510(k) Report

510(k) · Cleared

Cleared

MRidium 3870 MRI Infusion Pump System (3870)

K242752510(k) TraditionalCleared May 28, 2025258-day reviewClass II21 CFR 880.5725

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Iradimed Corporation · FRN Pump, Infusion

Cleared on May 28, 2025 as substantially equivalent to K143369 (Iradimed Corporation, 2016) after a 258-day review. Its family tree reaches back 5 generations to 2000. 2 of the 11 devices in its family, counting itself, have a Class I recall.

Decided
May 28, 2025
Review
258 days
Class
Class II
Product code
FRN
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K002211 (2000). 10 ancestors in all. No later device cites it as a predicate yet. 2 of the 11 devices in this family, counting this one, have a Class I recall.

  1. Root · 5 generations backK002211 Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric…Baxter Healthcare Corp · 2000 · Recalled
  2. 4 generations backK012383 Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE)…Alaris Medical Systems, Inc. · 2001 · Recalled (Class I)
  3. 3 generations backK050301 Mridium 3850 Mri Pump SystemIradimed Corporation · 2005 · Recalled
  4. 2 generations backK090087 Mridium 3860 Mri Infusion Pump/Monitoring SystemIradimed Corporation · 2009 · Recalled (Class I)
  5. 1 generation backK143369 MRidium 3860+ MRI infusion Pump/ Monitoring SystemIradimed Corporation · 2016 · Recalled
  6. This deviceK242752 MRidium 3870 MRI Infusion Pump System (3870)Iradimed Corporation · 2025

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K002211Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric Infusion Pump…Baxter Healthcare Corp20004
−4K012383Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem…Alaris Medical Systems, Inc.2001Class I
−3K050301Mridium 3850 Mri Pump SystemIradimed Corporation20052
−2K090087Mridium 3860 Mri Infusion Pump/Monitoring SystemIradimed Corporation2009Class I
−1K143369MRidium 3860+ MRI infusion Pump/ Monitoring SystemIradimed Corporation20161
0 (this device)K242752MRidium 3870 MRI Infusion Pump System (3870)Iradimed Corporation2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200210
200310
200620
FIG. 1 — Lineage of K242752, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K143369MRidium 3860+ MRI infusion Pump/ Monitoring System
Summary page 5
…MED: Predicate Device: MRidium 3860+ MRI Infusion Pump/Monitoring System Manufacturer: Iradimed Corporation 510(k) #: K143369 Clearance date: 12/15/2016 DEVICE DESCRIPTION: The MRidium 3870 MRI Infusion Pump System is indicated for genera…
Iradimed Corporation20161

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K090087Mridium 3860 Mri Infusion Pump/Monitoring SystemIradimed Corporation · 2009 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K012383Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem Iii Infusion (Msiii)…Alaris Medical Systems, Inc. · 2001 · 3 generations back · ◆ on the line of descent in FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 25 years and passes through 3 companies; a new generation was cleared every 4.0 years on median.
  • 8 of the 10 ancestors have a recall of any class; 2 have a Class I recall.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Iradimed Corporation's MRidium 3870 MRI Infusion Pump System is indicated for general hospital or clinical use by medical professionals whenever it is required to infuse patients with intravenous fluids where a large magnetic field could be present such as Magnetic Resonance Imaging (MRI). The MRidium 3870 MRI Infusion pump can also be used during transport within the clinical environment as part of the periprocedural care before or after the MRI…

Read the full statement (242 words)

The Iradimed Corporation's MRidium 3870 MRI Infusion Pump System is indicated for general hospital or clinical use by medical professionals whenever it is required to infuse patients with intravenous fluids where a large magnetic field could be present such as Magnetic Resonance Imaging (MRI). The MRidium 3870 MRI Infusion pump can also be used during transport within the clinical environment as part of the periprocedural care before or after the MRI exam. The MRidium 3870 MRI Infusion Pump can be used with up to a 3.0 Tesla MRI system when placed as close as clinically possible to the MR system, without being placed within the magnet’s bore, up to a 15,000 Gauss magnetic field line.

The pump and infusion sets are designed to deliver saline and IV medication. The system is useful in the administration of fluids requiring precisely controlled infusion rates. The system can operate in primary, loading, bolus, or KVO delivery mode.

The dedicated Iradimed 1056, 1058, and 1059 Infusion Sets for the MRidium 3870 MRI Infusion Pump are intended for single patient use up to 6 hours.

The device is intended for Adult and Pediatric patients, including Neonates. Careful clinical considerations involving drug and dosing should always be applied especially with regard to neonatal subpopulations (i.e., low and extremely low birth weight).

This device is available for sale only upon the order of a physician or other related licensed medical professional, and not intended for any home use applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K242752: MRidium 3870 MRI Infusion Pump System (3870).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K242752

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026