510(k) Report

510(k) · Cleared

Recalled · Class I · 2020

Langston Dual Lumen Catheter

K061565510(k) TraditionalCleared Aug 2, 200658-day reviewClass II21 CFR 870.1200Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vascular Solutions, Inc. · DQO Catheter, Intravascular, Diagnostic

Cleared on Aug 2, 2006 as substantially equivalent to K041909 (Vascular Solutions, Inc., 2004) and 2 other predicates after a 58-day review. Its family tree reaches back 2 generations to 1981. 3 of the 13 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 2 times, including a Class I recall.

Decided
Aug 2, 2006
Review
58 days
Class
Class II
Product code
DQO
Recalls
2 (1 Class I)
Predicates
3

Its longest line of descent runs back 2 generations to K810412 (1981). 6 ancestors in all. 6 later devices descend from it, 3 citing it directly. 3 of the 13 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2005: 2 other ancestors, 1 with a Class I recall11994: 1 other ancestor1991: 1 other ancestor2007: 2 descendants cleared, 1 with a Class I recall12017: 1 descendant cleared2008: 1 descendant cleared2023: 1 descendant cleared2009: 1 descendant clearedK810412 · Usci Cardiovascular Catheter · C.R. Bard, Inc. · 1981K041909 · Vascular Solutions Langston Dual Lumen Pigtail Catheter (DLP) · Vascular Solutions, Inc. · 200419801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 2 other ancestors 1 generation forward: 3 descendants, 1 with a Class I recall12 generations forward: 2 descendants3 generations forward: 1 descendant K041909 · Vascular Solutions Langston Dual Lumen Pigtail Catheter (DLP) · Vascular Solutions, Inc. · 2004K810412 · Usci Cardiovascular Catheter · C.R. Bard, Inc. · 19810
  1. Root · 2 generations backK810412 Usci Cardiovascular CatheterC.R. Bard, Inc. · 1981
  2. 1 generation backK041909 Vascular Solutions Langston Dual Lumen Pigtail Catheter (DLP)Vascular Solutions, Inc. · 2004
  3. This deviceK061565 Langston Dual Lumen CatheterVascular Solutions, Inc. · 2006
  4. Descendant · cites it directly · 2 of its ownK062627 Innerchange Micro-Introducer Catheter, Models 7900 Through 7905Vascular Solutions, Inc. · 2007
  5. Descendant · cites it directly · 1 of its ownK170544 Langston dual lumen catheterVascular Solutions, Inc. · 2017
  6. Descendant · 2 generations on · 1 of its ownK073585 Innerchange Micro-Introducer Catheter, Models 7900, 7901, 7902…Vascular Solutions, Inc. · 2008
  7. Descendant · 2 generations on · 0 of its ownK222023 RayFlowHexacath · 2023
  8. Descendant · 3 generations on · 0 of its ownK081846 Grebset Micro-Introducer Kit, Models 7950, 7951Vascular Solutions, Inc. · 2009 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2007: 22008: 12009: 12017: 12023: 1
2007high 2 (2007) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K062627Innerchange Micro-Introducer Catheter, Models 7900 Through 7905Vascular Solutions, Inc. · 2007 · cites it directly2descendants
  • K170544Langston dual lumen catheterVascular Solutions, Inc. · 2017 · cites it directly1descendant
  • K073585Innerchange Micro-Introducer Catheter, Models 7900, 7901, 7902, 7903, 7904 and 7905Vascular Solutions, Inc. · 2008 · 2 generations on1descendant
  • K222023RayFlowHexacath · 2023 · 2 generations on0descendants
  • K081846Grebset Micro-Introducer Kit, Models 7950, 7951Vascular Solutions, Inc. · 2009 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K810412Usci Cardiovascular CatheterC.R. Bard, Inc.1981—
−1K041909Vascular Solutions Langston Dual Lumen Pigtail Catheter (DLP)Vascular Solutions, Inc.2004—
0 (this device)K061565Langston Dual Lumen CatheterVascular Solutions, Inc.2006—
+1K062627Innerchange Micro-Introducer Catheter, Models 7900 Through 7905Vascular Solutions, Inc.2007—
+1K170544Langston dual lumen catheterVascular Solutions, Inc.2017—
+2K073585Innerchange Micro-Introducer Catheter, Models 7900, 7901, 7902, 7903, 7904 and…Vascular Solutions, Inc.2008—
+2K222023RayFlowHexacath2023—
+3K081846Grebset Micro-Introducer Kit, Models 7950, 7951Vascular Solutions, Inc.20091

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199410
200521

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200721
200810
200910
201710
202310
FIG. 1 — Lineage of K061565, 1981–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQO device

  1. #1 K254278 Arterial Pressure Monitoring Set/Trayscore 14.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041909Vascular Solutions Langston Dual Lumen Pigtail Catheter (DLP)
Summary page 1
…nfirmed the suitability of the device for its intended use. Predicate Devices: The Langston 7F/5F dual lumen catheter is substantially equivalent to the Langston 8F/5F (K041909), 7F/4F (K050168), and 6F/4F (K051395) dual lumen catheters. Conclusions: The Langston 7F outer lumen and 5F inner lumen is subs…
Vascular Solutions, Inc.2004—
*K050168Langston Dual Lumen Pressure Monitoring Catheter
Summary page 1
…bility of the device for its intended use. Predicate Devices: The Langston 7F/5F dual lumen catheter is substantially equivalent to the Langston 8F/5F (K041909), 7F/4F (K050168), and 6F/4F (K051395) dual lumen catheters. Conclusions: The Langston 7F outer lumen and 5F inner lumen is substantially equival…
Vascular Solutions, Inc.2005—
*K051395Langston Dual Lumen Pressure Monitoring Catheter
Summary page 1
…for its intended use. Predicate Devices: The Langston 7F/5F dual lumen catheter is substantially equivalent to the Langston 8F/5F (K041909), 7F/4F (K050168), and 6F/4F (K051395) dual lumen catheters. Conclusions: The Langston 7F outer lumen and 5F inner lumen is substantially equivalent to the currently…
Vascular Solutions, Inc.2005Class I

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K062627Innerchange Micro-Introducer Catheter, Models 7900 Through 7905Vascular Solutions, Inc.20072
K170544Langston dual lumen catheterVascular Solutions, Inc.20171
K062877Twin-Pass .023 Dual Access Catheter, Model 5230 Class IVascular Solutions, Inc.20070

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 31, 2020Class IZ-0113-2021Process controlThe inner lumen of the Langston catheter may separate from the hub during pressure injection.
Jun 9, 2011Not classifiedZ-2770-2011Packaging process controlVascular Solutions, Inc investigation of a recent Device Experience Report has made them aware of a potential problem with their 7F Langston dual lumen pigtail catheter (Model 5515) with lot number 548264. Recently it was reported that…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K051395Langston Dual Lumen Pressure Monitoring CatheterVascular Solutions, Inc. · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 15 Class I recalls
  • K062877Twin-Pass .023 Dual Access Catheter, Model 5230Vascular Solutions, Inc. · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 25 years and passes through 2 companies; a new generation was cleared every 23.6 years on median.
  • 2 of the 6 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K062627 (Vascular Solutions, Inc., 2007) has gone furthest, with 2 descendants of its own.

How does its category compare?

All DQO devices

DQO Catheter, Intravascular, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQO

2in 2025

2007: 82008: 32009: 82010: 72011: 62012: 152013: 22014: 32015: 102016: 42017: 102018: 102019: 62020: 32021: 12022: 42023: 42024: 42025: 22026 (partial year): 3
2007high 15 (2012) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQO

148in 2024

2007: 106 days median2008: 80 days median2009: 58 days median2010: 137 days median2011: 74 days median2012: 94 days median2014: 109 days median2015: 209 days median2016: 89 days median2017: 176 days median2018: 234 days median2019: 113 days median2020: 30 days median2022: 181 days median2023: 73 days median2024: 148 days median2026 (partial year): 85 days median
2007high 234 (2018) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQO

1in 2023

2007: 142008: 42009: 72010: 112011: 72012: 22013: 12015: 1, 1 Class I2016: 2, 1 Class I2017: 12018: 22019: 712020: 4, 4 Class I2022: 12023: 1, 1 Class I2026 (partial year): 24, 4 Class I
2007high 71 (2019) · low 0 (2014)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · DQO

656in 2025

2007: 5642008: 7262009: 9332010: 5562011: 6252012: 5322013: 7532014: 6012015: 6282016: 6142017: 8732018: 7772019: 3862020: 3122021: 3382022: 3792023: 3912024: 5112025: 656
2007high 933 (2009) · low 312 (2020)2025

MAUDE reports naming product code DQO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007810614564
20083804726
20098587933
2010713711556
20116747625
201215942532
20132—1753
201431090601
2015102091 (1)628
20164892 (1)614
2017101761873
2018102342777
2019611371386
20203304 (4)312
20211—0338
202241811379
20234731 (1)391
202441480511
20252—0656
202638524 (4)—

What is it indicated for?

The LangstonTM Dual Lumen Catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061565: Langston Dual Lumen Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K061565

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026