Solanas Posterior Stabilization System

K052201510(k) TraditionalCleared Nov 10, 200590-day reviewClass II21 CFR 888.3050Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Alphatec Spine, Inc. · KWP Appliance, Fixation, Spinal Interlaminal

Cleared on Nov 10, 2005 as substantially equivalent to K002733 (Depuyacromed, 2000) after a 90-day review. Its family tree reaches back 4 generations to 1985. 36 later devices descend from it. It has been recalled 3 times.

Decided
Nov 10, 2005
Review
90 days
Class
Class II
Product code
KWP
Recalls
3
Predicates
1

Its longest line of descent runs back 4 generations to K853033 (1985). 9 ancestors in all. 36 later devices descend from it, 8 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1999: 4 other ancestors2011: 4 descendants cleared2012: 2 descendants cleared2007: 2 descendants cleared2013: 2 descendants cleared2014: 2 descendants cleared2010: 2 descendants cleared2017: 2 descendants cleared2009: 1 descendant cleared2015: 2 descendants cleared2021: 2 descendants cleared2022: 3 descendants cleared2019: 2 descendants cleared2020: 4 descendants cleared2016: 1 descendant cleared2023: 2 descendants cleared2018: 1 descendant cleared2026: 2 descendants clearedK853033 · Hartshill Rect & Wires, Spinal Fixation System · A.W. Showell (Surgicraft), Ltd. · 1985K965221 · Ransford Cervical Fixation System · Surgicraft, Ltd. · 1997K982322 · Synthes Spine Occipital-Cervical Plate/Rod and Hook System · Synthes (Usa) · 1998K002733 · Summit Occipito-Cervico-Thoracic (Oct) Spinal System · Depuyacromed · 20001990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 5 other ancestors 1 generation forward: 8 descendants2 generations forward: 12 descendants3 generations forward: 8 descendants4 generations forward: 5 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant K002733 · Summit Occipito-Cervico-Thoracic (Oct) Spinal System · Depuyacromed · 2000K982322 · Synthes Spine Occipital-Cervical Plate/Rod and Hook System · Synthes (Usa) · 1998K965221 · Ransford Cervical Fixation System · Surgicraft, Ltd. · 1997K853033 · Hartshill Rect & Wires, Spinal Fixation System · A.W. Showell (Surgicraft), Ltd. · 19850+5
  1. Root · 4 generations backK853033 Hartshill Rect & Wires, Spinal Fixation SystemA.W. Showell (Surgicraft), Ltd. · 1985
  2. 3 generations backK965221 Ransford Cervical Fixation SystemSurgicraft, Ltd. · 1997
  3. 2 generations backK982322 Synthes Spine Occipital-Cervical Plate/Rod and Hook SystemSynthes (Usa) · 1998
  4. 1 generation backK002733 Summit Occipito-Cervico-Thoracic (Oct) Spinal SystemDepuyacromed · 2000
  5. This deviceK052201 Solanas Posterior Stabilization SystemAlphatec Spine, Inc. · 2005
  6. Descendant · cites it directly · 18 of its ownK071380 Modification to Solanas Posterior Stabilization SystemAlphatec Spine, Inc. · 2007 · Recalled
  7. Descendant · cites it directly · 6 of its ownK071905 Posterior Cervicothoracic Spinal Fixation System (PCFS)Lanx, LLC · 2007
  8. Descendant · 2 generations on · 5 of its ownK173522 Solanas® Posterior Stabilization SystemAlphatec Spine, Inc. · 2017
  9. Descendant · 2 generations on · 5 of its ownK111076 Solanas Avalon Posterior Fixation SystemAlphatec Spine, Inc. · 2011
  10. Descendant · 2 generations on · 5 of its ownK092656 Lanx Posterior Cervicothoracic Spinal Fixation SystemLanx, Inc. · 2009

Showing 5 of 36 descendants. See all descendants

Descendants cleared per year

2007: 22009: 12010: 22011: 42012: 22013: 22014: 22015: 22016: 12017: 22018: 12019: 22020: 42021: 22022: 32023: 22026: 2
2007high 4 (2011) · low 0 (2008)2026

Most influential descendants

Showing 5 of 36 · See all
  • K071380Modification to Solanas Posterior Stabilization SystemAlphatec Spine, Inc. · 2007 · cites it directly18descendants
  • K071905Posterior Cervicothoracic Spinal Fixation System (PCFS)Lanx, LLC · 2007 · cites it directly6descendants
  • K173522Solanas® Posterior Stabilization SystemAlphatec Spine, Inc. · 2017 · 2 generations on5descendants
  • K111076Solanas Avalon Posterior Fixation SystemAlphatec Spine, Inc. · 2011 · 2 generations on5descendants
  • K092656Lanx Posterior Cervicothoracic Spinal Fixation SystemLanx, Inc. · 2009 · 2 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K853033Hartshill Rect & Wires, Spinal Fixation SystemA.W. Showell (Surgicraft), Ltd.1985—
−3K965221Ransford Cervical Fixation SystemSurgicraft, Ltd.1997—
−2K982322Synthes Spine Occipital-Cervical Plate/Rod and Hook SystemSynthes (Usa)1998—
−1K002733Summit Occipito-Cervico-Thoracic (Oct) Spinal SystemDepuyacromed2000—
0 (this device)K052201Solanas Posterior Stabilization SystemAlphatec Spine, Inc.2005—
+1K071380Modification to Solanas Posterior Stabilization SystemAlphatec Spine, Inc.20073
+1K071905Posterior Cervicothoracic Spinal Fixation System (PCFS)Lanx, LLC2007—
+2K173522Solanas® Posterior Stabilization SystemAlphatec Spine, Inc.2017—
+2K111076Solanas Avalon Posterior Fixation SystemAlphatec Spine, Inc.2011—
+2K092656Lanx Posterior Cervicothoracic Spinal Fixation SystemLanx, Inc.2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199940
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200720
200910
201020
201140
201220
201320
201420
201520
201610
201720
201810
201920
202040
202120
202230
202320
202620
FIG. 1 — Lineage of K052201, 1985–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002733Summit Occipito-Cervico-Thoracic (Oct) Spinal System
Summary page 1
…rgery 7. Tumors VII Substantial Equivalence: The SOLANAS TM Posterior Stabilization System is substantially equivalent to Summit TM system (K002733). VIII Performance Data: Mechanical testing of the SOLANAS TM Posterior Stabilization System was performed and…
Depuyacromed2000—

Also cleared against K002733: 15 other devices →

What descends from it? (36)

See all 36

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071380Modification to Solanas Posterior Stabilization SystemAlphatec Spine, Inc.200718
K071905Posterior Cervicothoracic Spinal Fixation System (PCFS)Lanx, LLC20076
K122292Reliance Posterior Cervical- Thoracic SystemReliance Medical Systems, LLC20124
K112025Sure Lok Mini Posterior Cervical/Upper Thoracic SystemSpinal USA20114
K133214Blackbird Spinal SystemChoicespine, LP20143
K100191Lanx Posterior Cervicothoracic Spinal Fixation System (PCFS)Lanx, Inc.20103
K120467Spinal Elements Posterior Cervical/Thoracic Spinal SystemSpinal Elements, Inc.20121
K131119Solanas Posterior Stabilization SystemAlphatec Spine, Inc.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 1, 2010Not classifiedZ-0932-2012Device DesignThese lots were not manufactured correctly in that the instrument does not have the durability required to repeatedly perform over its full service life and, therefore, may experience breakage in the tips.
Aug 5, 2009Not classifiedZ-0502-2010Process controlAlphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40 degrees.
Aug 5, 2009Not classifiedZ-0503-2010Process controlAlphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40 degrees.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 5 companies; a new generation was cleared every 4.9 years on median.
  • None of the 9 ancestors has a recall.
  • Of its descendants, K071380 (Alphatec Spine, Inc., 2007) has gone furthest, with 18 descendants of its own.
  • 15 other devices also cite its predicate K002733.

How does its category compare?

All KWP devices

KWP Appliance, Fixation, Spinal Interlaminal: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KWP

9in 2015

2007: 142008: 132009: 122010: 142011: 122012: 182013: 122014: 132015: 9
2007high 18 (2012) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWP

113in 2015

2007: 96 days median2008: 50 days median2009: 89 days median2010: 186 days median2011: 67 days median2012: 112 days median2013: 99 days median2014: 138 days median2015: 113 days median
2007high 186 (2010) · low 50 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWP

2in 2022

2007: 22008: 12009: 232010: 22011: 62012: 12013: 32015: 12016: 32018: 12019: 22022: 2
2007high 23 (2009) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWP

494in 2025

2007: 2102008: 5942009: 6822010: 4742011: 7922012: 9182013: 1,0902014: 1,0222015: 1,3482016: 1,3792017: 8642018: 6572019: 5182020: 7962021: 5252022: 3352023: 3322024: 4362025: 494
2007high 1,379 (2016) · low 210 (2007)2025

MAUDE reports naming product code KWP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714962210
200813501594
2009128923682
2010141862474
201112676792
2012181121918
2013129931,090
20141313801,022
2015911311,348
20160—31,379
20170—0864
20180—1657
20190—2518
20200—0796
20210—0525
20220—2335
20230—0332
20240—0436
20250—0494
20260—0—

What is it indicated for?

Indications for Use: It is intended that this device, in any system configuration, be removed after development of solid fusion mass. Hook components are indicated for use at C1-C7. Polyaxial screws are limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. These screws are not intended for placement in the cervical spine.

1. Degenerative disk disease (DDD), defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.

2. Spondylolisthesis

3. Spinal stenosis

4. Fracture/Dislocation

5. Atlanto/Axial fracture with instability

6. Revision of previous cervical spine surgery

7. Tumors

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052201: Solanas Posterior Stabilization System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052201

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026