510(k) · Cleared
Recalled · Class II · 2025HD11 Diagnostic Ultrasound System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Ultrasound, Inc. · IYN System, Imaging, Pulsed Doppler, Ultrasonic
Cleared on Jan 6, 2005 after a 15-day review. The FDA summary names its predicate without a number. 2 of the 131 devices in its family, counting itself, have a Class I recall. 130 later devices descend from it. It has been recalled 36 times.
- Decided
- Jan 6, 2005
- Review
- 15 days
- Class
- Class II
- Product code
- IYN
- Recalls
- 36
- Predicates
- 0
130 later devices descend from it, 4 citing it directly. 2 of the 131 devices in this family, counting this one, have a Class I recall.
- This deviceK043535 HD11 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2005
- Descendant · cites it directly · 91 of its ownK120321 Clearvue 350/550 Diagnostic Ultrasound SystemPhilips Health Care · 2012 · Recalled
- Descendant · 2 generations on · 87 of its ownK133833 Nuvis Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2014 · Recalled
- Descendant · 3 generations on · 79 of its ownK152899 Philips Lumify Diagnostic Ultrasound SystemPhilips Healthcare, Inc. · 2015 · Recalled
- Descendant · cites it directly · 57 of its ownK081802 CX50 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2008 · Recalled
- Descendant · cites it directly · 56 of its ownK111513 CX50 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2011 · Recalled
Showing 5 of 130 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 130 · See all- K120321Clearvue 350/550 Diagnostic Ultrasound SystemPhilips Health Care · 2012 · cites it directly91descendants
- K133833Nuvis Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2014 · 2 generations on87descendants
- K152899Philips Lumify Diagnostic Ultrasound SystemPhilips Healthcare, Inc. · 2015 · 3 generations on79descendants
- K081802CX50 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2008 · cites it directly57descendants
- K111513CX50 Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2011 · cites it directly56descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K043535 | HD11 Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2005 | — |
| +1 | K120321 | Clearvue 350/550 Diagnostic Ultrasound System | Philips Health Care | 2012 | 8 |
| +2 | K133833 | Nuvis Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2014 | 3 |
| +3 | K152899 | Philips Lumify Diagnostic Ultrasound System | Philips Healthcare, Inc. | 2015 | 4 |
| +1 | K081802 | CX50 Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2008 | 4 |
| +1 | K111513 | CX50 Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2011 | 10 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 2 | 0 |
| 2011 | 1 | 0 |
| 2012 | 2 | 0 |
| 2014 | 4 | 0 |
| 2015 | 8 | 0 |
| 2016 | 10 | 0 |
| 2017 | 13 | 0 |
| 2018 | 5 | 0 |
| 2019 | 5 | 0 |
| 2020 | 8 | 0 |
| 2021 | 12 | 0 |
| 2022 | 10 | 0 |
| 2023 | 8 | 0 |
| 2024 | 10 | 0 |
| 2025 | 17 | 2 |
| 2026 | 15 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYN device
- #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Philips M2540A EnVisor Named without a submission number | |
| * | — | Philips HDI-5000 series Ultrasound systems Named without a submission number |
What descends from it? (130)
See all 1304 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 5, 2025 | Class III | Z-1622-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Sep 5, 2025 | Class III | Z-1618-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Sep 5, 2025 | Class III | Z-1619-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Sep 5, 2025 | Class III | Z-1620-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Sep 5, 2025 | Class III | Z-1624-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2390-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2393-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2389-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2368-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2377-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2372-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2385-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2378-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2396-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2346-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2381-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2365-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2395-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2376-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2397-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2359-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2358-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2357-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2347-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2353-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2363-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| May 28, 2025 | Class II | Z-2060-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2048-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2050-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2056-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2051-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2054-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2059-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| Jun 30, 2023 | Class II | Z-2517-2023 | Process control | Transducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock if the built-in… |
| Aug 30, 2019 | Class II | Z-1842-2020 | Process control | The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of… |
| Apr 25, 2007 | Not classified | Z-0987-2007 | Other | Potential for erroneous measurements which may lead to incorrect diagnosis based upon the presence of an inaccurate, undetectable error in clinical measurement or calculation. If the user exits an imaging mode while Active Annotation is… |
The 5 latest of 36 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
What does this lineage say?
- Of its descendants, K120321 (Philips Health Care, 2012) has gone furthest, with 91 descendants of its own.
How does its category compare?
All IYN devicesIYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · IYN
57in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYN
124in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYN
10in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · IYN
610in 2025
MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 13 | 26 | 16 | 3 |
| 2008 | 25 | 18 | 12 | 2 |
| 2009 | 18 | 29 | 8 | 9 |
| 2010 | 29 | 67 | 10 | 17 |
| 2011 | 39 | 46 | 8 | 18 |
| 2012 | 43 | 37 | 3 | 20 |
| 2013 | 51 | 60 | 8 | 18 |
| 2014 | 56 | 39 | 2 | 48 |
| 2015 | 54 | 55 | 3 | 116 |
| 2016 | 60 | 56 | 2 | 145 |
| 2017 | 64 | 83 | 0 | 146 |
| 2018 | 59 | 85 | 7 | 85 |
| 2019 | 47 | 79 | 0 | 59 |
| 2020 | 57 | 72 | 4 | 67 |
| 2021 | 46 | 105 | 8 | 213 |
| 2022 | 53 | 105 | 10 | 186 |
| 2023 | 44 | 115 | 4 | 229 |
| 2024 | 50 | 119 | 1 | 266 |
| 2025 | 57 | 124 | 10 | 610 |
| 2026 | 44 | 107 | 2 (1) | — |
What is it indicated for?
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K043535 · printed Nov 4, 2026
Cite this page
510(k) Report. “K043535: HD11 Diagnostic Ultrasound System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K043535