DAR-8000F

K052500510(k) AbbreviatedCleared Oct 17, 200534-day reviewClass II21 CFR 892.1600Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shimadzu Medical Systems · IZI System, X-Ray, Angiographic

Cleared on Oct 17, 2005 as substantially equivalent to K050925 (Shimadzu Corp., 2005) after a 34-day review. Its family tree reaches back 1 generation to 2005. 8 later devices descend from it. It has been recalled 2 times.

Decided
Oct 17, 2005
Review
34 days
Class
Class II
Product code
IZI
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K050925 (2005). 8 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2007: 1 descendant cleared2014: 1 descendant cleared2015: 1 descendant cleared2019: 2 descendants cleared2021: 1 descendant cleared2022: 2 descendants clearedK050925 · Digital Radiography System Radspeed Safire · Shimadzu Corp. · 200520052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 3 descendants5 generations forward: 1 descendant K050925 · Digital Radiography System Radspeed Safire · Shimadzu Corp. · 20050+5
  1. Root · 1 generation backK050925 Digital Radiography System Radspeed SafireShimadzu Corp. · 2005
  2. This deviceK052500 DAR-8000FShimadzu Medical Systems · 2005
  3. Descendant · cites it directly · 6 of its ownK131075 X-Ray TV System Sonial Vision G4Shimadzu Corp. · 2014
  4. Descendant · 2 generations on · 5 of its ownK142341 X-Ray TV System Sonialvision G4Shimadzu Corp. · 2015 · Recalled
  5. Descendant · 3 generations on · 4 of its ownK190373 SonialVision G4Shimadzu Corporation · 2019 · Recalled
  6. Descendant · 4 generations on · 1 of its ownK212212 Multitom Rax with True2scale Body Scan OptionSiemens Medical Solutions · 2022
  7. Descendant · 5 generations on · 0 of its ownK221281 Multitom RaxSiemens Medical Solutions · 2022 · Recalled

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2007: 12014: 12015: 12019: 22021: 12022: 2
2007high 2 (2019) · low 0 (2008)2026

Most influential descendants

Showing 5 of 8 · See all
  • K131075X-Ray TV System Sonial Vision G4Shimadzu Corp. · 2014 · cites it directly6descendants
  • K142341X-Ray TV System Sonialvision G4Shimadzu Corp. · 2015 · 2 generations on5descendants
  • K190373SonialVision G4Shimadzu Corporation · 2019 · 3 generations on4descendants
  • K212212Multitom Rax with True2scale Body Scan OptionSiemens Medical Solutions · 2022 · 4 generations on1descendant
  • K221281Multitom RaxSiemens Medical Solutions · 2022 · 5 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K050925Digital Radiography System Radspeed SafireShimadzu Corp.2005—
0 (this device)K052500DAR-8000FShimadzu Medical Systems2005—
+1K131075X-Ray TV System Sonial Vision G4Shimadzu Corp.2014—
+2K142341X-Ray TV System Sonialvision G4Shimadzu Corp.20152
+3K190373SonialVision G4Shimadzu Corporation20192
+4K212212Multitom Rax with True2scale Body Scan OptionSiemens Medical Solutions2022—
+5K221281Multitom RaxSiemens Medical Solutions20221

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
201410
201510
201920
202110
202220
FIG. 1 — Lineage of K052500, 2005–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IZI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZI device

  1. #1 K230727 SPY Portable Handheld Imaging (SPY-PHI) Systemscore 23.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050925Digital Radiography System Radspeed Safire
Summary page 3
…is total patient population. As mentioned at "Intended Use" in this clearance letter, the intended use of DAR-8000f is the same as that of predicate device of DAR-7000(K050925). Consequently, the indications for use of DAR-8000f are also the same as that of DAR-7000, and there are no additional indic…
Shimadzu Corp.2005—

Also cleared against K050925: 2 other devices →

What descends from it? (8)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131075X-Ray TV System Sonial Vision G4Shimadzu Corp.20146
K072006DAR-9400FShimadzu Corp.20070

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 12, 2018Class IIZ-2060-2018Software designWhen selecting serial radiography with a pulse rate of 7.5fps (including selecting a preset or changing the pulse rate using a temporary edit function) it was observed the indicated "irradiation time" on the X-ray Generator Console, as…
Apr 3, 2014Class IIZ-1569-2014Software designShimadzu Corporation is recalling the Shimadzu Digital X-Ray System because they have received a report of a lost image under certain circumstances.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K131075 (Shimadzu Corp., 2014) has gone furthest, with 6 descendants of its own.
  • 2 other devices also cite its predicate K050925.

How does its category compare?

All IZI devices

IZI System, X-Ray, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IZI

2in 2025

2007: 72008: 62009: 22010: 22011: 42012: 12013: 12014: 42016: 12017: 22018: 22019: 22020: 12021: 12022: 12023: 32024: 22025: 22026 (partial year): 2
2007high 7 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZI

81in 2023

2007: 28 days median2008: 68 days median2011: 104 days median2014: 111 days median2023: 81 days median
2007high 111 (2014) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZI

1in 2021

2007: 62008: 522009: 182010: 112011: 62012: 62013: 62014: 52015: 142016: 72017: 12018: 22019: 22020: 22021: 12026 (partial year): 1
2007high 52 (2008) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZI

2,171in 2025

2007: 922008: 1012009: 1322010: 2522011: 3772012: 772013: 582014: 1232015: 362016: 1022017: 882018: 492019: 302020: 142021: 292022: 2652023: 1,2042024: 1,3322025: 2,171
2007high 2,171 (2025) · low 14 (2020)2025

MAUDE reports naming product code IZI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007728692
200866852101
20092—18132
20102—11252
201141046377
20121—677
20131—658
201441115123
20150—1436
20161—7102
20172—188
20182—249
20192—230
20201—214
20211—129
20221—0265
202338101,204
20242—01,332
20252—02,171
20262—1—

What is it indicated for?

This device is intended to be used for the radiography in the hospital with X-ray devices.

This device is operated and used by the physicians and X-ray technologist.

The object of this device is total patient population.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K052500: DAR-8000F.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052500

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026