510(k) Report

510(k) · Cleared

Cleared

Spor-Test PA Biological Indicator Kit

K050664510(k) SpecialCleared Apr 7, 200523-day reviewClass II21 CFR 880.2800

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Getinge USA, Inc. · FRC Indicator, Biological Sterilization Process

Cleared on Apr 7, 2005 as substantially equivalent to K020205 (Getinge/Castle, Inc., 2002) after a 23-day review. Its family tree reaches back 3 generations to 1980.

Decided
Apr 7, 2005
Review
23 days
Class
Class II
Product code
FRC
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K792649 (1980). 5 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK792649 Unispore-Biological IndicatorsCastle Co. · 1980
  2. 2 generations backK960777 MDT Unispore and Spor-Test BiologicalindicatorMdt Corp., Inc. · 1996
  3. 1 generation backK020205 Castle Peracetic Acid Biological Indicator KitGetinge/Castle, Inc. · 2002
  4. This deviceK050664 Spor-Test PA Biological Indicator KitGetinge USA, Inc. · 2005
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K792649Unispore-Biological IndicatorsCastle Co.1980—
−2K960777MDT Unispore and Spor-Test BiologicalindicatorMdt Corp., Inc.1996—
−1K020205Castle Peracetic Acid Biological Indicator KitGetinge/Castle, Inc.2002—
0 (this device)K050664Spor-Test PA Biological Indicator KitGetinge USA, Inc.2005—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198020
FIG. 1 — Lineage of K050664, 1980–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRC device

  1. #1 K250044 Celerity 5 HP Biological Indicator (LCB052)…score 14.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020205Castle Peracetic Acid Biological Indicator Kit
Summary page 1
…POR-TEST PA Biological Indicator Kit's manufacturing material, manufacturing Kit (K020205). The storage conditions are the same as the Castle SPOR-TEST PA Biological Indicator…
Getinge/Castle, Inc.2002—

Also cleared against K020205: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 6.1 years on median.
  • None of the 5 ancestors has a recall.
  • 1 other device also cites its predicate K020205.

How does its category compare?

All FRC devices

FRC Indicator, Biological Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRC

6in 2025

2007: 32008: 12009: 62010: 42011: 62012: 92013: 42014: 132015: 32016: 22017: 102018: 182019: 52020: 102021: 52022: 52023: 62024: 82025: 62026 (partial year): 1
2007high 18 (2018) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRC

29in 2025

2007: 83 days median2009: 183 days median2010: 165 days median2011: 118 days median2012: 178 days median2013: 260 days median2014: 174 days median2015: 130 days median2017: 130 days median2018: 97 days median2019: 88 days median2020: 150 days median2021: 107 days median2022: 193 days median2023: 190 days median2024: 105 days median2025: 29 days median
2007high 260 (2013) · low 29 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRC

8so far in 2026, the most since 2007

2009: 12012: 22013: 22022: 42023: 12026 (partial year): 8
2007high 8 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRC

9in 2025

2007: 1002008: 602009: 452010: 572011: 372012: 462013: 392014: 512015: 1822016: 4822017: 1822018: 222019: 302020: 162021: 102022: 92023: 202024: 222025: 9
2007high 482 (2016) · low 9 (2022)2025

MAUDE reports naming product code FRC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073830100
20081—060
20096183145
20104165057
20116118037
20129178246
20134260239
201413174051
201531300182
20162—0482
2017101300182
20181897022
2019588030
202010150016
20215107010
2022519349
20236190120
20248105022
202562909
20261—8—

What is it indicated for?

The SPOR-TEST PA Biological Indicator Kit is only intended to monitor the Steris System I liquid chemical sterilization system, with the Steris 20 sterilant. Use in monitoring other sterilization processes is contraindicated.

SPOR-TEST PA Biological Indicators are qualified using Getinge Culture Media. When tested at 1,000 ppm peracetic acid, 50°C, the SPOR-TEST PA Biological Indicator will survive at 41 seconds and will be killed at 6 minutes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050664: Spor-Test PA Biological Indicator Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K050664

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026