Symphony Glucose Management System

K080639510(k) TraditionalCleared Jun 24, 2008110-day reviewClass II21 CFR 880.5725Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Animas Corp. · LZG Pump, Infusion, Insulin

Cleared on Jun 24, 2008 as substantially equivalent to K042873 (Animas Corp., 2004) and 6 other predicates after a 110-day review. Its family tree reaches back 2 generations to 1995. 1 of the 12 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 7 times.

Decided
Jun 24, 2008
Review
110 days
Class
Class II
Product code
LZG
Recalls
7
Predicates
7

Its longest line of descent runs back 2 generations to K950727 (1995). 9 ancestors in all. 2 later devices descend from it, 2 citing it directly. 1 of the 12 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2005: 2 other ancestors2004: 2 other ancestors, 1 with a Class I recall12003: 1 other ancestor2000: 1 other ancestor2009: 1 descendant cleared2008: 1 descendant clearedK950727 · One Touch Profile Complete Diabetes Tracking System · Lifescan, Inc. · 1995K021819 · One Touch Ultrasmart Blood Glucose Monitoring System · Lifescan, Inc. · 2002 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 6 other ancestors, 1 with a Class I recall12 generations back: 1 other ancestor 1 generation forward: 2 descendants K021819 · One Touch Ultrasmart Blood Glucose Monitoring System · Lifescan, Inc. · 2002 · recalledK950727 · One Touch Profile Complete Diabetes Tracking System · Lifescan, Inc. · 19950
  1. Root · 2 generations backK950727 One Touch Profile Complete Diabetes Tracking SystemLifescan, Inc. · 1995
  2. 1 generation backK021819 One Touch Ultrasmart Blood Glucose Monitoring SystemLifescan, Inc. · 2002 · Recalled
  3. This deviceK080639 Symphony Glucose Management SystemAnimas Corp. · 2008
  4. Descendant · cites it directly · 0 of its ownK083225 Freestyle Aviator Insulin Delivery and Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2009
  5. Descendant · cites it directly · 0 of its ownK082590 Symphony Meter RemoteLifescan, Inc. · 2008

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K083225Freestyle Aviator Insulin Delivery and Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2009 · cites it directly0descendants
  • K082590Symphony Meter RemoteLifescan, Inc. · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K950727One Touch Profile Complete Diabetes Tracking SystemLifescan, Inc.1995—
−1K021819One Touch Ultrasmart Blood Glucose Monitoring SystemLifescan, Inc.20021
0 (this device)K080639Symphony Glucose Management SystemAnimas Corp.2008—
+1K083225Freestyle Aviator Insulin Delivery and Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.2009—
+1K082590Symphony Meter RemoteLifescan, Inc.2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200310
200421
200520
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200910
FIG. 1 — Lineage of K080639, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZG device

  1. #1 K253534 Pivot Insulin Delivery Systemscore 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (7)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042873Animas Model IR1250 Insulin Infusion Pump
Summary page 1
…- Class II, One®Touch Ultra Test Strips, per 21 CFR § 862.1345 PREDICATE DEVICES: Animas® IR 1250 insulin pump (K042873) OneTouch® UltraSmart® Blood Glucose Monitoring System (K021819, K043197)…
Animas Corp.2004Class I
*K021819One Touch Ultrasmart Blood Glucose Monitoring System
Summary page 1
…DICATE DEVICES: Animas® IR 1250 insulin pump (K042873) OneTouch® UltraSmart® Blood Glucose Monitoring System (K021819, K043197) OneTouch® Ultra®2 Blood Glucose Monitoring System (K053529)…
Lifescan, Inc.20021
*K043197Onetouch Ultra, Induo and Ultrasmart Blood Glucose Monitoring System
Summary page 1
…VICES: Animas® IR 1250 insulin pump (K042873) OneTouch® UltraSmart® Blood Glucose Monitoring System (K021819, K043197) OneTouch® Ultra®2 Blood Glucose Monitoring System (K053529)…
Lifescan, Inc.20053
*K053529Onetouch Ultra 2 Blood Glucose Monitoring System
Summary page 1
…Monitoring System (K021819, K043197) OneTouch® Ultra®2 Blood Glucose Monitoring System (K053529) Medtronic MiniMed Paradigm Model 512 Insulin Pump and Paradigm Link Gluco…
Lifescan, Inc.2006—
*K030531Medtronic Minimed Paradigm Model 512 Insulin Pump and the BD Paradigm Link Glucose Meter
Summary page 1
…System (K053529) Medtronic MiniMed Paradigm Model 512 Insulin Pump and Paradigm Link Glucose Meter (K03053 1, K040603) Insulet OmniPod iXL-II Diabetes Management…
Minimed, Inc.20033
*K040603Paradigm Link Blood Glucose Meter
Summary page 1
…Medtronic MiniMed Paradigm Model 512 Insulin Pump and Paradigm Link Glucose Meter (K03053 1, K040603) Insulet OmniPod iXL-II Diabetes Management System (K042792) Animas Corpor…
Becton, Dickinson & CO2004—
*K042792Ixl-II Diabetes Management System
Summary page 1
…nk Glucose Meter (K03053 1, K040603) Insulet OmniPod iXL-II Diabetes Management System (K042792) Animas Corporation Section 5- 510(k) Summary Page 1 of 3 Symphony System…
Insulet Corporation2005—

Also cleared against K042873: 2 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083225Freestyle Aviator Insulin Delivery and Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.20090
K082590Symphony Meter RemoteLifescan, Inc.20080

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 1, 2012Not classifiedZ-1376-2012Software designThere is an error in the pump software where the OneTouch(R) Ping(R) insulin pump did not allow the user to save changes to the time and date on the pump if those changes were made on February 29, 2012 (Leap Day). Users who did not…
Mar 18, 2010Not classifiedZ-2394-2010Equipment maintenancesensor could be damaged
Dec 18, 2009Not classifiedZ-1560-2010Software designSoftware malfunction which disallows users to download, view and print information from their pump (including blood glucose, insulin delivery, carbohydrate and pump events). Additionally, users are unable to customize the built-in food…
Oct 27, 2008Not classifiedZ-0955-2009Software designincorrect dosing recommendation
Aug 4, 2008Not classifiedZ-0290-2009Nonconforming Material/ComponentUnintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure…
Aug 4, 2008Not classifiedZ-0291-2009Nonconforming Material/ComponentUnintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure…
Aug 4, 2008Not classifiedZ-0283-2009Software designdisplay old, inaccurate values for bolus amount delivered and amount of planned bolus totals

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K042873Animas Model IR1250 Insulin Infusion PumpAnimas Corp. · 2004 · Direct predicate · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 13 years and passes through 2 companies; a new generation was cleared every 7.2 years on median.
  • 4 of the 9 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K083225 (Abbott Diabetes Care, Inc., 2009) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K042873.

How does its category compare?

All LZG devices

LZG Pump, Infusion, Insulin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · LZG

2in 2024

2007: 32008: 62009: 22010: 22011: 42012: 32013: 22014: 42015: 12016: 42017: 12018: 12019: 22021: 12022: 12023: 12024: 22026 (partial year): 2
2007high 6 (2008) · low 0 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZG

116in 2016

2007: 112 days median2008: 129 days median2011: 189 days median2012: 73 days median2014: 123 days median2016: 116 days median
2007high 189 (2011) · low 73 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZG

1in 2022

2007: 12008: 132009: 22010: 42011: 22012: 42013: 3, 1 Class I2014: 2, 1 Class I2015: 2, 2 Class I2018: 3, 2 Class I2019: 52022: 1, 1 Class I2026 (partial year): 2, 2 Class I
2007high 13 (2008) · low 0 (2016)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · LZG

12,004in 2025

2007: 5,1102008: 5,7112009: 9,4712010: 10,1752011: 12,8582012: 18,3672013: 30,0592014: 49,1882015: 69,1362016: 56,5812017: 44,4722018: 32,6522019: 29,0272020: 32,5882021: 31,5222022: 28,2252023: 21,9722024: 18,2282025: 12,004
2007high 69,136 (2015) · low 5,110 (2007)2025

MAUDE reports naming product code LZG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007311215,110
20086129135,711
20092—29,471
20102—410,175
20114189212,858
2012373418,367
20132—3 (1)30,059
201441232 (1)49,188
20151—2 (2)69,136
20164116056,581
20171—044,472
20181—3 (2)32,652
20192—529,027
20200—032,588
20211—031,522
20221—1 (1)28,225
20231—021,972
20242—018,228
20250—012,004
20262—2 (2)—

What is it indicated for?

The Symphony System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood. The Symphony Pump is indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes. The Symphony Meter Remote is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood, and as a wireless (RF) remote control to…

Read the full statement (144 words)

The Symphony System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood.

The Symphony Pump is indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes.

The Symphony Meter Remote is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood, and as a wireless (RF) remote control to deliver insulin from the Symphony Pump. The Symphony Meter Remote is intended for use for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home and by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The Symphony Meter Remote is specifically indicated for use on the finger, forearm or palm. It should not be used for the diagnosis of diabetes or testing of newborns.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080639: Symphony Glucose Management System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K080639

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026