510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

GF-UE160-AL5 Ultrasound Endoscope

K051541510(k) TraditionalCleared Jul 1, 200521-day reviewClass II21 CFR 876.1500Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus America, Inc. · FDF Colonoscope And Accessories, Flexible/Rigid

Cleared on Jul 1, 2005 as substantially equivalent to K011314 (Aloka Co., Ltd., 2001) and 1 other predicate after a 21-day review. Its family tree reaches back 3 generations to 1986. 27 later devices descend from it. It has been recalled 2 times.

Decided
Jul 1, 2005
Review
21 days
Class
Class II
Product code
FDF
Recalls
2
Predicates
2

Its longest line of descent runs back 3 generations to K853585 (1986). 6 ancestors in all. 27 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1993: 2 other ancestors2025: 2 descendants cleared2007: 2 descendants cleared2009: 1 descendant cleared2012: 4 descendants cleared2011: 1 descendant cleared2026: 1 descendant cleared2016: 3 descendants cleared2018: 3 descendants cleared2013: 1 descendant cleared2020: 3 descendants cleared2023: 1 descendant cleared2014: 1 descendant cleared2019: 1 descendant cleared2022: 3 descendants clearedK853585 · Olympus Evs-Endoscopic Video Image & Data Sys · Olympus Corp. · 1986 · recalledK954451 · Evis 140 System · Olympus America, Inc. · 1996 · recalledK971660 · Olympus GF-UM130 Ultrasound Gastrovideoscope, Its Associated Accessories and Ancillary Equipment · Olympus America, Inc. · 1997 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 5 descendants3 generations forward: 6 descendants4 generations forward: 8 descendants5 generations forward: 5 descendants K971660 · Olympus GF-UM130 Ultrasound Gastrovideoscope, Its Associated Accessories and Ancillary Equipment · Olympus America, Inc. · 1997 · recalledK954451 · Evis 140 System · Olympus America, Inc. · 1996 · recalledK853585 · Olympus Evs-Endoscopic Video Image & Data Sys · Olympus Corp. · 1986 · recalled0+5
  1. Root · 3 generations backK853585 Olympus Evs-Endoscopic Video Image & Data SysOlympus Corp. · 1986 · Recalled
  2. 2 generations backK954451 Evis 140 SystemOlympus America, Inc. · 1996 · Recalled
  3. 1 generation backK971660 Olympus GF-UM130 Ultrasound Gastrovideoscope, Its Associated…Olympus America, Inc. · 1997 · Recalled
  4. This deviceK051541 GF-UE160-AL5 Ultrasound EndoscopeOlympus America, Inc. · 2005
  5. Descendant · cites it directly · 15 of its ownK063683 Olympus Xeu-M60A Endoscopic Ultrasound Center; XGF-UM180 Ultrasonic…Olympus Medical Systems Corporation · 2007
  6. Descendant · cites it directly · 8 of its ownK063847 Fujinon Ultrasonic Endoscope and Processor, Models EG-530UR…Fujinon, Inc. · 2007 · Recalled
  7. Descendant · 2 generations on · 7 of its ownK111243 Fujinon Ultrasonic ProcessorFujinon, Inc. · 2011
  8. Descendant · 2 generations on · 7 of its ownK092156 Cad 12, Colonoscopy Assist DeviceSoftscope Medical Technologies, Inc. · 2009
  9. Descendant · 2 generations on · 6 of its ownK121564 Endoscopic Ultrasound Center EU-Y0006Olympus Medical Systems Corp. · 2012

Showing 5 of 27 descendants. See all descendants

Descendants cleared per year

2007: 22009: 12011: 12012: 42013: 12014: 12016: 32018: 32019: 12020: 32022: 32023: 12025: 22026: 1
2007high 4 (2012) · low 0 (2008)2026

Most influential descendants

Showing 5 of 27 · See all
  • K063683Olympus Xeu-M60A Endoscopic Ultrasound Center; XGF-UM180 Ultrasonic GastrovideoscopeOlympus Medical Systems Corporation · 2007 · cites it directly15descendants
  • K063847Fujinon Ultrasonic Endoscope and Processor, Models EG-530UR, EG-530UT and SU-700Fujinon, Inc. · 2007 · cites it directly8descendants
  • K111243Fujinon Ultrasonic ProcessorFujinon, Inc. · 2011 · 2 generations on7descendants
  • K092156Cad 12, Colonoscopy Assist DeviceSoftscope Medical Technologies, Inc. · 2009 · 2 generations on7descendants
  • K121564Endoscopic Ultrasound Center EU-Y0006Olympus Medical Systems Corp. · 2012 · 2 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K853585Olympus Evs-Endoscopic Video Image & Data SysOlympus Corp.19862
−2K954451Evis 140 SystemOlympus America, Inc.19963
−1K971660Olympus GF-UM130 Ultrasound Gastrovideoscope, Its Associated Accessories and…Olympus America, Inc.19972
0 (this device)K051541GF-UE160-AL5 Ultrasound EndoscopeOlympus America, Inc.2005—
+1K063683Olympus Xeu-M60A Endoscopic Ultrasound Center; XGF-UM180 Ultrasonic…Olympus Medical Systems Corporation2007—
+1K063847Fujinon Ultrasonic Endoscope and Processor, Models EG-530UR, EG-530UT and SU-700Fujinon, Inc.20071
+2K111243Fujinon Ultrasonic ProcessorFujinon, Inc.2011—
+2K092156Cad 12, Colonoscopy Assist DeviceSoftscope Medical Technologies, Inc.2009—
+2K121564Endoscopic Ultrasound Center EU-Y0006Olympus Medical Systems Corp.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199320
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200720
200910
201110
201240
201310
201410
201630
201830
201910
202030
202230
202310
202520
202610
FIG. 1 — Lineage of K051541, 1986–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FDF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FDF device

  1. #1 K230332 Ambu® aScope™ Colon; Ambu® aBox™ 2score 20.1 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011314Olympus GF UC140P-AL5 Ultrasonic Endoscope with Aloka SSD-5500 Ultrasound SystemAloka Co., Ltd.20014
*K971660Olympus GF-UM130 Ultrasound Gastrovideoscope, Its Associated Accessories and Ancillary…
Summary page 2
…y 3. Predicate Devices: The Olympus GF-IJE I60-AL5 Endoscope is substantially equivalent to the Olympus GF-UCI14OP-AL5 Endoscope (K01I 1314) and the GF-UMI30 Endoscope (K971660). Both predicates share common indications and some features. 4. Device Description: The Olympus GF-UC14OP-AI5 is an electronic…
Olympus America, Inc.19972

Also cleared against K011314: 3 other devices →

What descends from it? (27)

See all 27

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063683Olympus Xeu-M60A Endoscopic Ultrasound Center; XGF-UM180 Ultrasonic GastrovideoscopeOlympus Medical Systems Corporation200715
K063847Fujinon Ultrasonic Endoscope and Processor, Models EG-530UR, EG-530UT and SU-700Fujinon, Inc.20078
K251859EVIS EUS Ultrasound Gastrointestinal Videoscope (Olympus GF-UE190)Olympus Medical Systems Corporation20251

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 25, 2023Class IIZ-0333-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Oct 16, 2020Class IIZ-0507-2021Device DesignA new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 2 companies; a new generation was cleared every 7.9 years on median.
  • 5 of the 6 ancestors have a recall of any class.
  • Of its descendants, K063683 (Olympus Medical Systems Corporation, 2007) has gone furthest, with 15 descendants of its own.
  • 3 other devices also cite its predicate K011314.

How does its category compare?

All FDF devices

FDF Colonoscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FDF

4in 2025

2007: 42008: 22009: 62010: 42011: 22012: 72013: 22014: 62015: 22016: 72017: 42018: 72019: 42020: 42021: 32022: 52023: 62024: 22025: 42026 (partial year): 5
2007high 7 (2012) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FDF

194in 2025

2007: 70 days median2009: 91 days median2010: 101 days median2012: 241 days median2014: 164 days median2016: 98 days median2017: 36 days median2018: 101 days median2019: 56 days median2020: 250 days median2021: 30 days median2022: 147 days median2023: 203 days median2025: 194 days median2026 (partial year): 147 days median
2007high 250 (2020) · low 30 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FDF

8in 2023

2015: 12016: 12020: 12023: 8
2007high 8 (2023) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FDF

17,231in 2025

2007: 812008: 682009: 832010: 482011: 772012: 882013: 632014: 602015: 1002016: 1702017: 3702018: 3882019: 4172020: 6872021: 2,4442022: 8,3452023: 11,7482024: 12,1282025: 17,231
2007high 17,231 (2025) · low 48 (2010)2025

MAUDE reports naming product code FDF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007470081
20082—068
2009691083
20104101048
20112—077
20127241088
20132—063
20146164060
20152—1100
20167981170
20174360370
201871010388
20194560417
202042501687
202133002,444
2022514708,345
20236203811,748
20242—012,128
20254194017,231
202651470—

What is it indicated for?

The Olympus GF-UE160-AL5 is an ultrasonic gastro video endoscope to be used with an Aloka diagnostic ultrasound system, video system center, light source, video monitor, endo-therapy accessories for endoscopic ultrasound imaging of the gastrointestinal wall, bile and pancreatic ducts and surrounding organs. The Aloka SSD-5000 (K033311) is an all-digital diagnostic ultrasonic scanner with a digital beam former supporting gray scale, spectral Doppler and…

Read the full statement (136 words)

The Olympus GF-UE160-AL5 is an ultrasonic gastro video endoscope to be used with an Aloka diagnostic ultrasound system, video system center, light source, video monitor, endo-therapy accessories for endoscopic ultrasound imaging of the gastrointestinal wall, bile and pancreatic ducts and surrounding organs.

The Aloka SSD-5000 (K033311) is an all-digital diagnostic ultrasonic scanner with a digital beam former supporting gray scale, spectral Doppler and Color Flow imaging. Depending on the probe, the Aloka SSD-5000 may be used for diagnostic ultrasound imaging in Cardiac, Gynecological, Neurological, Obstetrical, Neonatal, Pediatric, Perinatal, Radiological, Vascular, Urological, Trauma and Surgical applications. The Aloka SSD-5000 is a Track 3 system and is not indicated for ophthalmic applications.

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051541: GF-UE160-AL5 Ultrasound Endoscope.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051541

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026