Langston Dual Lumen Pressure Monitoring Catheter

K051395510(k) TraditionalCleared Jun 24, 200528-day reviewClass II21 CFR 870.1200Recalls 6 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vascular Solutions, Inc. · DQO Catheter, Intravascular, Diagnostic

Cleared on Jun 24, 2005 as substantially equivalent to K943739 (Mallinckrodt Medical, 1994) after a 28-day review. Its family tree reaches back 1 generation to 1994. 3 of the 10 devices in its family, counting itself, have a Class I recall. 8 later devices descend from it. It has been recalled 6 times, including 5 Class I recalls.

Decided
Jun 24, 2005
Review
28 days
Class
Class II
Product code
DQO
Recalls
6 (5 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K943739 (1994). 8 later devices descend from it, 3 citing it directly. 3 of the 10 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2006: 1 descendant cleared, 1 with a Class I recall12017: 1 descendant cleared2022: 1 descendant cleared2007: 2 descendants cleared, 1 with a Class I recall12023: 1 descendant cleared2008: 1 descendant cleared2009: 1 descendant clearedK943739 · Softouch Diagnostic Intravascular Catheter · Mallinckrodt Medical · 1994 · recalled1990200020102020
Descendants, by generation 1 generation forward: 3 descendants, 1 with a Class I recall12 generations forward: 3 descendants, 1 with a Class I recall13 generations forward: 1 descendant4 generations forward: 1 descendant K943739 · Softouch Diagnostic Intravascular Catheter · Mallinckrodt Medical · 1994 · recalled0
  1. Root · 1 generation backK943739 Softouch Diagnostic Intravascular CatheterMallinckrodt Medical · 1994 · Recalled
  2. This deviceK051395 Langston Dual Lumen Pressure Monitoring CatheterVascular Solutions, Inc. · 2005
  3. Descendant · cites it directly · 6 of its ownK061565 Langston Dual Lumen CatheterVascular Solutions, Inc. · 2006 · Recalled (Class I)
  4. Descendant · 2 generations on · 2 of its ownK062627 Innerchange Micro-Introducer Catheter, Models 7900 Through 7905Vascular Solutions, Inc. · 2007
  5. Descendant · cites it directly · 1 of its ownK170544 Langston dual lumen catheterVascular Solutions, Inc. · 2017
  6. Descendant · 3 generations on · 1 of its ownK073585 Innerchange Micro-Introducer Catheter, Models 7900, 7901, 7902…Vascular Solutions, Inc. · 2008
  7. Descendant · 2 generations on · 0 of its ownK222023 RayFlowHexacath · 2023

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2006: 12007: 22008: 12009: 12017: 12022: 12023: 1
2006high 2 (2007) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 8 · See all
  • K061565Langston Dual Lumen CatheterVascular Solutions, Inc. · 2006 · cites it directly6descendants
  • K062627Innerchange Micro-Introducer Catheter, Models 7900 Through 7905Vascular Solutions, Inc. · 2007 · 2 generations on2descendants
  • K170544Langston dual lumen catheterVascular Solutions, Inc. · 2017 · cites it directly1descendant
  • K073585Innerchange Micro-Introducer Catheter, Models 7900, 7901, 7902, 7903, 7904 and 7905Vascular Solutions, Inc. · 2008 · 3 generations on1descendant
  • K222023RayFlowHexacath · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K943739Softouch Diagnostic Intravascular CatheterMallinckrodt Medical19944
0 (this device)K051395Langston Dual Lumen Pressure Monitoring CatheterVascular Solutions, Inc.2005—
+1K061565Langston Dual Lumen CatheterVascular Solutions, Inc.2006Class I
+2K062627Innerchange Micro-Introducer Catheter, Models 7900 Through 7905Vascular Solutions, Inc.2007—
+1K170544Langston dual lumen catheterVascular Solutions, Inc.2017—
+3K073585Innerchange Micro-Introducer Catheter, Models 7900, 7901, 7902, 7903, 7904 and…Vascular Solutions, Inc.2008—
+2K222023RayFlowHexacath2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200611
200721
200810
200910
201710
202210
202310
FIG. 1 — Lineage of K051395, 1994–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQO device

  1. #1 K254278 Arterial Pressure Monitoring Set/Trayscore 14.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K943739Softouch Diagnostic Intravascular Catheter
Summary page 1
…ement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catheter (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased fr…
Mallinckrodt Medical19944
*—Langston Dual Lumen Pigtail Catheters
Named without a submission number
—
*—Cordis Dual Lumen Pressure Monitoring Catheter
Named without a submission number
—

Also cleared against K943739: 4 other devices →

What descends from it? (8)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061565Langston Dual Lumen Catheter Class IVascular Solutions, Inc.20066
K170544Langston dual lumen catheterVascular Solutions, Inc.20171
K221470Langston dual lumen catheterVascular Solutions, LLC20220

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 31, 2020Class IZ-0111-2021Process controlThe inner lumen of the Langston catheter may separate from the hub during pressure injection.
Aug 31, 2020Class IZ-0112-2021Process controlThe inner lumen of the Langston catheter may separate from the hub during pressure injection.
Aug 31, 2020Class IZ-0110-2021Process controlThe inner lumen of the Langston catheter may separate from the hub during pressure injection.
Mar 16, 2020Class IZ-1747-2020Process controlThe inner lumen of the Langston catheter may separate from the hub during pressure injection.
May 27, 2014Class IZ-1981-2014Component design/selectionThe inner catheter of some Langston V2 Dual Lumen Catheters have separated from the device hub during contrast pressure injections. This may require an intervention to retrieve the separated piece from the patient to prevent injury.
Sep 2, 2010Not classifiedZ-2608-2010Process controlInvestigation of a recent Device Experience Report has made us aware of a potential problem with our 6F Langston dual lumen pigtail catheter (Model 5540) with the following lot numbers: 548110, 548227, 548364, 548430, 548554 & 548723.…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K061565Langston Dual Lumen CatheterVascular Solutions, Inc. · 2006 · Cites it directly · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K062877Twin-Pass .023 Dual Access Catheter, Model 5230Vascular Solutions, Inc. · 2007 · 2 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K061565 (Vascular Solutions, Inc., 2006) has gone furthest, with 6 descendants of its own.
  • 4 other devices also cite its predicate K943739.

How does its category compare?

All DQO devices

DQO Catheter, Intravascular, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQO

2in 2025

2007: 82008: 32009: 82010: 72011: 62012: 152013: 22014: 32015: 102016: 42017: 102018: 102019: 62020: 32021: 12022: 42023: 42024: 42025: 22026 (partial year): 3
2007high 15 (2012) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQO

148in 2024

2007: 106 days median2008: 80 days median2009: 58 days median2010: 137 days median2011: 74 days median2012: 94 days median2014: 109 days median2015: 209 days median2016: 89 days median2017: 176 days median2018: 234 days median2019: 113 days median2020: 30 days median2022: 181 days median2023: 73 days median2024: 148 days median2026 (partial year): 85 days median
2007high 234 (2018) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQO

1in 2023

2007: 142008: 42009: 72010: 112011: 72012: 22013: 12015: 1, 1 Class I2016: 2, 1 Class I2017: 12018: 22019: 712020: 4, 4 Class I2022: 12023: 1, 1 Class I2026 (partial year): 24, 4 Class I
2007high 71 (2019) · low 0 (2014)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · DQO

656in 2025

2007: 5642008: 7262009: 9332010: 5562011: 6252012: 5322013: 7532014: 6012015: 6282016: 6142017: 8732018: 7772019: 3862020: 3122021: 3382022: 3792023: 3912024: 5112025: 656
2007high 933 (2009) · low 312 (2020)2025

MAUDE reports naming product code DQO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007810614564
20083804726
20098587933
2010713711556
20116747625
201215942532
20132—1753
201431090601
2015102091 (1)628
20164892 (1)614
2017101761873
2018102342777
2019611371386
20203304 (4)312
20211—0338
202241811379
20234731 (1)391
202441480511
20252—0656
202638524 (4)—

What is it indicated for?

The Vascular Solutions Langston TM Dual Lumen Pressure Monitoring Catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051395: Langston Dual Lumen Pressure Monitoring Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051395

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026