510(k) Report

510(k) · Cleared

Recalled · 2009

Lightwave Integrated Electrode Ablator and Light Wave Integrated Electrode Suction Ablator

K050923510(k) TraditionalCleared May 18, 200535-day reviewClass II21 CFR 878.4400Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Conmed Linvatec · JOS Electrode, Electrosurgical

Cleared on May 18, 2005 as substantially equivalent to K993885 (Linvatec Corp., 2000) and 4 other predicates after a 35-day review. Its family tree reaches back 3 generations to 1995. 2 later devices descend from it. It has been recalled once.

Decided
May 18, 2005
Review
35 days
Class
Class II
Product code
JOS
Recalls
1
Predicates
5

Its longest line of descent runs back 3 generations to K944992 (1995). 6 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor2009: 1 descendant cleared2006: 1 descendant clearedK944992 · Electrodes · Linvatec Corp. · 1995K993885 · Ultrablator Electrode · Linvatec Corp. · 2000K030720 · Ultrablator Electrode · Linvatec Corp. · 20031995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 2 descendants K030720 · Ultrablator Electrode · Linvatec Corp. · 2003K993885 · Ultrablator Electrode · Linvatec Corp. · 2000K944992 · Electrodes · Linvatec Corp. · 19950
  1. Root · 3 generations backK944992 ElectrodesLinvatec Corp. · 1995
  2. 2 generations backK993885 Ultrablator ElectrodeLinvatec Corp. · 2000
  3. 1 generation backK030720 Ultrablator ElectrodeLinvatec Corp. · 2003
  4. This deviceK050923 Lightwave Integrated Electrode Ablator and Light Wave Integrated…Conmed Linvatec · 2005
  5. Descendant · cites it directly · 0 of its ownK083281 EL Lightwave Suction Ablator, Model EL-2000-S, Lightwave Suction…Linvatec Corp. · 2009
  6. Descendant · cites it directly · 0 of its ownK052692 Suretek Medical Reprocessed Electrosurgical ElectrodesSuretek Medical · 2006

Most influential descendants

Showing 2 of 2 · See all
  • K083281EL Lightwave Suction Ablator, Model EL-2000-S, Lightwave Suction Ablator, Model…Linvatec Corp. · 2009 · cites it directly0descendants
  • K052692Suretek Medical Reprocessed Electrosurgical ElectrodesSuretek Medical · 2006 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K944992ElectrodesLinvatec Corp.1995—
−2K993885Ultrablator ElectrodeLinvatec Corp.2000—
−1K030720Ultrablator ElectrodeLinvatec Corp.2003—
0 (this device)K050923Lightwave Integrated Electrode Ablator and Light Wave Integrated Electrode…Conmed Linvatec2005—
+1K083281EL Lightwave Suction Ablator, Model EL-2000-S, Lightwave Suction Ablator…Linvatec Corp.2009—
+1K052692Suretek Medical Reprocessed Electrosurgical ElectrodesSuretek Medical2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910
200010
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200910
FIG. 1 — Lineage of K050923, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993885Ultrablator Electrode
Summary page 2
…an LightWave' Integrated ElectrodeSuto 5 10(k)# o• t April 5, 2005 Page 2 of 3 D. Predicate/Legally Marketed Devices K993885 - UltrAblator, Linvatec: Corporation K(002422 - VAPRTm 3.5mm 90' Suction Electrode for use with the VAPRTM Syst…
Linvatec Corp.2000—
*K002422Varpr 3.5MM 90 Suction Electrode for Use with Vapr System
Summary page 3
…ice subject to this application submission is attached. Substantially Equivalent Predicate Marketed Devices: K993885 - UltrAblator, Linvatec Corporation K002422 - VAPRTM 3.5mm 90' Suction Electrode for use with the VAPRTM System, Mitek Products K021299 - Heatwave Ha…
Mitek Products2000—
*K021299Conmed Heatwave Hand-Controlled Eletrode
Summary page 2
…Corporation K(002422 - VAPRTm 3.5mm 90' Suction Electrode for use with the VAPRTM System, Mitek Products K02 1299 - Heatwave Hand-Controlled Electrode, ConMed Corporation K(030720 - UltrAblator, Linvatec Corporation…
Conmedcorp2002—
*K030720Ultrablator Electrode
Summary page 2
…Device Description The ConMed/Linvatec LightWave TM Integrated Electrode Ablators are a modificati on of the currently marketed 3.2 mm UltrAblator electrode (Linvatec, K030720) and a Switch Pencil control handle (Linvatec, K983652) integrated into one device. One version incorporates hand- controlled but…
Linvatec Corp.2003—
*K983652Ablator Electrode
Summary page 2
…ntegrated Electrode Ablators are a modificati on of the currently marketed 3.2 mm UltrAblator electrode (Linvatec, K030720) and a Switch Pencil control handle (Linvatec, K983652) integrated into one device. One version incorporates hand- controlled buttons to actuate the "cut" and "coag" functions of the e…
Linvatec Corp.1999—

Also cleared against K993885: 2 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083281EL Lightwave Suction Ablator, Model EL-2000-S, Lightwave Suction Ablator, Model…Linvatec Corp.20090
K052692Suretek Medical Reprocessed Electrosurgical ElectrodesSuretek Medical20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 31, 2009Not classifiedZ-2051-2009Device DesignConMed Linvatec has determined there is a possibility that the device may not shut off if the ablate button is released, which may result in users/patients suffering burns.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 3.3 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K083281 (Linvatec Corp., 2009) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K993885.

How does its category compare?

All JOS devices

JOS Electrode, Electrosurgical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JOS

1in 2016

2016: 1
2007high 1 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOS

Too few JOS 510(k)s in any year since 2007 to give a median.

Recalls per year · JOS

2in 2019

2008: 12012: 22016: 12019: 2
2007high 2 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOS

6in 2025

2007: 292008: 372009: 622010: 142011: 52012: 72013: 52014: 72015: 132016: 62017: 72018: 102019: 172020: 122021: 52022: 82023: 92024: 32025: 6
2007high 62 (2009) · low 3 (2024)2025

MAUDE reports naming product code JOS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—029
20080—137
20090—062
20100—014
20110—05
20120—27
20130—05
20140—07
20150—013
20161—16
20170—07
20180—010
20190—217
20200—012
20210—05
20220—08
20230—09
20240—03
20250—06
20260—0—

What is it indicated for?

The ConMed/Linvatec LightWave"' Integrated Electrode Ablators are intended to be used for electrosurgical cutting and coagulation in shoulder, ankle, wrist, elbow, and knee arthroscopic procedures in a conductive fluid environment. Additionally, the suction/aspiration version also has capability for suctioning irrigation fluids.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050923: Lightwave Integrated Electrode Ablator and Light Wave Integrated Electrode Suction Ablator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K050923

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026