510(k) Report

510(k) · Cleared

Recalled · 2009

Omni Sphincterotome

K052051510(k) SpecialCleared Aug 5, 20057-day reviewClass II21 CFR 876.4300Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Wilson-Cook Medical, Inc. · KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Cleared on Aug 5, 2005 as substantially equivalent to K033203 (Wilson-Cook Medical, Inc., 2003) after a 7-day review. Its family tree reaches back 3 generations to 1997. 4 later devices descend from it. It has been recalled 2 times.

Decided
Aug 5, 2005
Review
7 days
Class
Class II
Product code
KNS
Recalls
2
Predicates
1

Its longest line of descent runs back 3 generations to K970053 (1997). 4 ancestors in all, the oldest K901443 (1990). 4 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1990: 1 other ancestor2018: 1 descendant cleared2025: 2 descendants cleared2020: 1 descendant clearedK970053 · Ultratomerx · Boston Scientific Corp · 1997K013153 · Autotome RX Model # 4515, 4516 · Boston Scientific Corp · 2001 · recalledK033203 · Wilson-Cook Tri-Tome Select Plus · Wilson-Cook Medical, Inc. · 20031990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 1 descendant K033203 · Wilson-Cook Tri-Tome Select Plus · Wilson-Cook Medical, Inc. · 2003K013153 · Autotome RX Model # 4515, 4516 · Boston Scientific Corp · 2001 · recalledK970053 · Ultratomerx · Boston Scientific Corp · 19970
  1. Root · 3 generations backK970053 UltratomerxBoston Scientific Corp · 1997
  2. 2 generations backK013153 Autotome RX Model # 4515, 4516Boston Scientific Corp · 2001 · Recalled
  3. 1 generation backK033203 Wilson-Cook Tri-Tome Select PlusWilson-Cook Medical, Inc. · 2003
  4. This deviceK052051 Omni SphincterotomeWilson-Cook Medical, Inc. · 2005
  5. Descendant · cites it directly · 3 of its ownK172288 Fusion OMNI SphincterotomeWilson-Cook Medical Inc./Cook Endoscopy · 2018
  6. Descendant · 2 generations on · 1 of its ownK192339 TeslaTome Bipolar Sphincterotome, Tesla Bipolar Active CordWilson-Cook Medical, Inc. · 2020
  7. Descendant · 2 generations on · 0 of its ownK251104 SureTome™ SW Sphincterotome with DomeTipWilson-Cook Medical · 2025
  8. Descendant · 3 generations on · 0 of its ownK243568 Teslatome Bipolar Sphincterotome; Tesla BiCord Active CordWilson-Cook Medical, Inc. · 2025 · Recalled

Most influential descendants

Showing 4 of 4 · See all
  • K172288Fusion OMNI SphincterotomeWilson-Cook Medical Inc./Cook Endoscopy · 2018 · cites it directly3descendants
  • K192339TeslaTome Bipolar Sphincterotome, Tesla Bipolar Active CordWilson-Cook Medical, Inc. · 2020 · 2 generations on1descendant
  • K251104SureTome™ SW Sphincterotome with DomeTipWilson-Cook Medical · 2025 · 2 generations on0descendants
  • K243568Teslatome Bipolar Sphincterotome; Tesla BiCord Active CordWilson-Cook Medical, Inc. · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K970053UltratomerxBoston Scientific Corp1997—
−2K013153Autotome RX Model # 4515, 4516Boston Scientific Corp200132
−1K033203Wilson-Cook Tri-Tome Select PlusWilson-Cook Medical, Inc.2003—
0 (this device)K052051Omni SphincterotomeWilson-Cook Medical, Inc.2005—
+1K172288Fusion OMNI SphincterotomeWilson-Cook Medical Inc./Cook Endoscopy2018—
+2K192339TeslaTome Bipolar Sphincterotome, Tesla Bipolar Active CordWilson-Cook Medical, Inc.2020—
+2K251104SureTome™ SW Sphincterotome with DomeTipWilson-Cook Medical2025—
+3K243568Teslatome Bipolar Sphincterotome; Tesla BiCord Active CordWilson-Cook Medical, Inc.20251

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202010
202520
FIG. 1 — Lineage of K052051, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KNS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KNS device

  1. #1 K241271 Aqua Medical RF Vapor Ablation Systemscore 17.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033203Wilson-Cook Tri-Tome Select Plus
Summary page 1
…21 CFR § 876.4300 PREDICATE DEVICES: Wilson-Cook Triple Tome Select Plus Sphincterotome (k033203) Wilson-Cook's OMNIlt Sphincterotome is INTENDED USE: intended for…
Wilson-Cook Medical, Inc.2003—

Also cleared against K033203: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172288Fusion OMNI SphincterotomeWilson-Cook Medical Inc./Cook Endoscopy20183

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 12, 2009Not classifiedZ-1598-2009Process controlThe wire control port of these devices may be missing due to omission of a manufacturing activity.
Oct 30, 2008Not classifiedZ-0325-2009Process controlThe wire control port may have been omitted during manufacture.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 2.2 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K172288 (Wilson-Cook Medical Inc./Cook Endoscopy, 2018) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K033203.

How does its category compare?

All KNS devices

KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KNS

11in 2025

2007: 12008: 32009: 32010: 42011: 12012: 22013: 22014: 12015: 32016: 32017: 32018: 52019: 72020: 42021: 42022: 32023: 92024: 32025: 112026 (partial year): 2
2007high 11 (2025) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KNS

171in 2025

2008: 61 days median2009: 212 days median2010: 168 days median2015: 252 days median2016: 30 days median2017: 48 days median2018: 62 days median2019: 66 days median2020: 212 days median2021: 139 days median2022: 182 days median2023: 83 days median2024: 198 days median2025: 171 days median
2007high 252 (2015) · low 30 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KNS

14in 2024

2007: 32008: 52009: 22011: 22016: 62017: 12021: 412024: 142026 (partial year): 4
2007high 41 (2021) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KNS

1,504in 2025

2007: 1642008: 5872009: 1,0272010: 9082011: 8602012: 8482013: 4492014: 4052015: 4062016: 4572017: 4462018: 5632019: 6682020: 9062021: 8142022: 9132023: 9762024: 1,5332025: 1,504
2007high 1,533 (2024) · low 164 (2007)2025

MAUDE reports naming product code KNS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—3164
20083615587
2009321221,027
201041680908
20111—2860
20122—0848
20132—0449
20141—0405
201532520406
20163306457
20173481446
20185620563
20197660668
202042120906
2021413941814
202231820913
20239830976
20243198141,533
20251117101,504
20262—4—

What is it indicated for?

Used for cannulation of the ductal system and sphincterotomy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052051: Omni Sphincterotome.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052051

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026