510(k) Report

510(k) · Cleared

Recalled · 2005

Onetouch Ultra, Induo and Ultrasmart Blood Glucose Monitoring System

K043197510(k) TraditionalCleared May 20, 2005183-day reviewClass II21 CFR 862.1345Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Lifescan, Inc. · CGA Glucose Oxidase, Glucose

Cleared on May 20, 2005 after a 183-day review. The FDA summary names its predicate without a number. 6 later devices descend from it. It has been recalled 3 times.

Decided
May 20, 2005
Review
183 days
Class
Class II
Product code
CGA
Recalls
3
Predicates
0

6 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2008: 2 descendants cleared2006: 1 descendant cleared2010: 1 descendant cleared2013: 1 descendant cleared2009: 1 descendant cleared20052010201520202025
  1. This deviceK043197 Onetouch Ultra, Induo and Ultrasmart Blood Glucose Monitoring SystemLifescan, Inc. · 2005
  2. Descendant · cites it directly · 2 of its ownK080639 Symphony Glucose Management SystemAnimas Corp. · 2008 · Recalled
  3. Descendant · cites it directly · 2 of its ownK062555 Ebsensor Blood Glucose Monitoring System, Model EB-GVisgeneer, Inc. · 2006
  4. Descendant · 2 generations on · 0 of its ownK122181 Ebchek Blood Glucose Monitoring SystemVisgeneer, Inc. · 2013
  5. Descendant · 2 generations on · 0 of its ownK091765 Ebcarryon Blood Glucose Monitoring SystemVisgeneer, Inc. · 2010
  6. Descendant · 2 generations on · 0 of its ownK083225 Freestyle Aviator Insulin Delivery and Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2009

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2006: 12008: 22009: 12010: 12013: 1
2006high 2 (2008) · low 0 (2007)2026

Most influential descendants

Showing 5 of 6 · See all
  • K080639Symphony Glucose Management SystemAnimas Corp. · 2008 · cites it directly2descendants
  • K062555Ebsensor Blood Glucose Monitoring System, Model EB-GVisgeneer, Inc. · 2006 · cites it directly2descendants
  • K122181Ebchek Blood Glucose Monitoring SystemVisgeneer, Inc. · 2013 · 2 generations on0descendants
  • K091765Ebcarryon Blood Glucose Monitoring SystemVisgeneer, Inc. · 2010 · 2 generations on0descendants
  • K083225Freestyle Aviator Insulin Delivery and Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2009 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K043197Onetouch Ultra, Induo and Ultrasmart Blood Glucose Monitoring SystemLifescan, Inc.2005—
+1K080639Symphony Glucose Management SystemAnimas Corp.20087
+1K062555Ebsensor Blood Glucose Monitoring System, Model EB-GVisgeneer, Inc.2006—
+2K122181Ebchek Blood Glucose Monitoring SystemVisgeneer, Inc.2013—
+2K091765Ebcarryon Blood Glucose Monitoring SystemVisgeneer, Inc.2010—
+2K083225Freestyle Aviator Insulin Delivery and Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.2009—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200820
200910
201010
201310
FIG. 1 — Lineage of K043197, 2005–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGA device

  1. #1 K252488 ABL90 FLEX PLUS Systemscore 12.8 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—OneTouch® Ultra® Test Strip
Named without a submission number
—
*—OneTouch® Ultra® Blood Glucose Monitoring System
Named without a submission number
—
*—OneTouch® InDuo® Blood Glucose Monitoring System
Named without a submission number
—
*—OneTouch® UltraSmart® Blood Glucose Monitoring System
Named without a submission number
—

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K080639Symphony Glucose Management SystemAnimas Corp.20082
K062555Ebsensor Blood Glucose Monitoring System, Model EB-GVisgeneer, Inc.20062

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 1, 2005Not classifiedZ-0435-06OtherProduct may be defective and may give low inaccurate glucose results.
Oct 19, 2005Not classifiedZ-0273-06OtherVials of OneTouch Ultra Test Strips marked with quantities of 25 per vial actually contain only 10 strips.
Oct 10, 2005Not classifiedZ-0272-06OtherUsers may inadvertently change units of measurement on demonstration kits of the OneTouch Ultra Blood Glucose Meters

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K080639 (Animas Corp., 2008) has gone furthest, with 2 descendants of its own.

How does its category compare?

All CGA devices

CGA Glucose Oxidase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CGA

2in 2024

2007: 72008: 62009: 102010: 82011: 92012: 102013: 32014: 52015: 42016: 42017: 22018: 22019: 22020: 22021: 12022: 12023: 42024: 22026 (partial year): 1
2007high 10 (2009) · low 0 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGA

249in 2023

2007: 122 days median2008: 56 days median2009: 155 days median2010: 122 days median2011: 217 days median2012: 123 days median2013: 441 days median2014: 265 days median2015: 150 days median2016: 157 days median2023: 249 days median
2007high 441 (2013) · low 56 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CGA

3in 2023

2007: 22008: 22009: 42010: 62011: 32013: 32014: 22015: 6, 2 Class I2016: 22019: 12023: 3
2007high 6 (2010) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · CGA

59in 2025

2007: 292008: 372009: 752010: 1722011: 4232012: 2332013: 1722014: 1222015: 1552016: 362017: 922018: 692019: 1162020: 682021: 482022: 2722023: 1,1412024: 2662025: 59
2007high 1,141 (2023) · low 29 (2007)2025

MAUDE reports naming product code CGA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077122229
2008656237
200910155475
201081226172
201192173423
2012101230233
201334413172
201452652122
201541506 (2)155
20164157236
20172—092
20182—069
20192—1116
20202—068
20211—048
20221—0272
2023424931,141
20242—0266
20250—059
20261—0—

What is it indicated for?

The OneTouch® Ultra® Blood Glucose Monitoring System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The OneTouch® Ultra® System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The OneTouch® Ultra® Blood Glucose Monitoring System is specifically…

Read the full statement (295 words)

The OneTouch® Ultra® Blood Glucose Monitoring System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The OneTouch® Ultra® System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The OneTouch® Ultra® Blood Glucose Monitoring System is specifically indicated for use on the finger, arm or palm.

The OneTouch® InDuo® System is intended to be used for quantitative measurement the body of glucose in fresh capillary whole blood. The OneTouch® Ultra® System is intended for use outside (in vitro diagnostic use) by people with diabetes at home and healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The OneTouch® InDuo® Blood Glucose Meter also functions as the cap for the InDuo® Insulin Doser. The two devices fit together to form a single unit for user convenience. The OneTouch InDuo® Blood Glucose Monitoring System is specifically indicated for use on the finger, arm or palm.

The OneTouch® UltraSmart® System is intended to be used for quantitative measurement of glucose in fresh capillary whole blood. The OneTouch® Ultra® System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The OneTouch UltraSmart® Blood Glucose Monitoring System provides the user with electronic logbook functions that store data such as insulin and oral medicationa data port that enables the user to download electronic data to a personal computer. The OneTouch® UltraSmart® Smart Blood Glucose Monitoring System is specifically indicated for use on the finger, arm or palm.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K043197: Onetouch Ultra, Induo and Ultrasmart Blood Glucose Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K043197

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026