Infusion Pump, Model 12348
Summary page 1
…sion Pump with HPL/RS Plug-and- Play Module (K031185) Abbott Plum A+3 MT Multichannel Infusion Pump (K021350) 6. Proposed Device Description: The Plum A+® and A+®3 Infusion Systems with MedNet M T…
Abbott Laboratories20026

Also cleared against K031185: 1 other device →

What descends from it? (27)

See all 27

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5 Class IHospira, Inc.200512
K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for… Class IHospira, Inc.20159
K042609Hospira Mednet Medication Management SuiteHospira, Inc.20044
K141389Sapphire Infusion PumpQ Core Medical, Ltd.20143
K042800Lifecare Pca Infusion System with Hospira Mednet SoftwareHospira, Inc.20043
K081412Plum a+ Hyperbaric Infusion System with Hospira Mednet Software Class IHospira, Inc.20080

Has anything in its family been recalled? (12 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 7, 2024Class IZ-2448-2024Process controlDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than…
May 7, 2024Class IZ-2449-2024Process controlDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than…
Mar 22, 2023Class IZ-1566-2023Under Investigation by firmDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than…
Jul 2, 2014Class IZ-1073-2015Component change controlOne lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
Jun 11, 2014Class IIZ-1050-2015Vendor change controlPlum A+ infusion system pole clamp assemblies were discovered to be cracked. If the knob on the pole clamp assembly is cracked, the infuser cannot be attached to the IV pole since the knob is used to tighten and loosen the grip of the…
Feb 1, 2013Class IIZ-0289-2014Component design/selectionPlum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and…
Feb 1, 2013Class IIZ-0290-2014Component design/selectionPlum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and…
Dec 7, 2012Class IIZ-0624-2013Device DesignThe door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
Dec 7, 2012Class IIZ-0625-2013Device DesignThe door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
Oct 26, 2012Class IIZ-0483-2013Device DesignThe volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a…
Feb 5, 2012Class IIZ-0209-2014Nonconforming Material/ComponentHospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186…
Sep 16, 2011Not classifiedZ-0009-2012Software designHospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This…
Sep 16, 2011Not classifiedZ-0010-2012Software designHospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This…
Sep 16, 2011Not classifiedZ-3284-2011Nonconforming Material/ComponentHospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged…
Sep 16, 2011Not classifiedZ-3285-2011Nonconforming Material/ComponentHospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged…
Feb 14, 2011Not classifiedZ-1645-2011Nonconforming Material/ComponentHospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.
Feb 14, 2011Not classifiedZ-1644-2011Nonconforming Material/ComponentHospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.
Aug 11, 2009Not classifiedZ-0027-2010Component design/selectionFire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Aug 11, 2009Not classifiedZ-0028-2010Component design/selectionFire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K081412Plum a+ Hyperbaric Infusion System with Hospira Mednet SoftwareHospira, Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5Hospira, Inc. · 2005 · Cites it directly · ◆ in the 2005 descendants bar of FIG. 11 Class I recall
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K083019Gemstar SP Infusion System with Gemstar SP Infusion Suite SoftwareHospira, Inc. · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 14 Class I recalls
  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K070398Hospira Plum a+ Infusion Pump, Model 12391Hospira, Inc. · 2007 · 2 generations on · ◆ in the 2007 descendants bar of FIG. 14 Class I recalls
  • K192860Sapphire Infusion Pump, Sapphire administration setsQ Core Medical, Ltd. · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls

What does this lineage say?

  • The line of descent spans 3 years and passes through 2 companies; a new generation was cleared every 1.9 years on median.
  • 3 of the 4 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K052052 (Hospira, Inc., 2005) has gone furthest, with 12 descendants of its own.
  • 1 other device also cites its predicate K031185.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

Plum A+® and Plum A+®3 Infusion Systems with Hospira MedNetTM Software are indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042081: Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K042081

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026