510(k) Report

510(k) · Cleared

Cleared

Harvard 2 Syringe Pump, Model 2001-001

K050107510(k) SpecialCleared Feb 11, 200524-day reviewClass II21 CFR 880.5725

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Harvard Clinical Technology · FRN Pump, Infusion

Cleared on Feb 11, 2005 after a 24-day review. The FDA summary names its predicate without a number. 7 later devices descend from it.

Decided
Feb 11, 2005
Review
24 days
Class
Class II
Product code
FRN
Recalls
None
Predicates
0

7 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared2011: 2 descendants cleared2012: 2 descendants cleared2013: 1 descendant cleared2015: 1 descendant cleared20052010201520202025
  1. This deviceK050107 Harvard 2 Syringe Pump, Model 2001-001Harvard Clinical Technology · 2005
  2. Descendant · cites it directly · 6 of its ownK053554 CompufloMilestone Scientific, Inc. · 2006
  3. Descendant · 2 generations on · 5 of its ownK101194 Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011
  4. Descendant · 3 generations on · 4 of its ownK112067 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011
  5. Descendant · 4 generations on · 3 of its ownK120830 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2012
  6. Descendant · 5 generations on · 2 of its ownK122215 Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2012

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2006: 12011: 22012: 22013: 12015: 1
2006high 2 (2011) · low 0 (2007)2026

Most influential descendants

Showing 5 of 7 · See all
  • K053554CompufloMilestone Scientific, Inc. · 2006 · cites it directly6descendants
  • K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc. · 2011 · 2 generations on5descendants
  • K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2011 · 3 generations on4descendants
  • K120830Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2012 · 4 generations on3descendants
  • K122215Navigator Delivery System (Navigator DS)Carticept Medical, Inc. · 2012 · 5 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K050107Harvard 2 Syringe Pump, Model 2001-001Harvard Clinical Technology2005—
+1K053554CompufloMilestone Scientific, Inc.2006—
+2K101194Navigator Delivery System (or Navigator DS) Model: Nav-010Carticept Medical, Inc.2011—
+3K112067Navigator Delivery System (Navigator DS)Carticept Medical, Inc.2011—
+4K120830Navigator Delivery System (Navigator DS)Carticept Medical, Inc.2012—
+5K122215Navigator Delivery System (Navigator DS)Carticept Medical, Inc.2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
201120
201220
201310
201510
FIG. 1 — Lineage of K050107, 2005–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Harvard 1 Syringe Pump
Named without a submission number
—
*—Harvard 2 Syringe Pump
Named without a submission number
—

What descends from it? (7)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053554CompufloMilestone Scientific, Inc.20066

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K053554 (Milestone Scientific, Inc., 2006) has gone furthest, with 6 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Harvard 2 provides intravenous, intra-arterial, or epidural delivery of drugs or other parenteral fluids when administered by health care professionals such as physicians and nurses.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050107: Harvard 2 Syringe Pump, Model 2001-001.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K050107

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026