510(k) · Cleared
Recalled · Class II · 2014Kodak Directview DR 7500 System, Model 8791345
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Eastman Kodak Company · KPR System, X-Ray, Stationary
Cleared on Jun 1, 2005 after a 16-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 32 later devices descend from it. It has been recalled 4 times.
- Decided
- Jun 1, 2005
- Review
- 16 days
- Class
- Class II
- Product code
- KPR
- Recalls
- 4
- Predicates
- 0
32 later devices descend from it, 2 citing it directly.
- This deviceK051258 Kodak Directview DR 7500 System, Model 8791345Eastman Kodak Company · 2005
- Descendant · cites it directly · 30 of its ownK091889 DRX-Evolution, Model VX3733-SysCarestream Health, Inc. · 2009 · Recalled
- Descendant · 2 generations on · 29 of its ownK141837 DRX-EvolutionCarestream Health, Inc. · 2015 · Recalled
- Descendant · 3 generations on · 26 of its ownK160997 GC85ASamsung Electronics Co., Ltd. · 2016
- Descendant · 4 generations on · 23 of its ownK163115 GC70Samsung Electronics Co., Ltd. · 2016
- Descendant · 4 generations on · 23 of its ownK162278 Digital Diagnostic Mobile X-ray SystemSamsung Electronics Co., Ltd. · 2016
Showing 5 of 32 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 32 · See all- K091889DRX-Evolution, Model VX3733-SysCarestream Health, Inc. · 2009 · cites it directly30descendants
- K141837DRX-EvolutionCarestream Health, Inc. · 2015 · 2 generations on29descendants
- K160997GC85ASamsung Electronics Co., Ltd. · 2016 · 3 generations on26descendants
- K163115GC70Samsung Electronics Co., Ltd. · 2016 · 4 generations on23descendants
- K162278Digital Diagnostic Mobile X-ray SystemSamsung Electronics Co., Ltd. · 2016 · 4 generations on23descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K051258 | Kodak Directview DR 7500 System, Model 8791345 | Eastman Kodak Company | 2005 | — |
| +1 | K091889 | DRX-Evolution, Model VX3733-Sys | Carestream Health, Inc. | 2009 | 3 |
| +2 | K141837 | DRX-Evolution | Carestream Health, Inc. | 2015 | 1 |
| +3 | K160997 | GC85A | Samsung Electronics Co., Ltd. | 2016 | — |
| +4 | K163115 | GC70 | Samsung Electronics Co., Ltd. | 2016 | — |
| +4 | K162278 | Digital Diagnostic Mobile X-ray System | Samsung Electronics Co., Ltd. | 2016 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2009 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 3 | 0 |
| 2017 | 2 | 0 |
| 2018 | 8 | 0 |
| 2019 | 2 | 0 |
| 2020 | 2 | 0 |
| 2021 | 2 | 0 |
| 2022 | 2 | 0 |
| 2023 | 1 | 0 |
| 2024 | 4 | 0 |
| 2025 | 1 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KPR device
- #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it? (32)
See all 322 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 9, 2014 | Class II | Z-2177-2014 | Device Design | An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering. Inspection of these parts… |
| Dec 26, 2013 | Class II | Z-0844-2014 | Software Manufacturing/Software Deployment | Carestream Health, Inc. has recalled DR 7500 Dual Detector System utilizing Version 5.X Software due to a possible patient image display error. |
| Nov 25, 2013 | Class II | Z-0798-2014 | Nonconforming Material/Component | Carestream Health Inc. has issued a recall for the Carestream DRX-Evolution (FF WS and CSJ WS) and the Kodak DirectView DR 7500 System due to the potential for unexpected device movement. |
| Mar 6, 2007 | Not classified | Z-0738-2007 | Other | The x-ray systems do not indicate the field size when the system's detector is turned off and the collimator is in manual mode. This results in the systems being noncompliant with the field indication performance standards of section… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K091889 (Carestream Health, Inc., 2009) has gone furthest, with 30 descendants of its own.
How does its category compare?
All KPR devicesKPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KPR
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KPR
155in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KPR
7in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · KPR
39in 2025
MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 31 | 10 | 120 |
| 2008 | 15 | 48 | 10 | 95 |
| 2009 | 9 | 35 | 5 | 61 |
| 2010 | 9 | 126 | 3 | 82 |
| 2011 | 9 | 152 | 5 | 103 |
| 2012 | 11 | 58 | 1 | 74 |
| 2013 | 5 | 163 | 5 | 24 |
| 2014 | 10 | 108 | 8 | 34 |
| 2015 | 18 | 81 | 1 | 59 |
| 2016 | 11 | 84 | 4 | 18 |
| 2017 | 6 | 75 | 3 | 31 |
| 2018 | 15 | 30 | 4 | 36 |
| 2019 | 12 | 33 | 1 | 77 |
| 2020 | 16 | 39 | 1 | 41 |
| 2021 | 6 | 55 | 0 | 37 |
| 2022 | 8 | 43 | 2 | 27 |
| 2023 | 9 | 76 | 4 | 39 |
| 2024 | 11 | 162 | 1 | 34 |
| 2025 | 8 | 155 | 7 | 39 |
| 2026 | 6 | 197 | 0 | — |
What is it indicated for?
The KODAK DirectView DR 7500 System is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K051258 · printed Nov 6, 2026
Cite this page
510(k) Report. “K051258: Kodak Directview DR 7500 System, Model 8791345.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051258
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026