510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

GE Viewpoint

K050943510(k) TraditionalCleared May 23, 200539-day reviewClass II21 CFR 892.2050Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

General Electric Co. · LLZ System, Image Processing, Radiological

Cleared on May 23, 2005 as substantially equivalent to K041029 (Siemens Medi Cal Solutions, Inc., 2004) after a 39-day review. Its family tree reaches back 5 generations to 1995. 3 later devices descend from it. It has been recalled once.

Decided
May 23, 2005
Review
39 days
Class
Class II
Product code
LLZ
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K950279 (1995). 9 ancestors in all, the oldest K925965 (1993). 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor1996: 2 other ancestors1993: 1 other ancestor2010: 1 descendant cleared2011: 1 descendant cleared2016: 1 descendant clearedK950279 · TOMTECP90 · Tomtec Imaging Systems, Inc. · 1995K964803 · Univision Echocardiology System · Network Concepts, Inc. · 1997K980060 · Echolink · Mpacs, LLC · 1998K992259 · Echocardiography System Series 95000 · Camtronics, Ltd. · 1999K041029 · Kinetdx · Siemens Medi Cal Solutions, Inc. · 2004 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant K041029 · Kinetdx · Siemens Medi Cal Solutions, Inc. · 2004 · recalledK992259 · Echocardiography System Series 95000 · Camtronics, Ltd. · 1999K980060 · Echolink · Mpacs, LLC · 1998K964803 · Univision Echocardiology System · Network Concepts, Inc. · 1997K950279 · TOMTECP90 · Tomtec Imaging Systems, Inc. · 1995−50
  1. Root · 5 generations backK950279 TOMTECP90Tomtec Imaging Systems, Inc. · 1995
  2. 4 generations backK964803 Univision Echocardiology SystemNetwork Concepts, Inc. · 1997
  3. 3 generations backK980060 EcholinkMpacs, LLC · 1998
  4. 2 generations backK992259 Echocardiography System Series 95000Camtronics, Ltd. · 1999
  5. 1 generation backK041029 KinetdxSiemens Medi Cal Solutions, Inc. · 2004 · Recalled
  6. This deviceK050943 GE ViewpointGeneral Electric Co. · 2005
  7. Descendant · cites it directly · 1 of its ownK103458 Viewpoint 6GE Healthcare · 2011 · Recalled
  8. Descendant · 2 generations on · 0 of its ownK162743 ViewPoint 6GE Medical Systems Ultrasound and Primary Care Diagnostics · 2016 · Recalled
  9. Descendant · cites it directly · 0 of its ownK101166 4D-View 9.1 (4D View PC Software) Model: H48651SZGE Healthcare · 2010

Most influential descendants

Showing 3 of 3 · See all
  • K103458Viewpoint 6GE Healthcare · 2011 · cites it directly1descendant
  • K162743ViewPoint 6GE Medical Systems Ultrasound and Primary Care Diagnostics · 2016 · 2 generations on0descendants
  • K1011664D-View 9.1 (4D View PC Software) Model: H48651SZGE Healthcare · 2010 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K950279TOMTECP90Tomtec Imaging Systems, Inc.1995—
−4K964803Univision Echocardiology SystemNetwork Concepts, Inc.1997—
−3K980060EcholinkMpacs, LLC1998—
−2K992259Echocardiography System Series 95000Camtronics, Ltd.1999—
−1K041029KinetdxSiemens Medi Cal Solutions, Inc.20041
0 (this device)K050943GE ViewpointGeneral Electric Co.2005—
+1K103458Viewpoint 6GE Healthcare20112
+2K162743ViewPoint 6GE Medical Systems Ultrasound and Primary Care Diagnostics20161
+1K1011664D-View 9.1 (4D View PC Software) Model: H48651SZGE Healthcare2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199620
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201610
FIG. 1 — Lineage of K050943, 1995–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041029Kinetdx
Summary page 1
…ing System System, Image Processing, Radiological , 21 CFR 892.2050, 90-LLZ 3. Marketed Device: KinetDx@ Diagnostic Workstation for Radiology - K041 029, 6/8/2004, 90-LLZ Currently placed in commercial distribution by Siemens Medical Solutions. 4. Devce es…
Siemens Medi Cal Solutions, Inc.20041

Also cleared against K041029: 2 other devices →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103458Viewpoint 6GE Healthcare20111
K1011664D-View 9.1 (4D View PC Software) Model: H48651SZGE Healthcare20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 8, 2024Class IIZ-3145-2024Device DesignThe National Institute of Standards and Technology (NIST) has identified in the National Vulnerability Database (NVD) that NexGen Healthcare Mirth Connect, prior to version 4.4.1, has a potential vulnerability allowing the deserialization…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 6 companies; a new generation was cleared every 1.4 years on median.
  • 2 of the 9 ancestors have a recall of any class.
  • Of its descendants, K103458 (GE Healthcare, 2011) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K041029.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

The GE ViewPoint is intended to accept, transfer, display, store and process medical images and data, including the ability to measure, calculate, annotate and prepare and print patient examination reports primarily for diagnostic ultrasound.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050943: GE Viewpoint.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K050943

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026